| Literature DB >> 23956842 |
Raul Velez-Montoya1, Chirag Patel, Scott C N Oliver, Hugo Quiroz-Mercado, Naresh Mandava, Jeffrey L Olson.
Abstract
Background. To assess the peeling forces exerted by different calibers of microsurgical forceps on an experimental model of epiretinal membrane. Methods. A model of epiretinal membrane was constructed using thin cellulose paper and heptanes-isopropyl alcohol 1% mixture. The model was mounted on a force censoring device. Subsequently, flaps were created with three different microsurgical forceps of different calibers. We recorded the number of attempts, the duration of the event, and the pushing and the pulling forces during the peeling. The results were compared by a one-way ANOVA and a Fisher unprotected least significant difference test with an alpha value of 0.05 for statistically significance. Results. There was a statistical significant difference on the pulling and pushing forces between the 25 gauge (13.79 mN; -13.27 mN) and the 23 (6.63 mN; -5.76 mN) and 20 (5.02 mN; -5.30 mN) gauge, being greater in the first (P < 0.001). There were no differences in the duration of all events, meaning that all the forces were measured within the same period of time. Conclusions. The 25 gauge microsurgical forceps exerted the greatest mechanical stress over our simulated epiretinal membrane model and required more attempts to create a surgical suitable flap. The clinical implication of this finding is still to be determined.Entities:
Year: 2013 PMID: 23956842 PMCID: PMC3707222 DOI: 10.1155/2013/784172
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Graphic representation of the basic settings used for the experiment. The MSF used were brand new and commercially available.
Figure 2Microphotograph taken with the surgical microscope during one of the attempts to create a flap.
Mean forces and number of attempts.
| Caliber | Duration (seconds) | Pushing (mN) | Pulling (mN) | Number of attemps |
|---|---|---|---|---|
| 20 gauge | 4.63 ± 1.62 | 5.02 ± 2.01 | −5.30 ± 1.51 | 3.7 ± 1.01 |
| 23 gauge | 4.38 ± 1.97 | 6.63 ± 3.96 | −5.76 ± 2.58 | 2.0 ± 0.7 |
| 25 gauge | 3.98 ± 1.93 | 13.79 ± 6.58 | −13.27 ± 5.10 | 7.3 ± 1.58 |
Summary of the duration and mean ± standard deviation of the forces applied during the experiment. mN: millinewtons.
Figure 3Graphic representation of the mean duration and principal vector forces with their corresponding standard error of measurement (SEM).