Sources of error in the measurement of aortic diameter in computed tomography scans.

Juan Parodi1, Ramon Berguer2, Patricia Carrascosa3, Khalil Khanafer4, Carlos Capunay3, Eric Wizauer5.   

Abstract

OBJECTIVE: This study was conducted to determine the differences in the diameter of the thoracic aorta when measured from electrocardiographic (ECG)-gated and nongated computed tomography (CT) angiography. Another aim was to define the difference in the aortic diameter when it is measured at peak systole and end diastole in ECG-gated scans.
METHODS: The gated and nongated CT angiograms of 27 patients (mean age, 58 ± 16 standard deviation [SD] years) obtained on a 256-slice multidetector CT scanner were used. The transverse and anteroposterior diameters and the lumen areas were measured at 1, 4, and 8 cm below the origin of the left subclavian artery.
RESULTS: There was a significant difference in the aortic measurements of diameter between gated and nongated scans found in samples taken at 1, 4, and 8 cm distal to the left subclavian artery (P < .0001). We found a considerable difference between the systolic and diastolic diameters (P < .0001). The maximum change in diameter between systole and diastole was 2.9 ± 0.9 (SD) mm (14.5%, P < .0001) at 1 cm, 5.4 mm (22.6%; median, 1.7 mm; P < .0001) at 4 cm, and 4.4 mm (16.9%; median, 1.3 mm; P < .0001) at 8 cm. There was a significant difference between the transverse and anteroposterior diameters in systole and diastole at all locations (P < .0001): The maximum change in diameter between transverse and anteroposterior diameters in systole was 5.4 ± 1.1 (SD) mm (15.7%, P < .0001) at 1 cm, 5.8 mm (19%; median, 1.4 mm; P < .0001) at 4 cm, and 5 mm (15%; median, 1.02 mm; P < .0001) at 8 cm. There was also a substantial difference between measuring the transverse diameter directly and deriving it from the lumen area (P < .0001).
CONCLUSIONS: Our results showed an important difference between systolic and diastolic diameters measurements in ECG-gated scans. The standard protocol for measuring aortic diameters in gated scans of the thoracic aorta uses images at end diastole because the lack of wall motion at this time provides better resolution. This is likely to result in undersizing that, in some instances, may threaten stability and the proper seal of the stent graft. The dimensions of the aorta in a gated CT should be measured at peak systole rather than the conventional end diastole used today. Most medical centers use nongated CT or gated CT scans in end diastole to calculate sizes of endografts. In view of our findings, the latter method could result in potential complications.
Copyright © 2014 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23958070     DOI: 10.1016/j.jvs.2013.07.005

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  5 in total

1.  A Deep Learning-Based and Fully Automated Pipeline for Thoracic Aorta Geometric Analysis and Planning for Endovascular Repair from Computed Tomography.

Authors:  Simone Saitta; Francesco Sturla; Alessandro Caimi; Alessandra Riva; Maria Chiara Palumbo; Giovanni Nano; Emiliano Votta; Alessandro Della Corte; Mattia Glauber; Dante Chiappino; Massimiliano M Marrocco-Trischitta; Alberto Redaelli
Journal:  J Digit Imaging       Date:  2022-01-26       Impact factor: 4.056

2.  Systolic excursion of the leaflets of the truncal valve: An unusual mechanism for pulmonary stenosis in common arterial trunk.

Authors:  Saurabh Kumar Gupta; Abhinav Aggarwal; Gurpreet S Gulati; Sivasubramanian Ramakrishnan; Shyam S Kothari; Anita Saxena; Sanjiv Sharma; Balram Airan; Robert H Anderson
Journal:  Ann Pediatr Cardiol       Date:  2020-06-29

3.  Dynamic Changes in the Aorta During the Cardiac Cycle Analyzed by ECG-Gated Computed Tomography.

Authors:  Wenying Zhu; Yingliang Wang; Yang Chen; Jiacheng Liu; Chen Zhou; Qin Shi; Songjiang Huang; Chongtu Yang; Tongqiang Li; Bin Xiong
Journal:  Front Cardiovasc Med       Date:  2022-05-19

4.  Normative reference values of thoracic aortic diameter in American College of Radiology Imaging Network (ACRIN 6654) arm of National Lung Screening Trial.

Authors:  Barbara L McComb; Reginald F Munden; Fenghai Duan; Amanda A Jain; Catherine Tuite; Caroline Chiles
Journal:  Clin Imaging       Date:  2016-04-26       Impact factor: 1.605

5.  On the Role and Effects of Uncertainties in Cardiovascular in silico Analyses.

Authors:  Simona Celi; Emanuele Vignali; Katia Capellini; Emanuele Gasparotti
Journal:  Front Med Technol       Date:  2021-12-01
  5 in total