Comparison of image quality using a X-ray stereotactical whole-body system and a direct flat-panel X-ray device in examinations of the pelvis and knee.

K B Krug1, C Weber1, H Schwabe1, N-M Sinzig1, B Wein2, D Müller1, K Wegmann3, S Peters4, V Sendler4, K Ewen4, M Hellmich5, D Maintz1.   

Abstract

PURPOSE: The novel biplanar X-ray unit "EOS", EOS Imaging, allows to acquire simultaneously 2 perpendicular planes of full-length spine and limbs and to measure spatially correct angles based on the acquired image data sets. This is to be seen alongside with a low spatial resolution, high investment costs and high operating expenses. The use of the biplanar X-ray unit in morphology based scelettal radiography might improve the cost-benefit-relation. Thus, the purpose of this study was to compare image quality of the EOS-unit and the flat panel (FP)-technology as reference in a clinical setting.
MATERIALS AND METHODS: All 114 patients of the Orthopedic Hospital Dept., who had a biplanar full-length lower limb radiograph and a FP-examination of the pelvis and/or the knee with maximum time interval of 3 months without changes in the clinical and radiological findings were included in the study. All X-ray examinations had been carried out due to clinical indications. Secondary captures comparable to the FP-images were extracted from the electronic EOS-image data sets. 4 radiologists independently from each other compared the visualization of normal anatomical structures of the pseudonymous EOS- and FP-images in a randomized order.
RESULTS: In the overwiew of all readers and all sceletal regions image quality of the FP-images was considered being superior in a mean of 83 ± 13 % standard deviation of the pair comparisons (minimum 48 %, maximum 100 %). Image quality of the EOS-images was assessed as being superior in 2 ± 3 % of the cases (0 %, 10 %). Image quality of 0.8 ± 3 % of the FP-images (0 %, 17 %) and 30 ± 34 % (0 %, 100 %) of the EOS-images was estimated as diagnostically inadequate. 30 ± 33 % of the pair comparisons (0 %, 100 %) showed a diagnostically inadequate image quality of the EOS-images and a diagnostically good image quality of the FP-images.
CONCLUSION: Image quality of biplanar full-length lower limb X-ray examinations is not suitable to be used for the diagnostic assessment of the morphological bone structure using the currently available technological setting. KEY POINTS: ▶ biplanar full-length lower limb X-ray examinations ▶ plat-panel radiography ▶ image quality. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2013        PMID: 23999782     DOI: 10.1055/s-0033-1350441

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  4 in total

1.  Stereoradiography imaging motion artifact: does it affect radiographic measures after spinal instrumentation?

Authors:  Anne-Laure Simon; Emmanuelle Ferrero; A N Larson; Kenton R Kaufman
Journal:  Eur Spine J       Date:  2016-03-03       Impact factor: 3.134

2.  Accuracy of biplanar linear radiography versus conventional radiographs when used for lower limb and implant measurements.

Authors:  Chen Xi Kasia Chua; Si Heng Sharon Tan; Andrew Kean Seng Lim; James Hoipo Hui
Journal:  Arch Orthop Trauma Surg       Date:  2021-01-02       Impact factor: 3.067

3.  Incidental extraspinal imaging findings on adult EOS full body radiographs: prevalence and clinical importance.

Authors:  Lily Wood; Christopher Martin; David Polly; Samuel Luchsinger; Takashi Takahashi
Journal:  BMC Med Imaging       Date:  2021-05-17       Impact factor: 1.930

4.  EOS(®) biplanar X-ray imaging: concept, developments, benefits, and limitations.

Authors:  Elias Melhem; Ayman Assi; Rami El Rachkidi; Ismat Ghanem
Journal:  J Child Orthop       Date:  2016-02-16       Impact factor: 1.548

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.