Literature DB >> 23978572

One-year outcomes from an international study of the Ovation Abdominal Stent Graft System for endovascular aneurysm repair.

Manish Mehta1, Francisco E Valdés2, Thomas Nolte3, Gregory J Mishkel4, William D Jordan5, Bruce Gray6, Mark K Eskandari7, Charles Botti8.   

Abstract

OBJECTIVE: This study evaluated 1-year safety and effectiveness outcomes of the United States regulatory trial for the Ovation Abdominal Stent Graft System (TriVascular Inc, Santa Rosa, Calif) for endovascular repair of abdominal aortic aneurysms (AAAs).
METHODS: This prospective, multicenter, single-arm trial was conducted at 36 sites in the United States, Germany, and Chile to evaluate the safety and effectiveness of the Ovation stent graft. From November 2009 to May 2011, 161 patients (88% males; mean age, 73 ± 8 years) with AAAs (mean diameter, 54 ± 9 mm) were treated with the Ovation stent graft. The main body is a modular two-docking limb device with a 14F outer diameter delivery system, active suprarenal fixation, and polymer-filled proximal rings that accommodate the aortic neck for seal. Main inclusion criteria included proximal aortic neck length ≥ 7 mm, inner neck diameter between 16 and 30 mm, distal iliac landing zones length ≥ 10 mm, and diameter between 8 and 20 mm. Patients were treated under a common protocol, including clinical and imaging follow-up at discharge, 30 days, 6 months, and annually through 5 years. A Clinical Events Committee adjudicated adverse events, an independent imaging core laboratory analyzed imaging, and a Data Safety and Monitoring Board provided study oversight. Complete 1-year follow-up data were available for this report.
RESULTS: The Ovation stent graft was implanted successfully in 161 patients (100%), including 69 (42.9%) by percutaneous access. General anesthesia was used in 106 patients (65.8%). Technical success was 100%, and mean procedure time was 110 minutes. Median procedural blood loss was 150 mL, and median hospital stay was 1 day. The 30-day major adverse event rate was 2.5%. At 1 year, AAA-related and all-cause mortality were 0.6% and 2.5%, respectively. Major adverse event and serious adverse event rates through 1 year were 6.2% and 38.5%, respectively. The 1-year treatment success rate was 99.3%. The imaging core laboratory reported no stent graft migration or type I, III, or IV endoleaks. At 1 year, type II endoleaks were identified in 34% of patients, and AAA enlargement was identified in one patient (0.7%). No AAA rupture or conversion to open surgery was reported. AAA-related secondary procedures were performed in 10 patients (6.2%) for 12 findings, including endoleak (six), aortic main body stenosis (three), and iliac limb stenosis or occlusion (three).
CONCLUSIONS: The 1-year results of the Ovation Abdominal Stent Graft System demonstrate excellent safety and effectiveness in treatment of patients with AAAs, particularly in patients with challenging anatomic characteristics, including short aortic necks and narrow iliac arteries. Longer-term follow-up is needed.
Copyright © 2014. Published by Mosby, Inc.

Entities:  

Mesh:

Year:  2013        PMID: 23978572     DOI: 10.1016/j.jvs.2013.06.065

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  18 in total

1.  Ultra-low profile polymer-filled stent graft for abdominal aortic aneurysm treatment: a two-year follow-up.

Authors:  Anna Maria Ierardi; Dimitrios Tsetis; Christos Ioannou; Domenico Laganà; Chiara Floridi; Mario Petrillo; Antonio Pinto; Gabriele Piffaretti; Gianpaolo Carrafiello
Journal:  Radiol Med       Date:  2015-01-29       Impact factor: 3.469

2.  Best surgical option for arch extension of type B dissection: the endovascular approach.

Authors:  Toru Kuratani
Journal:  Ann Cardiothorac Surg       Date:  2014-05

3.  Routine use of an aortic balloon to resolve possible inflow stenosis induced by the inflatable ring fixation mechanism of the Ovation endograft.

Authors:  Christos V Ioannou; Nikolaos Kontopodis; Efstratios Georgakarakos; Elias Kehagias; Eleni Metaxa; Stella Lioudaki; Yannis Papaharilaou; Dimitrios Tsetis
Journal:  Radiol Med       Date:  2016-07-23       Impact factor: 3.469

4.  Endovascular aneurysm repair with the Ovation TriVascular Stent Graft System utilizing a predominantly percutaneous approach under local anaesthesia.

Authors:  C V Ioannou; N Kontopodis; E Kehagias; A Papaioannou; A Kafetzakis; G Papadopoulos; D Pantidis; D Tsetis
Journal:  Br J Radiol       Date:  2015-05-12       Impact factor: 3.039

5.  Occlusive Shrinkage of Ovation Endograft Presenting as Acute Lower Limb Ischemia: Effective Endovascular Management.

Authors:  Paolo Bianchi; Filippo Scalise; Andrea Mortara; Guido Lanzillo; Giuseppe Scardina; Santi Trimarchi; Gianfranco Parati; Valerio Tolva
Journal:  Aorta (Stamford)       Date:  2017-02-01

Review 6.  Advanced Endovascular Approaches in the Management of Challenging Proximal Aortic Neck Anatomy: Traditional Endografts and the Snorkel Technique.

Authors:  Jon G Quatromoni; Ksenia Orlova; Paul J Foley
Journal:  Semin Intervent Radiol       Date:  2015-09       Impact factor: 1.513

7.  Treatment of Hostile Proximal Necks During Endovascular Aneurysm Repair.

Authors:  Tulio Pinho Navarro; Rodrigo de Castro Bernardes; Ricardo Jayme Procopio; Jose Oyama Leite; Alan Dardik
Journal:  Aorta (Stamford)       Date:  2014-02-01

Review 8.  Endovascular treatment of abdominal aortic aneurysms.

Authors:  Dominique B Buck; Joost A van Herwaarden; Marc L Schermerhorn; Frans L Moll
Journal:  Nat Rev Cardiol       Date:  2013-12-17       Impact factor: 32.419

9.  Hemodynamic impact of abdominal aortic aneurysm stent-graft implantation-induced stenosis.

Authors:  Nicolas Aristokleous; Nikolaos G Kontopodis; Konstantinos Tzirakis; Christos V Ioannou; Yannis Papaharilaou
Journal:  Med Biol Eng Comput       Date:  2015-12-16       Impact factor: 2.602

10.  Long-term outcomes of endovascular aortic aneurysm repair with the Zenith AAA endovascular graft: a single-center study.

Authors:  Hiroshi Kawamata; Hiroyuki Tajima; Tatsuo Ueda; Hidemasa Saito; Daisuke Yasui; Tadashi Kaneshiro; Naoko Takenoshita; Shouhei Mizushima; Takahiko Mine; Jiro Kurita; Yosuke Ishii; Tetsuro Morota; Takashi Nitta; Yuji Maruyama; Hajime Imura; Dai Nishina; Masahiro Fujii; Ryuzo Bessho
Journal:  Jpn J Radiol       Date:  2019-10-29       Impact factor: 2.374

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.