Evaluation of a new recombinant antigen-based Virotech Treponema pallidum screen ELISA for diagnosis of syphilis.

Clemens Busse1, Mojdeh Heidary Navid, Andreas Strubel, Paul Schnitzler.   

Abstract

BACKGROUND: The purpose of this study was to evaluate the diagnostic performance of the new Sekisui Virotech Treponema pallidum Screen ELISA in comparison with the currently used treponemal tests, TPPA and FTA-abs, and to compare this new ELISA to the FDA approved Phoenix Trep-Sure ELISA.
METHODS: To establish the sensitivity and specificity of the Virotech Screen, 421 serum samples from different panels of infected and noninfected patients, sera from seronegative pregnant women as well as international syphilis standard sera and panels were tested.
RESULTS: In comparison to combined TPPA/FTA-abs, Phoenix Trep-Sure and Virotech Screen demonstrated a sensitivity of 100% and a specificity of 93.9% and 98.3%, respectively. All samples of a well defined syphilis serum panel were correctly identified by the Virotech test, whereas the Phoenix test identified two Treponema negative samples as equivocal. Results of both ELISAs highly correlated with TPPA negative and positive samples. The analytical sensitivity of the Virotech Screen with international standards 05/122 and 05/132 was determined at 0.02 IU/mL and 0.03 IU/mL, respectively, and was slightly superior to the Phoenix Trep-Sure.
CONCLUSIONS: The Virotech Screen ELISA demonstrated good diagnostic sensitivity and specificity when evaluated as a screening test for syphilis among various patient populations, including samples with increased rates of false-positive nontreponemal test results. Thus, the new Virotech ELISA may be used in automatic analysers as an alternative to the manual TPPA. However, the use of a confirmatory test remains a must in order to avoid false-positive results.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23865350     DOI: 10.7754/clin.lab.2012.120519

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  5 in total

1.  Analytical performance of three diagnostic reagents for HBsAg on an automatic ELISA analyzer.

Authors:  Jiayi Hou; Jianping Ren; Ling Song; Fen Zhao; Peijun Liang
Journal:  J Clin Lab Anal       Date:  2017-02-10       Impact factor: 2.352

2.  Performance of Treponemal Tests for the Diagnosis of Syphilis.

Authors:  Ina U Park; Yetunde F Fakile; Joan M Chow; Kathleen J Gustafson; Heather Jost; Jeffrey M Schapiro; Susan Novak-Weekley; Anthony Tran; Jim H Nomura; Victor Chen; Manie Beheshti; Townson Tsai; Karen Hoover; Gail Bolan
Journal:  Clin Infect Dis       Date:  2019-03-05       Impact factor: 9.079

3.  Performance of the Syphilis Health Check in Clinic and Laboratory-Based Settings.

Authors:  Yetunde F Fakile; Myra Brinson; Victoria Mobley; Ina U Park; Anne M Gaynor
Journal:  Sex Transm Dis       Date:  2019-04       Impact factor: 2.830

4.  Syphilis and a pregnant woman: a real danger for the woman and the child.

Authors:  Katarzyna Plagens-Rotman; Renata Przybylska; Katarzyna Gerke; Małgorzata Piskorz-Szymendera; Magdalena Tomaszewska; Anna Sadowska-Przytocka; Zygmunt Adamski; Magdalena Czarnecka-Operacz
Journal:  Postepy Dermatol Alergol       Date:  2019-02-22       Impact factor: 1.837

5.  Assessment of recombinant antigens Tp0100 and Tp1016 of Treponema pallidum for serological diagnosis of syphilis.

Authors:  Dejun Chen; Siqian Wang; Yuxing He; Yue Fu; Feijun Zhao; Xiuping Zhou; Haoquan Yin; Jia Wan; Yunting Huang; Yimou Wu; Longgu Cao; Tiebing Zeng
Journal:  J Clin Lab Anal       Date:  2022-07-31       Impact factor: 3.124

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.