How influential is the duration of contrast material bolus injection in perfusion CT? evaluation in a swine model.

Sonja M Kandel1, Henning Meyer, Markus Boehnert, Bernice Hoppel, Narinder Singh Paul, Patrik Rogalla.   

Abstract

PURPOSE: To analyze the effect of the duration of contrast material bolus injection on perfusion values in a swine model by using the maximum slope method.
MATERIALS AND METHODS: This study was approved by the institutional animal care committee. Twenty pigs (weight range, 63-77 kg) underwent dynamic volume computed tomography (CT) of the kidneys during suspended respiration. Before the CT examination, a miniature cuff-shaped ultrasonographic flow probe encircling the right renal artery was surgically implanted in each pig to obtain true perfusion values. Two sequential perfusion CT series were performed in 30 seconds, each comprising 30 volumes with identical parameters (100 kV, 200 mAs, 0.5 sec rotation time). The duration of contrast material bolus (0.5 mL/kg of body weight) was 3.8 seconds in the first series (short bolus series) and 11.5 seconds in the second series (long bolus series), and the injection flow rate was adapted accordingly. In each pig, cortical kidney volume was determined by using the volume with the highest cortical enhancement. CT perfusion values were calculated for both series by using the maximum slope method and were statistically compared and correlated with the true perfusion values from the flow probe by using linear regression analysis.
RESULTS: Mean true perfusion and CT perfusion values (in minutes(-1)) for the short bolus series were 1.95 and 2.03, respectively (P = .22), and for the long bolus series, they were 2.02 and 1.92, respectively (P = .12). CT perfusion showed very good correlation with true perfusion in both the short (slope, 1.01; 95% confidence interval: 0.91, 1.11) and long (slope, 0.92; 95% confidence interval: 0.78, 1.04) series. On the basis of the regression analysis, CT perfusion values in the short bolus series were overestimated by 1% and those in the long bolus series were underestimated by 8%.
CONCLUSION: Duration of contrast material bolus injection does not influence CT perfusion values substantially. The longer, clinically preferred intravenous injection scheme is sufficiently accurate for CT perfusion. © RSNA, 2013.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24029648     DOI: 10.1148/radiol.13130024

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  4 in total

1.  Evaluation of the effect of image noise on CT perfusion measurements using digital perfusion phantoms.

Authors:  Stephan Skornitzke; Jessica Hirsch; Hans-Ulrich Kauczor; Wolfram Stiller
Journal:  Eur Radiol       Date:  2018-10-11       Impact factor: 5.315

2.  Arterial input function placement effect on computed tomography lung perfusion maps.

Authors:  Laura Jimenez-Juan; Hatem Mehrez; Chris Dey; Shabnam Homampour; Anastasia Oikonomou; Fatima Ursani; Narinder Paul
Journal:  Quant Imaging Med Surg       Date:  2016-02

3.  Relationship between low tube voltage (70 kV) and the iodine delivery rate (IDR) in CT angiography: An experimental in-vivo study.

Authors:  Michael M Lell; Ulrike Fleischmann; Hubertus Pietsch; Johannes G Korporaal; Ulrike Haberland; Andreas H Mahnken; Thomas G Flohr; Michael Uder; Gregor Jost
Journal:  PLoS One       Date:  2017-03-20       Impact factor: 3.240

4.  Comparison of Renal Blood Flow Using Maximum Slope-Based Computed Tomography Perfusion and Ultrasound Flow Probe in Healthy Dogs.

Authors:  Sang-Kwon Lee; Youjung Jang; Jin-Woo Jung; Hyejin Je; Jihye Choi
Journal:  Front Vet Sci       Date:  2020-10-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.