Literature DB >> 23842028

The effects of loading on the preload and dimensions of the abutment screw for a 3-unit cantilever-fixed prosthesis design.

Gaurav Setia1, Hoda Yousef, David Ehrenberg, Allyn Luke, Saul Weiner.   

Abstract

OBJECTIVE: The purpose of this study was to use an in vitro model system to compare the effects on the screw torque and screw dimensions within 2 commercially available implant systems from occlusal loading on a cantilevered-fixed partial denture.
MATERIALS AND METHODS: Cantilevered implant-supported 3-unit prostheses with 2 premolar abutments and 1 premolar pontic (7.3 mm in length) were made on resin casts containing 2 implant analogs for 2 implant systems: BioLok Silhouette Tapered Implant System (Birmingham, AL) and Zimmer Tapered Screw-Vent Implant System (Carlsbad, CA) with 10 samples in each group. Each sample was loaded with either of 2 protocols: (1) a load of 50 N on the cantilevered pontic unit and (2) a loading of 150 N on all 3 units. The outcome measures were (1) changes in residual torque of the abutment screws and (2) changes in screw dimension.
RESULTS: The BioLok Silhouette Tapered Implant group demonstrated slight but statistically significant torque loss 18.8% to 28.5% in both abutment screws for both protocols, P ≤ 0.05, without any changes in screw dimension. In the Zimmer Tapered Screw-Vent Implant group, there was a significant elongation of the abutment screws and a markedly significant 44.4%, (P ≤ 0.01) loss in torque in the mesial screw and a 28.5%, (P ≤ 0.05) loss in torque in the distal screw when the cantilever alone was loaded.
CONCLUSIONS: Differences in screw design influence the maintenance of preload and distortion of the shank. The influence of the interface design, namely an internal hex of 1 mm versus an external hex did not influence the preload. Cantilevered prostheses can cause loss of torque and dimensional changes in abutment screws.

Mesh:

Year:  2013        PMID: 23842028     DOI: 10.1097/ID.0b013e31829c227a

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  1 in total

1.  Effect of lateral oblique cyclic loading on microleakage and screw loosening of implants with different connections.

Authors:  Hakimeh Siadat; Hossain Najafi; Marzieh Alikhasi; Babak Falahi; Elaheh Beyabanaki; Farid Zayeri
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2018-09-18
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.