| Literature DB >> 24014467 |
Atsushi Ogata1, Makoto Sonobe2, Noriyuki Kato2, Tomosato Yamazaki2, Hiromichi Kasuya2, Go Ikeda2, Shunichiro Miki2, Toshio Matsushima3.
Abstract
PURPOSE: To evaluate the clinical outcome and MRI findings after carotid artery stenting (CAS) without post-dilatation.Entities:
Keywords: Artery; Cervical; Intervention; Stent; Technique
Mesh:
Year: 2013 PMID: 24014467 PMCID: PMC4145433 DOI: 10.1136/neurintsurg-2013-010873
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1An octogenarian with symptomatic carotid artery stenosis who underwent carotid artery stenting using the GuardWire system. (A) Black blood T1-weighted MRI showing a high-intensity plaque lesion in the right internal carotid artery (white arrow). (B) Angiogram of the right common carotid artery showing severe stenosis of the internal carotid artery with a long lesion. (C) Post-procedural angiogram of the right common carotid artery (pre-dilatation with 4×30 mm balloon and stent deployment with 10×24 mm carotid Wallstent) showing improvement of the internal carotid artery stenosis.
Procedural outcomes
| Total (n=176) | Asymptomatic (n=68) | Symptomatic (n=108) | |
|---|---|---|---|
| Any stroke | 4 (2.3%) | 1 (1.5%) | 3 (2.8%) |
| Ischemic stroke | 2 (1.1%) | 1 (1.5%) | 1 (0.9%) |
| Hemorrhagic stroke | 2 (1.1%) | 0 | 2 (1.9%) |
| Myocardial infarction | 0 | 0 | 0 |
| Death | 0 | 0 | 0 |
| Major adverse event* | 4 (2.3%) | 1 (1.5%) | 3 (2.8%) |
| TIA | 3 (1.7%) | 1 (1.5%) | 2 (1.9%) |
| DWI high intensity | 26 (14.8%) | 7 (10.3%) | 19 (17.6%) |
*Major adverse event defined as any stroke/myocardial infarction/death.
DWI, diffusion-weighted imaging; TIA, transient ischemic attack.
Comparison of the procedural outcomes according to the residual stenosis immediately after carotid artery stenting (CAS)
| Residual stenosis rate immediately after CAS | |||
|---|---|---|---|
| >30% (n=27) | ≤30% (n=149) | p Value | |
| Restenosis | 3 (11.1%) | 3 (2.0%) | 0.047 |
| In-stent thrombosis | 2 (7.4%) | 0 | 0.023 |
| Restenosis or in-stent thrombosis | 5 (18.5%) | 3 (2.0%) | 0.002 |
Comparison of the procedural outcomes between patients treated with the GuardWire (GW) and FilterWire EZ (FW) systems
| Total (n=176) | GW (n=116) | FW (n=60) | p Value | |
|---|---|---|---|---|
| Any stroke | 4 (2.3%) | 4 (3.4%) | 0 | 0.30 |
| Ischemic stroke | 2 (1.1%) | 2 (1.7%) | 0 | 0.55 |
| Hemorrhagic stroke | 2 (1.1%) | 2 (1.7%) | 0 | 0.55 |
| DWI high intensity | 26 (14.8%) | 16 (13.8%) | 10 (16.7%) | 0.82 |
DWI, diffusion-weighted imaging.