A method for online verification of adapted fields using an independent dose monitor.

Jina Chang1, Robert K Heaton, Robert Mahon, Bernhard D Norrlinger, David A Jaffray, Young-Bin Cho, Mohammad K Islam.   

Abstract

PURPOSE: Clinical implementation of online adaptive radiotherapy requires generation of modified fields and a method of dosimetric verification in a short time. We present a method of treatment field modification to account for patient setup error, and an online method of verification using an independent monitoring system.
METHODS: The fields are modified by translating each multileaf collimator (MLC) defined aperture in the direction of the patient setup error, and magnifying to account for distance variation to the marked isocentre. A modified version of a previously reported online beam monitoring system, the integral quality monitoring (IQM) system, was investigated for validation of adapted fields. The system consists of a large area ion-chamber with a spatial gradient in electrode separation to provide a spatially sensitive signal for each beam segment, mounted below the MLC, and a calculation algorithm to predict the signal. IMRT plans of ten prostate patients have been modified in response to six randomly chosen setup errors in three orthogonal directions.
RESULTS: A total of approximately 49 beams for the modified fields were verified by the IQM system, of which 97% of measured IQM signal agree with the predicted value to within 2%.
CONCLUSIONS: The modified IQM system was found to be suitable for online verification of adapted treatment fields.

Entities:  

Mesh:

Year:  2013        PMID: 23822448     DOI: 10.1118/1.4811204

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Real-time beam monitoring for error detection in IMRT plans and impact on dose-volume histograms : A multi-center study.

Authors:  Livia Marrazzo; Chiara Arilli; Marlies Pasler; Martijn Kusters; Richard Canters; Luca Fedeli; Silvia Calusi; Marta Casati; Cinzia Talamonti; Gabriele Simontacchi; Lorenzo Livi; Stefania Pallotta
Journal:  Strahlenther Onkol       Date:  2017-12-18       Impact factor: 3.621

2.  X-Ray Beam Segment Size and Entrance Location Effects on the Integral Quality Monitor (IQM®) Signal and Usefulness in Predicting Complex Segment Output Signals.

Authors:  Mahuvava C; Du Plessis F C P
Journal:  J Biomed Phys Eng       Date:  2020-08-01

3.  Sensitivity Analysis of the Integral Quality Monitoring System® Using Monte Carlo Simulation.

Authors:  Oluwaseyi M Oderinde; F C P du Plessis
Journal:  Comput Math Methods Med       Date:  2017-08-27       Impact factor: 2.238

4.  Characterization and evaluation of an integrated quality monitoring system for online quality assurance of external beam radiation therapy.

Authors:  David Hoffman; Eunah Chung; Clayton Hess; Robin Stern; Stanley Benedict
Journal:  J Appl Clin Med Phys       Date:  2016-11-29       Impact factor: 2.102

5.  Two-dimensional solid-state array detectors: A technique for in vivo dose verification in a variable effective area.

Authors:  Kananan Utitsarn; Giordano Biasi; Nauljun Stansook; Ziyad A Alrowaili; Marco Petasecca; Martin Carolan; Vladimir L Perevertaylo; Wolfgang A Tomé; Tomas Kron; Michael L F Lerch; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2019-10-14       Impact factor: 2.102

6.  Diode-based transmission detector for IMRT delivery monitoring: a validation study.

Authors:  Taoran Li; Q Jackie Wu; Thomas Matzen; Fang-Fang Yin; Jennifer C O'Daniel
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.