Literature DB >> 23879524

A clinically relevant accuracy study of computer-planned implant placement in the edentulous maxilla using mucosa-supported surgical templates.

Luc M Verhamme1, Gert J Meijer, Tiny Boumans, Anton F J de Haan, Stefaan J Bergé, Thomas J J Maal.   

Abstract

PURPOSE: The purpose of the study is to determine the clinically relevant accuracy of implant placement in the edentulous maxilla using computer planning and a mucosa-supported surgical template.
MATERIALS AND METHODS: In each of in total 30 consecutive edentulous patients suffering from retention problems of their upper denture, two or four Brånemark MkIII Groovy (Nobel Biocare®, Zürich, Switzerland) implants in the maxilla were installed. Preoperatively, first, a cone-beam computer tomography (cone beam computer tomography) scan was acquired, followed by virtual implant planning. Hereafter, a surgical template was designed to allow flapless implant placement using the template as a guide. To inventory the accuracy of implant placement, a postoperative CBCT scan was obtained and matched to the preoperative scan. The accuracy of implant placement was validated three-dimensionally. The Implant Position Orthogonal Projection validation method was applied to measure the clinically relevant implant deviations (i.e., in both the bucco-lingual and mesio-distal plane). Also, the influence of type of surgery, use of fixation pins, and position on the dental arch were investigated with regard to implant deviations.
RESULTS: In total, 104 implants were installed. In bucco-lingual direction, a mean implant deviation of 0.67 mm was scored at the implant tip, of 0.51 mm at the shoulder, of -0.83 mm in depth, as also a mean deviation of angulation of 1.74°. In mesio-distal direction, a mean implant deviation of 0.75 mm was found at the implant tip, of 0.60 mm at the implant shoulder, of -0.75 mm in depth, and a deviation of angulation of 1.94°. Of all implants, 74% was placed not deep enough compared with the planning. Implant position on the dental arch, the use of fixation pins, and type of surgery showed no significant effect on implant deviations. However, a significant difference for implant deviations in both buccal and mesial direction was observed, explained by a nonoptimal positioning of the surgical template.
CONCLUSIONS: Computer-aided implant planning showed to be a clinically relevant tool for the placement of two or four implants in the maxilla of fully edentulous patients. Exact positioning of the surgical template in anterior/posterior direction is crucial in reducing implant deviations both in buccal and mesial direction.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  cone-beam CT; implant placement; implantology; oral and maxillofacial surgery; validation; validation method; virtual implant planning

Mesh:

Substances:

Year:  2013        PMID: 23879524     DOI: 10.1111/cid.12112

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  8 in total

1.  [Research advances in the use of digital surgical guides in implantology].

Authors:  Xiao-Hua Wang; Ai-Peng Liu; Wen-Zheng Deng
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-02-01

2.  Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study.

Authors:  Du-Hyeong Lee; Seo-Young An; Min-Ho Hong; Kyoung-Bae Jeon; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2016-06-17       Impact factor: 1.904

Review 3.  Role of three-dimensional printing in periodontal regeneration and repair: Literature review.

Authors:  Meisha Gul; Aysha Arif; Robia Ghafoor
Journal:  J Indian Soc Periodontol       Date:  2019 Nov-Dec

4.  Immediate Loading of Implants Placed by Guided Surgery in Geriatric Edentulous Mandible Patients.

Authors:  Eugenio Velasco-Ortega; Alvaro Jiménez-Guerra; Ivan Ortiz-Garcia; Jesús Moreno-Muñoz; Enrique Núñez-Márquez; Daniel Cabanillas-Balsera; José López-López; Loreto Monsalve-Guil
Journal:  Int J Environ Res Public Health       Date:  2021-04-13       Impact factor: 3.390

Review 5.  Accuracy of Computer-Assisted Flapless Implant Placement by Means of Mucosa-Supported Templates in Complete-Arch Restorations: A Systematic Review.

Authors:  Paolo Carosi; Claudia Lorenzi; Fabrizio Lio; Pierluigi Cardelli; Alessandro Pinto; Andrea Laureti; Alessandro Pozzi
Journal:  Materials (Basel)       Date:  2022-02-16       Impact factor: 3.623

Review 6.  3D Printed and Bioprinted Membranes and Scaffolds for the Periodontal Tissue Regeneration: A Narrative Review.

Authors:  Irina-Georgeta Sufaru; Georgiana Macovei; Simona Stoleriu; Maria-Alexandra Martu; Ionut Luchian; Diana-Cristala Kappenberg-Nitescu; Sorina Mihaela Solomon
Journal:  Membranes (Basel)       Date:  2022-09-19

7.  The accuracy of a 3D printing surgical guide determined by CBCT and model analysis.

Authors:  Boyoung Ma; Taeseok Park; Inkon Chun; Kwidug Yun
Journal:  J Adv Prosthodont       Date:  2018-08-17       Impact factor: 1.904

8.  Accuracy of implant surgical guides fabricated using computer numerical control milling for edentulous jaws: a pilot clinical trial.

Authors:  Jinyou Chai; Xiaoqian Liu; Ramona Schweyen; Jürgen Setz; Shaoxia Pan; Jianzhang Liu; Yongsheng Zhou
Journal:  BMC Oral Health       Date:  2020-10-21       Impact factor: 2.757

  8 in total