Literature DB >> 23963035

The effect of different bearing surfaces on metal ion levels in urine following 28 mm metal-on-metal and 28 mm metal-on-polyethylene total hip arthroplasty.

H Tiusanen1, K Mäkelä, M Kiilunen, P Sarantsin, E Sipola, M Pesola.   

Abstract

BACKGROUND AND AIMS: Recent advancements in manufacturing technology have enabled more precise tolerances and surface finishes using metal-on-metal bearing surfaces in total hip arthroplasty. The aim of this study was to compare the level of metal ions in urine after implantation of a 28-mm metal-on-metal bearing manufactured from high-carbon wrought alloy and a 28-mm metal-on-polyethylene bearing.
MATERIAL AND METHODS: A total of 92 total hip arthroplasty patients were prospectively randomized into two groups: those receiving metal-on-metal bearings and those receiving metal-on-polyethylene bearings. Chromium, cobalt, and molybdenum ion levels in urine were measured preoperatively and at 1 year and 2 years postoperatively.
RESULTS: In the metal-on-polyethylene group, there was a slight increase in mean chromium and cobalt concentrations at 2-year follow-up compared to the preoperative level (p = 0.02 for both chromium and cobalt). In the metal-on-metal group, there was a 15-fold increase in chromium and a 26-fold increase in cobalt at 2-year follow-up compared to the preoperative level (p < 0.001 for both chromium and cobalt). However, the quantity of chromium and cobalt in urine from the metal-on-metal group was not higher at 2-year follow-up than at 1-year follow-up (p = 0.5 and p = 0.6, respectively).
CONCLUSIONS: The 28-mm metal-on-metal bearings yield chromium and cobalt concentrations in urine that can be higher than those recommended for occupational exposure. However, our results also indicate that a steady state in wear and ion production using metal-on-metal total hip arthroplasty can occur.

Entities:  

Keywords:  Total hip arthroplasty; bearing surfaces; head size; metal ion levels; metal-on-metal; wear

Mesh:

Substances:

Year:  2013        PMID: 23963035     DOI: 10.1177/1457496913491874

Source DB:  PubMed          Journal:  Scand J Surg        ISSN: 1457-4969            Impact factor:   2.360


  2 in total

1.  Differences in concentration of metal debris in blood, serum, and plasma samples of patients with metal-on-metal hip resurfacing arthroplasty.

Authors:  M Khan; J H Kuiper; Christine Sieniawska; J B Richardson
Journal:  J Orthop       Date:  2015-12-09

Review 2.  Increased Mortality in Metal-on-Metal versus Non-Metal-on-Metal Primary Total Hip Arthroplasty at 10 Years and Longer Follow-Up: A Systematic Review and Meta-Analysis.

Authors:  B G Pijls; J M T A Meessen; J W Schoones; M Fiocco; H J L van der Heide; A Sedrakyan; R G H H Nelissen
Journal:  PLoS One       Date:  2016-06-13       Impact factor: 3.240

  2 in total