Literature DB >> 23818460

Biomechanical comparison of 4 different lateral plate constructs for distal fibula fractures.

Tobin T Eckel1, Richard R Glisson, Prashanth Anand, Selene G Parekh.   

Abstract

BACKGROUND: Displaced lateral malleolar fractures are often treated with reduction and surgical stabilization. However, there has not been a comprehensive laboratory comparison to determine the most appropriate device for treating these patients. This study subjected a range of contemporary lateral fibular plates to a series of mechanical tests designed to reveal performance differences.
METHODS: Forty fresh frozen lower extremities were divided into 4 groups. A Weber B distal fibula fracture was simulated with an osteotomy and stabilized using 1 of 4 plate systems: a standard Synthes one-third tubular plate with interfragmentary lag screw, a Synthes LCP locking plate with lag screw, an Orthohelix MaxLock Extreme low-profile locking plate with lag screw, or a TriMed Sidewinder nonlocking plate. Controlled monotonic bending and cyclic torsional loading were applied and bending stiffness, torsional stiffness, and fracture site motion were quantified. Resistance to cyclic torsional loading was determined by quantifying the number of loads withstood before excessive rotation occurred. Correlation between bone mineral density and each of the mechanical measures was determined.
RESULTS: There was no difference in angulation or bending stiffness between plates. All plates except the LCP showed greater lateral deflection than in the other bending directions. Bending stiffness was lowest in lateral distal fragment deflection for all 4 plates. There was a positive correlation between bone mineral density and bending stiffness for all plate types. There was no difference in fracture site rotation between plate types in internal or external torsion, but internal rotation of the distal fragment consistently exceeded external rotation. Torsional stiffness in external rotation exceeded stiffness in internal rotation in nearly all specimens. LCP plates performed relatively poorly under cyclic torsion.
CONCLUSIONS: Significant differences in plate performance were not demonstrated. The effects of bone quality variability and differences in interfragmentary screw purchase resulted in data dispersion that confounded absolute ranking of plate performance. CLINICAL RELEVANCE: Identification of an optimal lateral fibular plating system has the potential to improve the clinical outcome of malleolar fracture fixation, particularly when patient conditions are unfavorable.

Entities:  

Keywords:  biomechanical comparison; fibula fracture; lateral plate constructs; open reduction internal fixation; trauma

Mesh:

Year:  2013        PMID: 23818460     DOI: 10.1177/1071100713496223

Source DB:  PubMed          Journal:  Foot Ankle Int        ISSN: 1071-1007            Impact factor:   2.827


  6 in total

1.  Distal fibula oblique fracture fixation using one-third tubular plate with and without lag screw - A biomechanical study of stability.

Authors:  Gopikanthan Manoharan; Rohit Singh; Jan Herman Kuiper; Leonard Derek Martin Nokes
Journal:  J Orthop       Date:  2018-05-07

2.  Is early full weight bearing safe following locking plate ORIF of distal fibula fractures?

Authors:  Michael Zyskowski; Markus Wurm; Frederik Greve; Sebastian Pesch; Francesca von Matthey; Patrick Pflüger; Moritz Crönlein; Peter Biberthaler; Chlodwig Kirchhoff
Journal:  BMC Musculoskelet Disord       Date:  2021-02-09       Impact factor: 2.362

Review 3.  Fibula fractures management.

Authors:  Gianluca Canton; Andrea Sborgia; Guido Maritan; Roberto Fattori; Federico Roman; Marko Tomic; Massimo Max Morandi; Luigi Murena
Journal:  World J Orthop       Date:  2021-05-18

4.  Stacked 1/3 Tubular Plates for Fixation of Pediatric Forearm Fractures: A Biomechanical Study.

Authors:  Benjamin J Cooper; Alexander C Wendling; Sharon R Isaacs-Pullins; Joel White; Steven M Hollenbeck
Journal:  Kans J Med       Date:  2022-02-09

5.  One-Third Tubular Plate Remains a Clinically Good Option in Danis-Weber Type B Distal Fibular Fracture Fixation.

Authors:  Jae Hoon Ahn; Sung Hyun Cho; Mingi Jeong; Yoon-Chung Kim
Journal:  Orthop Surg       Date:  2021-10-27       Impact factor: 2.071

6.  Tubular vs Profile Plate in Peroneal or Bimalleolar Fractures: is There a Real Difference in Skin Complication? A Retrospective Study in Three Level I Trauma Center.

Authors:  Rosario Petruccelli; Michele Bisaccia; Giuseppe Rinonapoli; Giuseppe Rollo; Luigi Meccariello; Gabriele Falzarano; Paolo Ceccarini; Olga Bisaccia; Marco Giaracuni; Auro Caraffa
Journal:  Med Arch       Date:  2017-08
  6 in total