Influence of bag volume on reproducibility of inert gas rebreathing pulmonary blood flow measurements in patients with pulmonary diseases.

Joachim Saur1, Franziska Kraus, Joachim Brade, Dariusch Haghi, Michael Behnes, Ursula Hoffmann, Martin Borggrefe, Jens J Kaden, Frederik Trinkmann.   

Abstract

PURPOSE: Non-invasive inert gas rebreathing (IGR) has shown promising results in the determination of pulmonary blood flow. The volume of the rebreathing bag (V bag) is proposed by the system. However, elderly patients or those with severe pulmonary disease may be unable to rebreathe this volume entirely. We evaluated the effect of adapting V bag on the reproducibility of IGR.
METHODS: A total of 270 valid measurements were obtained from 45 patients with obstruction (group A), restriction (group B), and in healthy controls (group C). Two measurements for each of three different V bag of 1,200, 1,700, and 2,200 ml were conducted in the supine position.
RESULTS: We found no statistically significant difference of the repeated measurements neither between the different V bag in groups A to C nor between the three groups for identical V bag. There was a weak yet significantly worse coefficient of variation between a V bag of 2,200 ml in group A compared with group C with 2,200 and 1,200 ml, respectively. Intraclass correlation coefficient and repeatability coefficient yielded significantly worse values in group A for a V bag of 2,200 ml compared with healthy controls and lower bag volumes. No difference could be found intraclass nor interclass in groups B and C.
CONCLUSIONS: V bag can be altered between 1,200 and 2,200 ml in most situations without affecting the reproducibility. Attention has to be paid to extreme volumes in obstructive patients. Nevertheless, V bag should be chosen as large as possible and therefore has to be carefully adapted, particularly in patients with obstruction or restriction.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23884622     DOI: 10.1007/s00408-013-9494-0

Source DB:  PubMed          Journal:  Lung        ISSN: 0341-2040            Impact factor:   2.584


  21 in total

Review 1.  Cardiovascular magnetic resonance.

Authors:  D Pennell
Journal:  Heart       Date:  2001-05       Impact factor: 5.994

Review 2.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

3.  Comparison of left ventricular ejection fraction and volumes in heart failure by echocardiography, radionuclide ventriculography and cardiovascular magnetic resonance; are they interchangeable?

Authors:  N G Bellenger; M I Burgess; S G Ray; A Lahiri; A J Coats; J G Cleland; D J Pennell
Journal:  Eur Heart J       Date:  2000-08       Impact factor: 29.983

Review 4.  Clinical use of cardiovascular magnetic resonance.

Authors:  Gerald M Pohost; Lynne Hung; Mark Doyle
Journal:  Circulation       Date:  2003-08-12       Impact factor: 29.690

5.  Noninvasive measurement of cardiac output during exercise by inert gas rebreathing technique: a new tool for heart failure evaluation.

Authors:  Piergiuseppe Agostoni; Gaia Cattadori; Anna Apostolo; Mauro Contini; Pietro Palermo; Giancarlo Marenzi; Karlman Wasserman
Journal:  J Am Coll Cardiol       Date:  2005-10-10       Impact factor: 24.094

6.  The impact of pulmonary disease on noninvasive measurement of cardiac output by the inert gas rebreathing method.

Authors:  Joachim Saur; Frederik Trinkmann; Christina Doesch; Armin Scherhag; Joachim Brade; Stefan O Schoenberg; Martin Borggrefe; Jens J Kaden; Theano Papavassiliu
Journal:  Lung       Date:  2010-07-31       Impact factor: 2.584

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  A critical discussion of intraclass correlation coefficients.

Authors:  R Müller; P Büttner
Journal:  Stat Med       Date:  1994 Dec 15-30       Impact factor: 2.373

9.  A modified photo- and magnetoacoustic multigas analyzer applied in gas exchange measurements.

Authors:  P Clemensen; P Christensen; P Norsk; J Grønlund
Journal:  J Appl Physiol (1985)       Date:  1994-06

10.  Noninvasive determination of cardiac output by the inert-gas-rebreathing method--comparison with cardiovascular magnetic resonance imaging.

Authors:  Joachim Saur; Stephan Fluechter; Frederik Trinkmann; Theano Papavassiliu; Stefan Schoenberg; Joerg Weissmann; Dariusch Haghi; Martin Borggrefe; Jens J Kaden
Journal:  Cardiology       Date:  2009-08-07       Impact factor: 1.869

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.