Literature DB >> 23930382

Vacuum-assisted venous drainage, angel or demon: PRO?

Yves Durandy1.   

Abstract

Vacuum-assisted venous drainage (VAVD) was proposed to optimize venous drainage during bypass through femoral venous cannulation. It is currently used in both adult and pediatric surgery when siphon gravity venous drainage is suboptimal. In pediatric surgery, the major advantages of VAVD are a significant decrease in cardiopulmonary bypass prime volume and an improved drainage with all collateral benefits. To limit gravity drainage, we use a two-level heart-lung machine dedicated to pediatric perfusion. The top level of the cardiotomy reservoir is positioned at the patient atrial level, making it possible to downsize the length and diameter of venous and arterial lines. Since 2008, a negative pressure of approximately -30 mmHg has been used for all patients. Initiation of bypass is performed in a classical way with a cardiotomy reservoir open; vacuum is added as soon as the maximal gravity drainage is reached. During bypass, when the blood level in the reservoir decreases to the safety limit level, a small increase in negative pressure is used to improve venous drainage. For weaning from bypass, the negative pressure is gradually decreased to zero, then the reservoir is opened and the venous line progressively closed. Prime volumes were significantly reduced to 100 mL for small neonates, 125 mL for infants, and 175 mL for older children with flow up to 1.5 L/min(-1). A low prime volume is expected to improve blood conservation and decrease donor exposure, prevent drawbacks of transfusion (immunomodulation, infection), increase the incidence of blood-free surgery in smaller babies, and decrease whole body systemic inflammation by decreasing surface of foreign material in contact with blood and inflammation associated with blood transfusion. The main drawbacks described have been retrograde flow in the venous line with cerebral air embolus and an increased incidence of gaseous microemboli. These drawbacks are avoidable through appropriate training of perfusionists. When negative pressure is "reasonable," complications are more theoretical than significant in clinical practice. A technique with a benefit/drawback ratio of 1:0 is utopian, but the advantages of VAVD far outweigh any potential drawbacks when applied properly.

Entities:  

Mesh:

Year:  2013        PMID: 23930382      PMCID: PMC4557578     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  32 in total

1.  Venous air in the bypass circuit: a source of arterial line emboli exacerbated by vacuum-assisted drainage.

Authors:  T W Willcox; S J Mitchell; D F Gorman
Journal:  Ann Thorac Surg       Date:  1999-10       Impact factor: 4.330

2.  Assisted venous drainage presents the risk of undetected air microembolism.

Authors:  A Lapietra; E A Grossi; B B Pua; R A Esposito; A C Galloway; C C Derivaux; L R Glassman; A T Culliford; G H Ribakove; S B Colvin
Journal:  J Thorac Cardiovasc Surg       Date:  2000-11       Impact factor: 5.209

3.  How effective are cardiopulmonary bypass circuits at removing gaseous microemboli?

Authors:  Timothy J Jones; Dwight D Deal; Jason C Vernon; Noel Blackburn; David A Stump
Journal:  J Extra Corpor Technol       Date:  2002-03

4.  Vacuum-assisted venous drainage: to air or not to air, that is the question. Has the bubble burst?

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2002-03

5.  The influence of vacuum assisted drainage on arterial line emboli.

Authors:  Timothy W Willcox; Simon J Mitchell
Journal:  J Extra Corpor Technol       Date:  2002-06

6.  Vacuum-assisted venous return reduces blood usage.

Authors:  Michael K Banbury; Jennifer A White; Eugene H Blackstone; Delos M Cosgrove
Journal:  J Thorac Cardiovasc Surg       Date:  2003-09       Impact factor: 5.209

7.  In vitro assessment of the vacuum-assisted venous drainage (VAVD) system: risks and benefits.

Authors:  Angelo Fiorucci; Pier Silvio Gerometta; Massimo DeVecchi; Cesare Guzman; Maria Laura Costantino; Vincenzo Arena
Journal:  Perfusion       Date:  2004-03       Impact factor: 1.972

8.  Does vacuum-assisted venous drainage increase gaseous microemboli during cardiopulmonary bypass?

Authors:  Timothy J Jones; Dwight D Deal; Jason C Vernon; Noel Blackburn; David A Stump
Journal:  Ann Thorac Surg       Date:  2002-12       Impact factor: 4.330

9.  Vacuum assisted venous drainage does not increase trauma to blood cells.

Authors:  X M Mueller; H T Tevaearai; J Horisberger; M Augstburger; M Burki; L K von Segesser
Journal:  ASAIO J       Date:  2001 Nov-Dec       Impact factor: 2.872

10.  Augmented venous return for minimally invasive open heart surgery with selective caval cannulation.

Authors:  D Jegger; H T Tevaearai; J Horisberger; X M Mueller; Y Boone; N Pierrel; I Seigneul; L K von Segesser
Journal:  Eur J Cardiothorac Surg       Date:  1999-09       Impact factor: 4.191

View more
  1 in total

Review 1.  American Society of ExtraCorporeal Technology: Development of Standards and Guidelines for Pediatric and Congenital Perfusion Practice (2019).

Authors:  Molly E Oldeen; Ronald E Angona; Ashley Hodge; Tom Klein
Journal:  J Extra Corpor Technol       Date:  2020-12
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.