| Literature DB >> 23811915 |
Jianfeng Wang1, Mads Sylvest Bergholt, Wei Zheng, Zhiwei Huang.
Abstract
We report on the development of a beveled fiber-optic confocal Raman probe coupled with a ball lens for enhancing in vivo epithelial tissue Raman measurements at endoscopy. Our Monte Carlo simulations show that by selecting a proper fiber-ball lens distance and beveled angle of collection fibers, the confocal Raman probe design can be optimized for maximizing shallower tissue Raman measurements in epithelial tissue; in addition, the ratio of epithelium to stromal Raman photons collected using an optimized confocal Raman probe is approximately 19-fold higher than that using a volume-type Raman probe. Further experiments confirm that the confocal Raman endoscopic probe developed is in favor of probing superficial tissue Raman signals from a two-layer tissue phantom as well as esophagus tissue in vivo during endoscopy. This work suggests the great potential of applying the beveled fiber-optic confocal Raman probe for improving in vivo diagnosis of precancer occurring in epithelial tissue at endoscopy.Mesh:
Year: 2013 PMID: 23811915 DOI: 10.1364/OL.38.002321