Gantry-angle resolved VMAT pretreatment verification using EPID image prediction.

Henry C Woodruff1, Todsaporn Fuangrod, Pejman Rowshanfarzad, Boyd M C McCurdy, Peter B Greer.   

Abstract

PURPOSE: Pretreatment verification of volumetric modulated arc therapy (VMAT) dose delivery with electronic portal imaging device (EPID) uses images integrated over the entire delivery or over large subarcs. This work aims to develop a new method for gantry-angle-resolved verification of VMAT dose delivery using EPID.
METHODS: An EPID dose prediction model was used to calculate EPID images as a function of gantry angle for eight prostate patient deliveries. EPID image frames at 7.5 frames per second were acquired during delivery via a frame-grabber system. The gantry angle for each image was encoded in kV frames which were synchronized to the MV frames. Gamma analysis results as a function of gantry angle were assessed by integrating the frames over 2° subarcs with an angle-to-agreement tolerance of 0.5° about the measured image angle.
RESULTS: The model agreed with EPID images integrated over the entire delivery with average Gamma pass-rates at 2%, 2 mm of 99.7% (10% threshold). The accuracy of the kV derived gantry angle for each image was found to be 0.1° (1 SD) using a phantom test. For the gantry-resolved analysis all Gamma pass-rates were greater than 90% at 3%, 3 mm criteria (with only two exceptions), and more than 90% had a 95% pass-rate, with an average of 97.3%. The measured gantry angle lagged behind the predicted angle by a mean of 0.3°±0.3°, with a maximum lag of 1.3°.
CONCLUSIONS: The method provides a comprehensive and highly efficient pretreatment verification of VMAT delivery using EPID. Dose delivery accuracy is assessed as a function of gantry angle to ensure accurate treatment.

Entities:  

Mesh:

Year:  2013        PMID: 23927312     DOI: 10.1118/1.4816384

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

Authors:  P Rowshanfarzad; H L Riis; S J Zimmermann; M A Ebert
Journal:  Br J Radiol       Date:  2015-04-23       Impact factor: 3.039

2.  Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy.

Authors:  Aiping Ding; Lei Xing; Bin Han
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

3.  Camera selection for real-time in vivo radiation treatment verification systems using Cherenkov imaging.

Authors:  Jacqueline M Andreozzi; Rongxiao Zhang; Adam K Glaser; Lesley A Jarvis; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

4.  Clinical translation of a new flat-panel detector for beam's-eye-view imaging.

Authors:  T C Harris; J Seco; D Ferguson; M Lehmann; P Huber; M Shi; M Jacobson; I Valencia Lozano; M Myronakis; P Baturin; R Fueglistaller; D Morf; R Berbeco
Journal:  Phys Med Biol       Date:  2020-12-07       Impact factor: 4.174

5.  Pixel response-based EPID dosimetry for patient specific QA.

Authors:  Bin Han; Aiping Ding; Minghui Lu; Lei Xing
Journal:  J Appl Clin Med Phys       Date:  2016-12-15       Impact factor: 2.102

6.  A novel and independent method for time-resolved gantry angle quality assurance for VMAT.

Authors:  Todsaporn Fuangrod; Peter B Greer; Benjamin J Zwan; Michael P Barnes; Joerg Lehmann
Journal:  J Appl Clin Med Phys       Date:  2017-07-13       Impact factor: 2.102

7.  A method for in vivo treatment verification of IMRT and VMAT based on electronic portal imaging device.

Authors:  Jun Zhang; Xiuqing Li; Miaomiao Lu; Qilin Zhang; Xile Zhang; Ruijie Yang; Maria F Chan; Junhai Wen
Journal:  Radiat Oncol       Date:  2021-12-04       Impact factor: 3.481

8.  Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device.

Authors:  Pejman Rowshanfarzad; Peter Häring; Hans L Riis; Sune J Zimmermann; Martin A Ebert
Journal:  Med Devices (Auckl)       Date:  2015-11-05

9.  An EPID-based system for gantry-resolved MLC quality assurance for VMAT.

Authors:  Benjamin J Zwan; Michael P Barnes; Todsaporn Fuangord; Cameron J Stanton; Daryl J O'Connor; Paul J Keall; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

10.  Two-dimensional in vivo dose verification using portal imaging and correlation ratios.

Authors:  Stefano Peca; Derek W Brown
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.