The impact of column connection on band broadening in very high pressure liquid chromatography.

Joseph J Stankovich1, Fabrice Gritti, Paul G Stevenson, Georges Guiochon.   

Abstract

A series of experiments was conducted to evaluate the degree of band broadening in very high pressure LC due to column connections. Different column manufacturers use slightly different designs for their column fittings. If the same column connections are repeatedly used to attach columns of different origins, different void volumes form between capillary tubes and column inlets. An Agilent Ultra Low Dispersion Kit (tubing id 75 μm) was installed on an Agilent Infinity 1290 ultra HPLC and used to connect successively an Agilent, a Phenomenex, and a Waters column. A series of uracil (unretained) samples were injected and eluted at a wide range of flow rates with a water/acetonitrile mixture as eluent. In order to determine the variance contribution from column connections as accurately as possible a nonretained probe compound was selected because the variance contribution from the column is the smallest for analytes, which have very low k values. Yet, this effect still has an impact on the resolution for moderately retained compounds (k > 2) for narrow-bore columns packed with fine particles, since variance contributions are additive for linear chromatographic systems. Each injection was replicated five times under the same experimental conditions. Then NanoViper column connections (tubing id 75 μm) were used and the same injections were made. This system was designed to minimize connection void volumes for any column. Band variances were calculated as the second central moment of elution peaks and used to assess the degree of band broadening due to the column connections. Band broadening may increase from 3.8 to 53.9% when conventional metal ferrules were used to join columns to connection sites. The results show that the variance contribution from improper connections can generate as much as 60.5% of the total variance observed. This demonstrates that column connections can play a larger role than the column packing with respect to band dispersion.
© 2013 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

Entities:  

Keywords:  Band broadening; Column connections; Very high pressure liquid chromatography; Void volumes

Year:  2013        PMID: 23900740     DOI: 10.1002/jssc.201300175

Source DB:  PubMed          Journal:  J Sep Sci        ISSN: 1615-9306            Impact factor:   3.645


  2 in total

1.  Retention and effective diffusion of model metabolites on porous graphitic carbon.

Authors:  Daniel B Lunn; Young J Yun; James W Jorgenson
Journal:  J Chromatogr A       Date:  2017-11-14       Impact factor: 4.759

Review 2.  [Research advances in nano liquid chromatography instrumentation].

Authors:  Sandong Yang; Naijie Li; Zhou Ma; Tao Tang; Tong Li
Journal:  Se Pu       Date:  2021-10
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.