Literature DB >> 24016832

Radiation dose reduction in soft tissue neck CT using adaptive statistical iterative reconstruction (ASIR).

Behroze Vachha1, Harald Brodoefel, Carol Wilcox, David B Hackney, Gul Moonis.   

Abstract

PURPOSE: To compare objective and subjective image quality in neck CT images acquired at different tube current-time products (275 mAs and 340 mAs) and reconstructed with filtered-back-projection (FBP) and adaptive statistical iterative reconstruction (ASIR).
MATERIALS AND METHODS: HIPAA-compliant study with IRB approval and waiver of informed consent. 66 consecutive patients were randomly assigned to undergo contrast-enhanced neck CT at a standard tube-current-time-product (340 mAs; n = 33) or reduced tube-current-time-product (275 mAs, n = 33). Data sets were reconstructed with FBP and 2 levels (30%, 40%) of ASIR-FBP blending at 340 mAs and 275 mAs. Two neuroradiologists assessed subjective image quality in a blinded and randomized manner. Volume CT dose index (CTDIvol), dose-length-product (DLP), effective dose, and objective image noise were recorded. Signal-to-noise ratio (SNR) was computed as mean attenuation in a region of interest in the sternocleidomastoid muscle divided by image noise.
RESULTS: Compared with FBP, ASIR resulted in a reduction of image noise at both 340 mAs and 275 mAs. Reduction of tube current from 340 mAs to 275 mAs resulted in an increase in mean objective image noise (p=0.02) and a decrease in SNR (p = 0.03) when images were reconstructed with FBP. However, when the 275 mAs images were reconstructed using ASIR, the mean objective image noise and SNR were similar to those of the standard 340 mAs CT images reconstructed with FBP (p>0.05). Subjective image noise was ranked by both raters as either average or less-than-average irrespective of the tube current and iterative reconstruction technique.
CONCLUSION: Adapting ASIR into neck CT protocols reduced effective dose by 17% without compromising image quality.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  ASIR; CT; Neck; Radiation-dose

Mesh:

Year:  2013        PMID: 24016832     DOI: 10.1016/j.ejrad.2013.08.014

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  9 in total

1.  Impact of model-based iterative reconstruction on image quality of contrast-enhanced neck CT.

Authors:  S Gaddikeri; J B Andre; J Benjert; D S Hippe; Y Anzai
Journal:  AJNR Am J Neuroradiol       Date:  2014-10-09       Impact factor: 3.825

2.  Adaptive statistical iterative reconstruction use for radiation dose reduction in pediatric lower-extremity CT: impact on diagnostic image quality.

Authors:  Amisha Shah; Mitchell Rees; Erica Kar; Kimberly Bolton; Vincent Lee; Ashok Panigrahy
Journal:  Skeletal Radiol       Date:  2017-12-26       Impact factor: 2.199

3.  Reaching for better image quality and lower radiation dose in head and neck CT: advanced modeled and sinogram-affirmed iterative reconstruction in combination with tube voltage adaptation.

Authors:  Andrea I Schmid; Michael Uder; Michael M Lell
Journal:  Dentomaxillofac Radiol       Date:  2016-09-08       Impact factor: 2.419

4.  Dose reduction potential of iterative reconstruction algorithms in neck CTA-a simulation study.

Authors:  Stephan Ellmann; Ferdinand Kammerer; Thomas Allmendinger; Michael Brand; Rolf Janka; Matthias Hammon; Michael M Lell; Michael Uder; Manuel Kramer
Journal:  Dentomaxillofac Radiol       Date:  2016-08-19       Impact factor: 2.419

5.  Head CT: Image quality improvement with ASIR-V using a reduced radiation dose protocol for children.

Authors:  Hyun Gi Kim; Ho-Joon Lee; Seung-Koo Lee; Hyun Ji Kim; Myung-Joon Kim
Journal:  Eur Radiol       Date:  2017-01-23       Impact factor: 5.315

6.  Effect of an iterative reconstruction quantum noise reduction technique on computed tomography radiomic features.

Authors:  Joseph J Foy; Mena Shenouda; Sahar Ramahi; Samuel Armato; Daniel Thomas Ginat
Journal:  J Med Imaging (Bellingham)       Date:  2020-12-30

7.  Radiation exposure of ovarian cancer patients: contribution of CT examinations performed on different MDCT (16 and 64 slices) scanners and image quality evaluation: an observational study.

Authors:  Stefania Rizzo; Daniela Origgi; Sarah Brambilla; Federica De Maria; Riccardo Foà; Sara Raimondi; Nicoletta Colombo; Massimo Bellomi
Journal:  Medicine (Baltimore)       Date:  2015-05       Impact factor: 1.889

8.  Variation in tube voltage for pediatric neck 64VCT: Effect on radiation dose and image quality.

Authors:  Li-Guo Chen; Ping-An Wu; Hsing-Yang Tu; Ming-Huei Sheu; Li-Chuan Huang
Journal:  PLoS One       Date:  2021-11-12       Impact factor: 3.240

9.  Effect of reduced z-axis scan coverage on diagnostic performance and radiation dose of neck computed tomography in patients with suspected cervical abscess.

Authors:  Jakob Weiss; Michael Maurer; Dominik Ketelsen; Mike Notohamiprodjo; Dominik Zinsser; Julian L Wichmann; Konstantin Nikolaou; Fabian Bamberg; Ahmed E Othman
Journal:  PLoS One       Date:  2017-07-05       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.