Late failure of annealed highly cross-linked polyethylene acetabular liner.

Daisuke Hara1, Yasuharu Nakashima, Takuaki Yamamoto, Shinshichiro Higashihara, Mitsugu Todo, Masanobu Hirata, Mio Akiyama, Yukihide Iwamoto.   

Abstract

Highly cross-linked polyethylene (HXPE) in total hip arthroplasty (THA) has been shown to significantly decrease wear rates compared with conventional polyethylene (CPE). However, crosslinking, thermal treatment and oxidation can decrease the mechanical properties of PE, and several cases of fracture of remelted HXPE liners were reported. We present, for the first time, unexpected failures of THA with the use of annealed HXPE liners in two patients occurring at 7 and 8 years after operation. Operative findings revealed dislocated liners from the metal shell and a fracture of the superior rim at the rim-dome junction in both liners. Scanning electron microscopy showed that the cracks initiated at the rim and propagated toward the articular surface. Both liners showed generally a low amount of oxidation (less than 1.00) at the articular surface and low wear rates; however, oxidation at the rim was relatively higher (mean 1.55). These findings suggested that decreased mechanical properties at the rim-dome junction due to cross-linking, annealing and oxidation might have been caused breakage of the HXPE liners after a long implantation time, although the annealed HXPE achieved low degree of wear.
© 2013 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Annealing; Cross-linked polyethylene; Fracture; Mechanical property; Total hip arthroplasty

Mesh:

Substances:

Year:  2013        PMID: 23995342     DOI: 10.1016/j.jmbbm.2013.08.003

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  5 in total

1.  The Impact of Free Radical Stabilization Techniques on in vivo Mechanical Changes in Highly Cross-Linked Polyethylene Acetabular Liners.

Authors:  Michael Decker; Amber Price; Aria Khalili; Robert Klassen; Mary Jane Walzak; Matthew Teeter; Richard McCalden; Brent Lanting
Journal:  Orthop Res Rev       Date:  2021-08-17

2.  Gait instability may indicate liner failure in patients with total hip arthroplasty. A report of three cases.

Authors:  F Díaz-Dilernia; M Lattore; G Zanotti; F Comba; F Piccaluga; M Buttaro
Journal:  Ann R Coll Surg Engl       Date:  2021-08-20       Impact factor: 1.951

3.  Incarcerated fracture fragments of Longevity polyethylene liners after total hip arthroplasty.

Authors:  Michael A Yee; Thomas J O'Keefe; Suzanne Winter
Journal:  Arthroplast Today       Date:  2016-01-13

4.  Highly Cross-linked Polyethylene Liner Dissociation from a Cement-less Modular Acetabular Shell: Two Case Reports.

Authors:  Shunsuke Kawano; Motoki Sonohata; Masaru Kitajima; Masaaki Mawatari
Journal:  Open Orthop J       Date:  2016-12-26

5.  Future Prospects for Clinical Applications of Nanocarbons Focusing on Carbon Nanotubes.

Authors:  Naoto Saito; Hisao Haniu; Kaoru Aoki; Naoyuki Nishimura; Takeshi Uemura
Journal:  Adv Sci (Weinh)       Date:  2022-06-26       Impact factor: 17.521

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.