Imaging capabilities of the Inveon SPECT system using single-and multipinhole collimators.

Frederic Boisson1, David Zahra, Arvind Parmar, Marie-Claude Gregoire, Steven R Meikle, Hasar Hamse, Anthonin Reilhac.   

Abstract

UNLABELLED: The Inveon small-animal SPECT system comes with several types of multipinhole collimator plates. We evaluate here the performance measurements of the Inveon SPECT system using 6 different collimators: 3 dedicated for mouse imaging and 3 for rat imaging.
METHODS: The measured performance parameters include the sensitivity, the spatial resolution using line sources, the ultra-micro Derenzo phantom, the recovery coefficient and the noise measurements using the National Electrical Manufacturers Association NU-4 image quality phantom, obtained with the 2 reconstruction algorithms available with the Inveon Acquisition Workplace, version 1.5-the 3-dimensional ordered-subset expectation maximization (3DOSEM) and the 3-dimensional maximum a posteriori (3DMAP). Further, the overall performance of the system is illustrated by an animal experiment.
RESULTS: The results show that the Inveon SPECT scanner offers a spatial resolution, measured at the center of the field of view, ranging from 0.6 to 1 mm with the collimator plates dedicated to mouse imaging and from 1.2 to less than 2 mm with rat collimator plates. The system sensitivity varies from 29 to 404 cps/MBq for mouse collimators and from 53 to 175 cps/MBq for rat collimators. The image quality study showed that 3DMAP allows better noise reduction while preserving the recovery coefficient, compared with other regularization strategies such as the premature termination of the 3DOSEM reconstruction or 3DOSEM followed by gaussian filtering.
CONCLUSION: The acquisition parameters, such as the collimator set and the radius of rotation, offer a wide range of possibilities to apply to a large number of biologic studies. However, special care must be taken because this increase in sensitivity can be offset by image degradation, such as image artifacts caused by projection overlap and statistical noise due to a higher number of iterations required for convergence. 3DMAP allowed better noise reduction while maintaining relatively constant recovery coefficients, as compared with other reconstruction strategies.

Entities:  

Keywords:  NEMA; image quality; multi-pinhole; single photon emission computed tomography; small-animal

Mesh:

Substances:

Year:  2013        PMID: 24009279     DOI: 10.2967/jnumed.112.117572

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  5 in total

Review 1.  Standardization of Small Animal Imaging-Current Status and Future Prospects.

Authors:  Julia G Mannheim; Firat Kara; Janine Doorduin; Kerstin Fuchs; Gerald Reischl; Sayuan Liang; Marleen Verhoye; Felix Gremse; Laura Mezzanotte; Marc C Huisman
Journal:  Mol Imaging Biol       Date:  2018-10       Impact factor: 3.488

2.  A whole-body dual-modality radionuclide optical strategy for preclinical imaging of metastasis and heterogeneous treatment response in different microenvironments.

Authors:  Gilbert O Fruhwirth; Seckou Diocou; Philip J Blower; Tony Ng; Greg E D Mullen
Journal:  J Nucl Med       Date:  2014-03-06       Impact factor: 10.057

3.  Preclinical molecular imaging: development of instrumentation for translational research with small laboratory animals.

Authors:  Jorge Mejia; Ana Claudia Camargo Miranda; Ana Claudia Ranucci Durante; Larissa Rolim de Oliveira; Marycel Rosa Felisa Figols de Barboza; Katerin Taboada Rosell; Daniele Pereira Jardim; Alexandre Holthausen Campos; Marilia Alves Dos Reis; Marcela Forli Catanoso; Orfa Yineth Galvis-Alonso; Francisco Romero Cabral
Journal:  Einstein (Sao Paulo)       Date:  2016 Jul-Sep

4.  Accurate molecular imaging of small animals taking into account animal models, handling, anaesthesia, quality control and imaging system performance.

Authors:  Christian Vanhove; Jens P Bankstahl; Stefanie D Krämer; Eric Visser; Nicola Belcari; Stefaan Vandenberghe
Journal:  EJNMMI Phys       Date:  2015-11-11

5.  Capabilities of multi-pinhole SPECT with two stationary detectors for in vivo rat imaging.

Authors:  Jan P Janssen; Jan V Hoffmann; Takayuki Kanno; Naoko Nose; Jan-Peter Grunz; Masahisa Onoguchi; Xinyu Chen; Constantin Lapa; Andreas K Buck; Takahiro Higuchi
Journal:  Sci Rep       Date:  2020-10-29       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.