Literature DB >> 23827787

Multiplex loop-mediated isothermal amplification (M-LAMP) assay for the detection of influenza A/H1, A/H3 and influenza B can provide a specimen-to-result diagnosis in 40 min with single genome copy sensitivity.

James Mahony1, Sylvia Chong, David Bulir, Alexandra Ruyter, Ken Mwawasi, Daniel Waltho.   

Abstract

BACKGROUND: Rapid isothermal amplification methods have recently been introduced and they offer significant advantages over PCR.
OBJECTIVE: To develop a rapid and sensitive M-LAMP assay for the detection of influenza A (H1 and H3) and B that does not require RNA extraction. STUDY
DESIGN: We designed six primers targeting the matrix genes of influenza H1 and H3 and the NS1 gene of influenza B and developed a M-LAMP assay using a commercially available Master Mix and a real time fluorometer (Genie II, Optigene, UK) that displays real time amplification, time to positivity and amplicon annealing temperature (Tm). M-LAMP was evaluated against PCR by testing 202 nasopharyngeal (NP) specimens.
RESULTS: Optimized M-LAMP was rapid with a mean amplification time of 12 min (compared with 90-120 min for PCR), had an analytical sensitivity of 1 genome equivalent (ge), and could distinguish influenza A including subtypes A/H1 and A/H3 from influenza B by Tm. M-LAMP detected 26/28 influenza A/H1, 27/27 influenza A/H3 and 39/39 influenza B specimens and had a combined sensitivity and specificity for detecting influenza (A and B) of 97.9% (92/94) and 100% (108/108), respectively. The rapid amplification time of LAMP coupled with a novel 10-min specimen preparation procedure consisting of vortexing and heating in M-Swab diluent (Copan Italia) provided a rapid result.
CONCLUSIONS: M-LAMP had excellent sensitivity and specificity for detecting influenza A and B in NP specimens and when used together with a rapid specimen processing method provided a specimen-to-result diagnosis in 30 min.
Copyright © 2013 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Influenza; LAMP; Rapid detection

Mesh:

Substances:

Year:  2013        PMID: 23827787     DOI: 10.1016/j.jcv.2013.06.006

Source DB:  PubMed          Journal:  J Clin Virol        ISSN: 1386-6532            Impact factor:   3.168


  30 in total

Review 1.  Miniaturized devices for point of care molecular detection of HIV.

Authors:  Michael Mauk; Jinzhao Song; Haim H Bau; Robert Gross; Frederic D Bushman; Ronald G Collman; Changchun Liu
Journal:  Lab Chip       Date:  2017-01-31       Impact factor: 6.799

2.  Initial Evaluation of a Mobile SARS-CoV-2 RT-LAMP Testing Strategy.

Authors:  Christina M Newman; Mitchell D Ramuta; Matthew T McLaughlin; Roger W Wiseman; Julie A Karl; Dawn M Dudley; Miranda R Stauss; Robert J Maddox; Andrea M Weiler; Mason I Bliss; Katrina N Fauser; Luis A Haddock; Cecilia G Shortreed; Amelia K Haj; Molly A Accola; Anna S Heffron; Hailey E Bussan; Matthew R Reynolds; Olivia E Harwood; Ryan V Moriarty; Laurel M Stewart; Chelsea M Crooks; Trent M Prall; Emma K Neumann; Elizabeth D Somsen; Corrie B Burmeister; Kristi L Hall; William M Rehrauer; Thomas C Friedrich; Shelby L O'Connor; David H O'Connor
Journal:  J Biomol Tech       Date:  2021-09

3.  Evaluation of the Alere I influenza A&B nucleic acid amplification test by use of respiratory specimens collected in viral transport medium.

Authors:  J Jeremiah Bell; Rangaraj Selvarangan
Journal:  J Clin Microbiol       Date:  2014-09-10       Impact factor: 5.948

4.  Simple Identification of Human Taenia Species by Multiplex Loop-Mediated Isothermal Amplification in Combination with Dot Enzyme-Linked Immunosorbent Assay.

Authors:  Agathe Nkouawa; Yasuhito Sako; Munehiro Okamoto; Akira Ito
Journal:  Am J Trop Med Hyg       Date:  2016-04-04       Impact factor: 2.345

5.  A Rapid On-Site Assay for the Detection of Influenza A by Capillary Convective PCR.

Authors:  Zhihao Zhuo; Jin Wang; Wendi Chen; Xiaosong Su; Mengyuan Chen; Mujin Fang; Shuizhen He; Shiyin Zhang; Shengxiang Ge; Jun Zhang; Ningshao Xia
Journal:  Mol Diagn Ther       Date:  2018-04       Impact factor: 4.074

6.  Real-time sequence-validated loop-mediated isothermal amplification assays for detection of Middle East respiratory syndrome coronavirus (MERS-CoV).

Authors:  Sanchita Bhadra; Yu Sherry Jiang; Mia R Kumar; Reed F Johnson; Lisa E Hensley; Andrew D Ellington
Journal:  PLoS One       Date:  2015-04-09       Impact factor: 3.240

7.  Multiplex Reverse-Transcription Loop-Mediated Isothermal Amplification Coupled with Cascade Invasive Reaction and Nanoparticle Hybridization for Subtyping of Influenza A Virus.

Authors:  Ying Chi; Yiyue Ge; Kangchen Zhao; Bingjie Zou; Bin Liu; Xian Qi; Qian Bian; Zhiyang Shi; Fengcai Zhu; Minghao Zhou; Lunbiao Cui; Chuan Su
Journal:  Sci Rep       Date:  2017-03-21       Impact factor: 4.379

8.  Closed tube loop-mediated isothermal amplification assay for rapid detection of hepatitis B virus in human blood.

Authors:  Nguyen Bao Quoc; Nguyen Doan Nguyen Phuong; Nguyen Ngoc Bao Chau; Do Thi Phuong Linh
Journal:  Heliyon       Date:  2018-03-06

9.  Detection of Middle East respiratory syndrome coronavirus using reverse transcription loop-mediated isothermal amplification (RT-LAMP).

Authors:  Kazuya Shirato; Takuya Yano; Syouhei Senba; Shigehiro Akachi; Takashi Kobayashi; Takamichi Nishinaka; Tsugunori Notomi; Shutoku Matsuyama
Journal:  Virol J       Date:  2014-08-08       Impact factor: 4.099

10.  Detection of influenza viruses by coupling multiplex reverse-transcription loop-mediated isothermal amplification with cascade invasive reaction using nanoparticles as a sensor.

Authors:  Yiyue Ge; Qiang Zhou; Kangchen Zhao; Ying Chi; Bin Liu; Xiaoyan Min; Zhiyang Shi; Bingjie Zou; Lunbiao Cui
Journal:  Int J Nanomedicine       Date:  2017-04-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.