Literature DB >> 24003371

Inadvertent Self-Detachment of Solitaire AB Stent during the Mechanical Thrombectomy for Recanalization of Acute Ischemic Stroke: Lessons Learned from the Removal of Stent via Surgical Embolectomy.

Dong-Hun Kang1, Jaechan Park, Yang-Ha Hwang, Yong-Sun Kim.   

Abstract

We recently experienced self-detachment of the Solitaire stent during mechanical thrombectomy of acute ischemic stroke. Then, we tried to remove the detached stent and to recanalize the occlusion, but failed with endovascular means. The following diffusion weighted image MRI revealed no significant increase in infarction size, therefore, we performed surgical removal of the stent to rescue the patient and to elucidate the reason why the self-detachment occurred. Based upon the operative findings, the stent grabbed the main thrombi but inadvertently detached at a severely tortuous, acutely angled, and circumferentially calcified segment of the internal carotid artery. Postoperative angiography demonstrated complete recanalization of the internal carotid artery. The patient's neurological deficits gradually improved, and the modified Rankin scale score was 2 at three months after surgery. In the retrospective case review, bone window images of the baseline computed tomography (CT) scan corresponded to the operative findings. According to this finding, we hypothesized that bone window images of a baseline CT scan can play a role in terms of anticipating difficult stent retrieval before the procedure.

Entities:  

Keywords:  CT scan; Self-detachment; Stent; Stroke; Thrombectomy

Year:  2013        PMID: 24003371      PMCID: PMC3756129          DOI: 10.3340/jkns.2013.53.6.360

Source DB:  PubMed          Journal:  J Korean Neurosurg Soc        ISSN: 1225-8245


  7 in total

1.  Mechanical thrombectomy with the Solitaire AB device in large artery occlusions of the anterior circulation: a pilot study.

Authors:  Carlos Castaño; Laura Dorado; Cristina Guerrero; Monica Millán; Meritxell Gomis; Natalia Perez de la Ossa; Mar Castellanos; M Rosa García; Sira Domenech; Antoni Dávalos
Journal:  Stroke       Date:  2010-06-10       Impact factor: 7.914

2.  Direct thrombus retrieval using the reperfusion catheter of the penumbra system: forced-suction thrombectomy in acute ischemic stroke.

Authors:  D-H Kang; Y-H Hwang; Y-S Kim; J Park; O Kwon; C Jung
Journal:  AJNR Am J Neuroradiol       Date:  2010-11-18       Impact factor: 3.825

3.  Solitaire flow restoration device versus the Merci Retriever in patients with acute ischaemic stroke (SWIFT): a randomised, parallel-group, non-inferiority trial.

Authors:  Jeffrey L Saver; Reza Jahan; Elad I Levy; Tudor G Jovin; Blaise Baxter; Raul G Nogueira; Wayne Clark; Ronald Budzik; Osama O Zaidat
Journal:  Lancet       Date:  2012-08-26       Impact factor: 79.321

4.  Treatment of acute middle cerebral artery occlusion with a Solitaire AB stent: preliminary experience.

Authors:  S Nayak; G Ladurner; M Killer
Journal:  Br J Radiol       Date:  2010-12       Impact factor: 3.039

5.  Extended superciliary approach for middle cerebral artery embolectomy after unsuccessful endovascular recanalization therapy: technical note.

Authors:  Jaechan Park; Yang-Ha Hwang; Yongsun Kim
Journal:  Neurosurgery       Date:  2009-12       Impact factor: 4.654

6.  Intracranial carotid artery atherosclerosis: prevalence and risk factors in the general population.

Authors:  Daniel Bos; Maggee J M van der Rijk; Tychon E A Geeraedts; Albert Hofman; Gabriel P Krestin; Jacqueline C M Witteman; Aad van der Lugt; M Arfan Ikram; Meike W Vernooij
Journal:  Stroke       Date:  2012-05-08       Impact factor: 7.914

7.  Intra-arterial thrombolysis using double devices: mechanicomechanical or chemicomechanical techniques.

Authors:  Hyun Park; Gyo Jun Hwang; Sung-Chul Jin; Jae Seung Bang; Chang Wan Oh; O-Ki Kwon
Journal:  J Korean Neurosurg Soc       Date:  2012-02-29
  7 in total
  5 in total

1.  Inadvertent Stent Retriever Detachment: A Multicenter Case Series and Review of Device Experience FDA Reports.

Authors:  Hesham Masoud; Thanh N Nguyen; Coleman O Martin; William E Holloway; Sudheer Ambekar; Dileep R Yavagal; Diogo C Haussen; Raul Nogueira; Diego J Lozano; Ajit Puri; Ayman Quateen; Daniela Iancu; Michael G Abraham; Michael Chen; Sonal Mehta; Tim Malisch; Franklin Marden; Robin Novakovic; Daniel Roy; Alain Weill; Alexander M Norbash
Journal:  Interv Neurol       Date:  2015-11-20

2.  Unexpected Detachment of Solitaire Stents during Mechanical Thrombectomy.

Authors:  Sung Tae Kim; Sung-Chul Jin; Hae Woong Jeong; Jung Hwa Seo; Sam Yeol Ha; Hae Wook Pyun
Journal:  J Korean Neurosurg Soc       Date:  2014-12-31

3.  Solitaire Stentectomy Using a Stent-Retriever Technique in a Porcine Model.

Authors:  Andreas Simgen; Michael Kettner; Frida Juliane Webelsiep; Toshiki Tomori; Ruben Mühl-Benninghaus; Umut Yilmaz; Pervinder Bhogal; Matthias W Laschke; Michael D Menger; Wolfgang Reith; Philipp Dietrich
Journal:  Clin Neuroradiol       Date:  2020-05-18       Impact factor: 3.649

4.  Retained Solitaire FR device after mechanical thrombectomy: Case review and management strategies.

Authors:  Jay P Kinariwala; Gary B Rajah; Ali W Luqman
Journal:  Brain Circ       Date:  2018-12-31

5.  Direct Aspiration Catheter Fracture and Retrieval during Neurothrombectomy.

Authors:  Pouria Moshayedi; Ashutosh P Jadhav
Journal:  Interv Neurol       Date:  2018-01-25
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.