Abstract

OBJECTIVE: To optimize image-guided prostate biopsy by minimizing the target error with trocar-sharpened needle tips instead of beveled needles, which constantly deviate away from the bevel.
MATERIALS AND METHODS: We performed stereotactic biopsies on two prostate phantoms, which incorporate three randomly placed TRUS-visible lesions. Four stereotactic biopsies per lesion were taken under live-ultrasound guidance through a template: two biopsies with conventional beveled needles and two biopsies with novel trocar-sharpened needles. The procedural targeting error (PTE) between the virtually planned biopsy trajectory and the manually registered 3D needle position of every single biopsy core taken was calculated.
RESULTS: The absolute overall targeting error using the novel needle-tip design was 0.13 mm (SD: ± 0.15 mm) with the highest PTE in the sagittal plane (0.18 ± 0.16 mm), followed by the coronal (0.13 ± 0.17 mm) and axial (0.09 ± 0.05 mm) planes. Comparing the PTE of the novel trocar-shaped needles with conventional beveled needles, there was a statistically significant difference in the axial plane [p (overall) = 0.47, p(axial) = 0.03].
CONCLUSION: The targeting error of stereotactic biopsies using trocar-sharpened needles is significantly lower than the targeting error of classical beveled needles. Thus, trocar-tip configurations improve the accuracy of computer-assisted biopsies and allow precise assessment of suspicious lesions in the prostate and in other organs accessible to image-guided biopsy.
Copyright © 2013 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2013        PMID: 23838372     DOI: 10.1159/000350653

Source DB:  PubMed          Journal:  Urol Int        ISSN: 0042-1138            Impact factor:   2.089


  1 in total

1.  Deflection Analysis of Different Needle Designs for Prostate Biopsy and Focal Therapy.

Authors:  Nelson N Stone; Vladimir Mouraviev; David Schechter; Josh Goetz; M Scott Lucia; E Erin Smith; E David Crawford