Predictive parameters of CyberKnife fiducial-less (XSight Lung) applicability for treatment of early non-small cell lung cancer: a single-center experience.

Houda Bahig1, Marie-Pierre Campeau, Toni Vu, Robert Doucet, Dominic Béliveau Nadeau, Bernard Fortin, David Roberge, Louise Lambert, Jean-François Carrier, Edith Filion.   

Abstract

PURPOSE: To determine which parameters allow for CyberKnife fiducial-less tumor tracking in stereotactic body radiation therapy (SBRT) for early-stage non-small cell lung cancer. METHODS AND MATERIALS: A total of 133 lung SBRT patients were preselected for direct soft-tissue tracking based on manufacturer recommendations (peripherally located tumors ≥1.5 cm with a dense appearance) and staff experience. Patients underwent a tumor visualization test to verify adequate detection by the tracking system (orthogonal radiographs). An analysis of potential predictors of successful tumor tracking was conducted looking at: tumor stage, size, histology, tumor projection on the vertebral column or mediastinum, distance to the diaphragm, lung-to-soft tissue ratio, and patient body mass index.
RESULTS: Tumor visualization was satisfactory for 88 patients (66%) and unsatisfactory for 45 patients (34%). Median time to treatment start was 6 days in the success group (range, 2-18 days) and 15 days (range, 3-63 days) in the failure group. A stage T2 (P=.04), larger tumor size (volume of 15.3 cm(3) vs 6.5 cm(3) in success and failure group, respectively) (P<.0001), and higher tumor density (0.86 g/cm(3) vs 0.79 g/cm(3)) were predictive of adequate detection. There was a 63% decrease in failure risk with every 1-cm increase in maximum tumor dimension (relative risk for failure = 0.37, CI=0.23-0.60, P=.001). A diameter of 3.6 cm predicted a success probability of 80%. Histology, lung-to-soft tissue ratio, distance to diaphragm, patient's body mass index, and tumor projection on vertebral column and mediastinum were not found to be predictive of success.
CONCLUSIONS: Tumor size, volume, and density were the most predictive factors of a successful XSight Lung tumor tracking. Tumors >3.5 cm have ≥80% chance of being adequately visualized and therefore should all be considered for direct tumor tracking. Crown
Copyright © 2013. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23953636     DOI: 10.1016/j.ijrobp.2013.06.2048

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  23 in total

1.  Accuracy of Breath-hold CT in Treatment Planning for Lung Stereotactic Ablative Radiotherapy.

Authors:  Dominique Mathieu; Charles Martel; Marie-Pierre Campeau; Édith Filion; David Roberge; Houda Bahig; Toni Vu; Louise Lambert; Karim Boudam; Jean-Francois Carrier
Journal:  Cureus       Date:  2014-12-22

2.  Scatter imaging during lung stereotactic body radiation therapy characterized with phantom studies.

Authors:  Kevin C Jones; Julius Turian; Gage Redler; Gizem Cifter; John Strologas; Alistair Templeton; Damian Bernard; James C H Chu
Journal:  Phys Med Biol       Date:  2020-08-10       Impact factor: 3.609

3.  A dosimetric parameter to limit chest wall toxicity in SABR of NSCLC.

Authors:  Raphaël Jumeau; Édith Filion; Houda Bahig; Toni Vu; Louise Lambert; David Roberge; Robert Doucet; Marie-Pierre Campeau
Journal:  Br J Radiol       Date:  2017-06-07       Impact factor: 3.039

4.  Comparison of CT artifacts and image recognition of various fiducial markers including two types of thinner fiducial markers for CyberKnife treatment.

Authors:  Toshihiro Suzuki; Masahide Saito; Hiroshi Onishi; Koji Mochizuki; Kenichiro Satani; Akihiro Yamazaki; Kenichi Miura; Shinji Taka; Naoki Sano; Takafumi Komiyama; Hiroshi Takahashi
Journal:  Rep Pract Oncol Radiother       Date:  2019-12-09

5.  Real-time markerless tumour tracking with patient-specific deep learning using a personalised data generation strategy: proof of concept by phantom study.

Authors:  Wataru Takahashi; Shota Oshikawa; Shinichiro Mori
Journal:  Br J Radiol       Date:  2020-02-28       Impact factor: 3.039

6.  Cyberknife® stereotactic radiation therapy for stage I lung cancer and pulmonary metastases: evaluation of local control at 24 months.

Authors:  Myriam Khadige; Julia Salleron; Vincent Marchesi; Guillaume Oldrini; Didier Peiffert; Véronique Beckendorf
Journal:  J Thorac Dis       Date:  2018-08       Impact factor: 2.895

7.  Medically inoperable peripheral lung cancer treated with stereotactic body radiation therapy.

Authors:  K D Kelley; D L Benninghoff; J S Stein; J Z Li; R T Byrnes; L Potters; J P S Knisely; H D Zinkin
Journal:  Radiat Oncol       Date:  2015-05-28       Impact factor: 3.481

8.  Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy.

Authors:  Charles A Kunos; Jeffrey M Fabien; John P Shanahan; Christine Collen; Thierry Gevaert; Kenneth Poels; Robbe Van den Begin; Benedikt Engels; Mark De Ridder
Journal:  J Vis Exp       Date:  2015-06-07       Impact factor: 1.355

9.  Markerless Real-Time 3-Dimensional kV Tracking of Lung Tumors During Free Breathing Stereotactic Radiation Therapy.

Authors:  Kimmie de Bruin; Max Dahele; Hassan Mostafavi; Berend J Slotman; Wilko F A R Verbakel
Journal:  Adv Radiat Oncol       Date:  2021-04-20

10.  Magnetic Resonance-Guided Stereotactic Body Radiotherapy of Liver Tumors: Initial Clinical Experience and Patient-Reported Outcomes.

Authors:  Fabian Weykamp; Philipp Hoegen; Sebastian Klüter; C Katharina Spindeldreier; Laila König; Katharina Seidensaal; Sebastian Regnery; Jakob Liermann; Carolin Rippke; Stefan A Koerber; Carolin Buchele; Jürgen Debus; Juliane Hörner-Rieber
Journal:  Front Oncol       Date:  2021-06-09       Impact factor: 6.244

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.