Jianfeng Wang1, Kan Lin1, Wei Zheng1, Khek Yu Ho2, Ming Teh3, Khay Guan Yeoh2, Zhiwei Huang1. 1. Optical Bioimaging Laboratory, Department of Biomedical Engineering, Faculty of Engineering, National University of Singapore, Singapore 117576. 2. Department of Medicine, Yong Loo Lin School of Medicine, National University of Singapore and National University Health System, Singapore 119260. 3. Department of Pathology, Yong Loo Lin School of Medicine, National University of Singapore and National University Health System, Singapore 119074.
Abstract
This work aims to evaluate clinical value of a fiber-optic Raman spectroscopy technique developed for in vivo diagnosis of esophageal squamous cell carcinoma (ESCC) during clinical endoscopy. We have developed a rapid fiber-optic Raman endoscopic system capable of simultaneously acquiring both fingerprint (FP)(800-1800 cm(-1)) and high-wavenumber (HW)(2800-3600 cm(-1)) Raman spectra from esophageal tissue in vivo. A total of 1172 in vivo FP/HW Raman spectra were acquired from 48 esophageal patients undergoing endoscopic examination. The total Raman dataset was split into two parts: 80% for training; while 20% for testing. Partial least squares-discriminant analysis (PLS-DA) and leave-one patient-out, cross validation (LOPCV) were implemented on training dataset to develop diagnostic algorithms for tissue classification. PLS-DA-LOPCV shows that simultaneous FP/HW Raman spectroscopy on training dataset provides a diagnostic sensitivity of 97.0% and specificity of 97.4% for ESCC classification. Further, the diagnostic algorithm applied to the independent testing dataset based on simultaneous FP/HW Raman technique gives a predictive diagnostic sensitivity of 92.7% and specificity of 93.6% for ESCC identification, which is superior to either FP or HW Raman technique alone. This work demonstrates that the simultaneous FP/HW fiber-optic Raman spectroscopy technique improves real-time in vivo diagnosis of esophageal neoplasia at endoscopy.
This work aims to evaluate clinical value of a fiber-optic Raman spectroscopy technique developed for in vivo diagnosis of esophageal squamous cell carcinoma (ESCC) during clinical endoscopy. We have developed a rapid fiber-optic Raman endoscopic system capable of simultaneously acquiring both fingerprint (FP)(800-1800 cm(-1)) and high-wavenumber (HW)(2800-3600 cm(-1)) Raman spectra from esophageal tissue in vivo. A total of 1172 in vivo FP/HW Raman spectra were acquired from 48 esophagealpatients undergoing endoscopic examination. The total Raman dataset was split into two parts: 80% for training; while 20% for testing. Partial least squares-discriminant analysis (PLS-DA) and leave-one patient-out, cross validation (LOPCV) were implemented on training dataset to develop diagnostic algorithms for tissue classification. PLS-DA-LOPCV shows that simultaneous FP/HW Raman spectroscopy on training dataset provides a diagnostic sensitivity of 97.0% and specificity of 97.4% for ESCC classification. Further, the diagnostic algorithm applied to the independent testing dataset based on simultaneous FP/HW Raman technique gives a predictive diagnostic sensitivity of 92.7% and specificity of 93.6% for ESCC identification, which is superior to either FP or HW Raman technique alone. This work demonstrates that the simultaneous FP/HW fiber-optic Raman spectroscopy technique improves real-time in vivo diagnosis of esophageal neoplasia at endoscopy.
Esophageal cancer is the eighth most frequent malignancy worldwide with a 5-year survival
rate of approximately 19% in the Unites States12. Esophageal squamous
cell carcinoma (ESCC) is one of the deadly esophageal cancers3. Current
routine diagnosis of ESCC is based on white light reflectance (WLR) endoscopy which
heavily relies on visual identification of gross morphological tissue changes, leading
to poor diagnosis accuracy123. Existing diagnostic guidelines
recommend extensive biopsy samplings (typically four-quadrant samplings) at every
1–2 cm interval along suspicious lesions segments during
endoscopic inspections of patients4. This approach produces a vast number
of negative biopsies without much improving the diagnostic yield, but is clinically
labor intensive and a burden to the patients. There is a need to develop advanced
optical diagnostic techniques for objective diagnosis and characterization of esophageal
tissue with biomolecular specificity.Optical spectroscopic techniques (e.g., fluorescence, diffuse reflectance, and Raman)
have been comprehensively investigated for precancer and cancer diagnosis in internal
organs (e.g., stomach, esophagus, colon, bladder, and lung, etc.)5678910111213141516171819202122232425. Raman spectroscopy is an optical vibrational technique based on the fundamental
principle of inelastic light scattering26, which is capable of probing
biochemical and biomolecular structures and conformations of cells and tissue associated
with disease transformation78910111213141516171819202122232425.
Currently, biomedical Raman research in diagnosing precancer and cancer is mostly
centered on the fingerprint (FP) range (i.e.,
800–1800 cm−1) that contains rich
biochemical information about the tissue789101112131415161718. The unrivaled advantage of the
FP Raman spectroscopy technique stems from its capability to uncover specific
information about backbone structures of proteins, lipids and nucleic acid assemblies in
cells and tissue789101112131415161718. The
efficiency of the FP Raman spectroscopy technique is, however, compromised in certain
organ sites owing to extremely weak tissue Raman signals but overwhelming tissue
autofluorescence (AF) background. Recent attentions have been directed towards the use
of high-wavenumber (HW) regime (e.g.,
2800–3600 cm−1), as the HW
spectral range exhibits stronger tissue Raman signals with less AF interference19202122232425. The clinical rationales for combining the FP
and HW spectral ranges for in vivo esohphageal Raman measurements are therefore
manifold: (i) For tissues that could exhibit intense AF overwhelming the tissue FP Raman
signals, the HW range could still contain intense tissue Raman peaks with diagnostic
information, (ii) The FP and HW Raman spectra offer complementary biomolecular
information, and the integrated FP/HW Raman technique could improve tissue
characterization and diagnosis2425. To date, the simultaneous FP and HW
Raman spectroscopic technique has not yet been investigated for distinguishing ESCC from
normal esophagus. This work aims to assess the clinical value of a simultaneous FP and
HW fiber-optic Raman spectroscopy technique developed for improving real-time in
vivo diagnosis of ESCC at endoscopy. Unpaired two-sided Student’s
t-test is employed to extract the statistically different Raman spectral
features between ESCC and normal esophageal tissues. Partial least squares -
discriminant analysis (PLS-DA) and leave-one patient-out, cross validation (LOPCV) are
further implemented to develop robust diagnostic algorithms for differentiation between
ESCC and normal esophagus.
Results
A total of 1172 in vivo FP/HW tissue Raman spectra (normal
(n = 860); ESCC (n = 312)) were
acquired from 48 esophagealpatients undergoing routine endoscopic examination (see
Supplementary Fig. S1 online). The
total in vivo Raman dataset acquired was split into two parts: i.e., 80% of
the total dataset for training (938 in vivo FP/HW Raman spectra [normal
(n = 736); ESCC (n = 202)] from 34
esophagealpatients); while the remaining 20% of the total dataset for predictive
testing (234 in vivo FP/HW Raman spectra [normal
(n = 124); ESCC (n = 110)] from 14
esophagealpatients). Figure 1(a) shows the mean in
vivo FP/HW tissue Raman
spectra ± 1standard deviation (SD) (shaded area)
of the training dataset (80% of the total dataset) for tissue diagnostic algorithms
development. The corresponding images of the WLR-guided FP/HW Raman procedures are
also shown in Fig. 1. Prominent esophageal tissue Raman peaks
with tentative assignments789101112131415161718
can be observed in the FP region, i.e., 853 (v(C-C) proteins), 1004
(νs(C-C) ring breathing of phenylalanine), 1078
(ν(C-C) of lipids), 1265 (amide III v(C-N) and
δ(N-H) of proteins), 1302 (CH2 twisting and wagging of
lipids), 1335 (CH3CH2 twisting of proteins and nucleic acids),
1445 (δ(CH2) deformation of proteins and lipids), 1618
(v(C = C) of porphyrins), 1655 (amide I
v(C = O) of proteins) and
1745 cm−1
(v(C = O) of phospholipids). Intense Raman peaks are also
observed in the HW region1819202122232425, i.e., 2850
and 2885 cm−1 (symmetric and asymmetric
CH2 stretching of lipids),
2940 cm−1 (CH3 stretching of
proteins), ~3300 cm−1 (amide A
(NH stretching of proteins)) and the broad Raman band of water (OH stretching
vibrations peaking at ~3250 and
~3400 cm−1) that are related
to the local conformation and interactions of OH-bonds in the intracellular and
extracellular space of esophageal tissue2728. The intense broad
Raman band of water above 3000 cm−1 has also
been observed in other soft tissues (e.g., brain, oral tissue)2728.
Figure 1(b) shows the difference Raman spectra between
ESCC and normal esophageal tissue ± 1SD (shaded
area), reflecting the Raman-active component changes associated with cancerous
progression in the esophagus. The significant difference
(p = 1.3E-8, unpaired two-sided Student’s
t-test) in Raman spectra of ESCC and normal tissue discerned (Fig.
1(b)) demonstrates the potential of simultaneous FP/HW Raman endoscopy
for in vivo diagnosis of esophageal cancer.
Figure 1
(a) The mean in vivo FP/HW Raman
spectra ± 1 standard deviation (SD) of
the training dataset (80% of the total dataset) (normal
(n = 736); ESCC (n = 202))
for diagnostic algorithms development; (b) Difference spectra (ESCC -
normal) ± 1 SD resolving the unique
spectral features of ESCC. The corresponding images of the WLR-guided FP/HW
Raman procedures on normal esophagus and ESCC are also shown.
To elucidate the diagnostically important Raman-active components, Fig. 2(a) shows a logarithmic plot of the calculated p-values
(unpaired two-sided Student’s t-test) for each of the Raman intensities
in the entire spectral range (i.e.,
800–1800 cm−1 and
2800–3600 cm−1). In
particular, seven spectral sub-regions with statically significant difference
(p < 1E-10) between ESCC and normal esophagus were
found: i.e., 840–940 cm−1,
1025–1100 cm−1,
1310–1355 cm−1,
1585–1690 cm−1, and
2830–2975 cm−1 related to
proteins, lipids and nucleic acids. Significant spectral differences were also
observed in bound water in the ranges of
3160–3260 cm−1 and
3370–3420 cm−1. Figure 2(b) displays a histogram of the most statistically
different Raman peak intensities (mean ± 1SD)
for both the FP and HW ranges, i.e., (i)
853 cm−1, (ii)
1078 cm−1, (iii)
1335 cm−1, (iv)
1618 cm−1, (v)
1655 cm−1, (vi)
2850 cm−1, (vii)
2885 cm−1, (viii)
3250 cm−1, and (ix)
3400 cm−1. The histopathology identifies
prominent cellular and architectural anomalies in ESCC (Fig.
3), while the relatively higher or lower FP/HW tissue Raman bands
representing different Raman-active components reveal the specific
biochemical/biomolecular changes of esophageal tissue accompanied with ESCC
transformation. The changes of FP/HW Raman spectra related to lipids, proteins, DNA
and water contents in tissue reconfirm the capability of simultaneous FP/HW Raman
spectroscopy to detect ESCC at the molecular level.
Figure 2
(a) Unpaired two-sided Student’s t-test on Raman peak
intensities of the training dataset (80% of the total dataset) (normal
(n = 736); ESCC (n = 202))
over the entire spectral range (i.e.,
800–1800 cm−1 and
2800–3600 cm−1).
Seven Raman spectra sub-regions containing the diagnostically significant
information were identified. (b)
Histogram ± 1 SD of the most
diagnostically significant Raman peaks
(*p < 1E-10).
Figure 3
Representative hematoxylin and eosin (H&E)-stained histopathologic
slides (original magnification, ×200) corresponding to different
esophageal tissue types measured:
(a) Normal superficial keratinised squamous epithelium and the basal
layer; (b) Invasive esophageal squamous cell carcinoma showing
prominent architectural and cytological atypia.
Capitalizing on the complementary biochemical/biomolecular information identified in
both the FP and HW spectral ranges, PLS-DA and LOPCV are implemented on the training
dataset (80% of the total dataset) (Fig. 1(a)) to develop
robust diagnostic model for enhancing in vivo ESCC diagnosis. A
Cohen’s kappa of 0.91 demonstrated a high level of agreement between the
independent pathologists for the esophageal tissue groupings29. Figure 4 shows the scattered plots of cross-validated PLS-DA
posterior probability of each Raman prediction for (a) FP, (b) HW, and (c)
integrated FP/HW, respectively. The diagnostic accuracy with integrated FP/HW Raman
spectroscopy is 97.3% [sensitivity of 97.0% (196/202) and specificity of 97.4%
(717/736)], superior to using either FP (accuracy of 90.9%, sensitivity of 93.6%
(189/202), and specificity of 90.2% (664/736)) or HW (accuracy of 85.5%, sensitivity
of 78.2%(158/202), and specificity of 87.5% (644/736)) Raman technique alone. The
receiver operating characteristic (ROC) curves are also generated (Fig. 5), with the integration areas under the ROC curves of being 0.972,
0.928 and 0.995, respectively, for the FP, HW and the integrated FP/HW techniques.
The above results confirm that the integrated FP/HW Raman technique provides the
best diagnostic performance for in vivo ESCC detection as compared to FP or
HW Raman technique.
Figure 4
The posterior probabilities of in vivo Raman spectra belonging to (i)
normal esophagus (n = 736), and (ii) ESCC
(n = 202) of the training dataset (80% of the total
dataset), using partial lease square-discriminant analysis and leave-one-patient out cross validation based on the FP, HW and the integrated FP/HW Raman
techniques, respectively. (◯) normal; (▲) ESCC.
Figure 5
Receiver operating characteristic (ROC) curves for separating ESCC from
normal esophageal tissue for the training dataset (80% of the total
dataset).
The areas under the ROC curves (AUC) are 0.972, 0.928 and 0.995,
respectively, using the FP, HW and the integrated FP/HW Raman
techniques.
In the light of the promising diagnostic results presented (Figs
4 and 5), we have applied the simultaneous FP/HW
Raman spectroscopy and diagnostic algorithms developed for predictive diagnosis of
the independent testing dataset (20% of the total dataset)(see Supplementary Fig. S2 online). The predictive
accuracy of 93.2% (i.e., sensitivity of 92.7% (102/110) and specificity of 93.6%
(116/124)) can be achieved by using the integrated FP/HW Raman spectroscopy,
substantiating the advantages of simultaneous FP/HW Raman spectroscopy over FP
(predictive accuracy of 91.0%; sensitivity of 90.9% (100/110), and specificity of
91.9% (113/124)) or HW (predictive accuracy of 80.3%; sensitivity of 76.4% (84/110),
and specificity of 83.9% (104/124)) Raman technique for in vivo ESCC
diagnosis (see Supplementary Fig. S3
online).
Discussion
We have successfully acquired high-quality in vivo FP/HW Raman spectra from
esophagus tissue in real-time (<1 sec) during clinical endoscopic
examination. Significant FP and HW Raman spectral differences between ESCC and
normal esophageal tissue are observed. Statistically significant different
sub-regions of FP Raman bands (840–940, 1025–1100,
1310–1355, and
1585–1690 cm−1) were
identified, uncovering highly specific biochemical changes of esophageal tissue
associated with ESCC transformation. For instance, ESCC was characterised by
significantly reduced Raman peak intensities at
1078 cm−1
(v(C = C)) (p = 2.3E-59), indicating
a relative reduction in lipid content that is likely due to the thickening of the
cancerous esophageal mucosa obscuring the Raman signals from deeper tissue
layers30. The increase of proteins content was indicated by the
sensitive biomarkers at 1618 cm−1 and the
band broadening of 1655 cm−1 (amide I
v(C = O) of proteins) for ESCC. This observation is in
agreement with the molecular biology study whereby an over-expression of various
proteins associated with ESCC has been reported31. Moreover, ESCC
shows an increased DNA content exhibiting much stronger Raman intensity at
1335 cm−1 (CH3CH2
twisting of proteins and nucleic acids), which is complied with the DNA changes
associated with precancer and cancer transformation7891011121314151617182425. On the other hand,
unpaired two-sided Student’s t-test identified three statistically
different sub-regions of HW Raman bands (2830-2975, 3160-3260 and
3370–3420 cm−1) related to
lipids, proteins and water changes associated with ESCC transformation. ESCC was
characterised by a relatively decreased lipid and increased protein content as
indicated by the higher Raman peak intensity ratio of
I2940/I2850in ESCC. The result is consistent with the
changes of lipids and proteins content observed in other clinical studies of Raman
spectroscopy for disease diagnosis78910111213141516171819202122232425. Different water content (OH stretching vibrations peaking at ~3250
and ~3400 cm−1) was observed
between ESCC and normal esophagus as indicated by the significantly different ratio
(p = 3.2E-7) of asymmetric to symmetric OH stretching
peaking at ~3250 and
~3400 cm−1, reflecting the
local structure of hydrogen-bonded networks in the epithelium32. The
relative increase of the OH stretching vibrational intensity suggests that
neoplastic epithelium is accompanied with an increased content of water in tissue
associated with ESCC, which is probably due to the increased cellularity and
proliferation of ESCC in the esophagus3233. Three Raman peak
intensity ratios (i.e., I2885/I2940,
I2940/I3250, and I2885/I3250) were
also calculated, suggesting that ESCC could be separated from normal esophageal
tissue with sensitivities of 48.5% (98/202), 57.9% (117/202), 73.3% (148/202), and
specificities of 54.6% (402/736), 64.1% (472/736), 64.3% (473/736), respectively.
These findings indicate that the water perfusion in ESCC is one of the important
biomarkers for esophageal neoplasia detection in vivo.To employ the Raman-based biochemical/biomolecular information for ESCC diagnosis, we
further developed a robust PLS-DA classification model constructed from the training
dataset (80% of the total dataset). The PLS-DA algorithms based on the integrated
FP/HW Raman spectra could differentiate ESCC from normal esophagus mucosa with a
good diagnostic accuracy (accuracy of 97.3% (913/938)). The FP Raman spectrum
contains highly specific information about proteins, lipids and DNA conformations.
On the other hand, the HW spectral range contains complementary information related
to local conformation of bound and unbound water as well as CH2 and
CH3 stretching moieties of lipids and proteins that are not reflected
in the FP range, enabling back-tracking of misclassified Raman spectra. For
instance, FP and HW Raman spectroscopy misclassified 85 and 136 spectra,
respectively, from the 11 and 23 patients recruited, in which 9 misclassified
patients were identical in both the modalities. The integrated FP/HW Raman technique
reduced the number of misclassified patients to 4 (25 spectra). Hence, the
complementary biochemical/biomolecular information harvested by the simultaneous
FP/HW Raman spectroscopy could greatly improve the in vivo diagnosis of ESCC
at endoscopy. Furthermore, misclassified spectra by the FP Raman (that were
correctly classified by the integrated FP/HW Raman) are comprised of weaker tissue
Raman signals and relatively high AF background; while the addition of the HW Raman
containing stronger tissue Raman peaks (e.g., 2885, 2940, 3250 and
3400 cm−1) but much reduced AF could
improve the overall signal-to-noise ratio (SNR) of integrated FP/HW Raman spectra,
therefore, enhancing the performance of classification. In addition, prospective
study on the independent testing dataset (20% of the total dataset) (see Supplementary Figs. S2-S3 online) shows
that the integrated FP/HW Raman spectroscopy could provide a predication accuracy of
93.2% (218/234) for ESCC identification. The above results affirm that the
integrated FP/HW Raman spectroscopy could be a potent optical diagnostics means to
revealing comprehensive but complementary endogenous biomolecular information of
epithelial tissue, thereby enhancing in vivo objective diagnosis of
esophageal neoplasia.The introduction of FP/HW Raman spectroscopy could have a major impact on current
endoscopic practice and decision making by providing objective real-time diagnosis.
Several clinical scenarios exist for the applications of FP/HW Raman spectroscopy
during clinical endoscopic examinations. One scenario is for the screening and
surveillance. FP/HW Raman spectroscopy can be used by the endoscopists as a guiding
tool to reduce negative biopsy numbers by objectively evaluating wide lesion regions
in real-time without extensive excisional biopsies of the tissues, reducing the risk
and the cost of the procedure. Further, the individual patients having higher risk
of developing ESCC could be identified by Raman spectroscopy for quick surveillance
and examination at endoscopy. Therefore, FP/HW Raman spectroscopy is expected to
provide a greater opportunity for curative treatment with improved patient outcome.
In another scenario of interventional procedures, FP/HW Raman spectroscopy could be
used to assist in determining the lesion margins during endoscopic mucosal resection
(EMR) and endoscopic submucosal dissection (ESD) procedures for the treatment of
ESCC3435. The FP/HW Raman spectroscopy system developed is now
routinely used by both experienced and novice endoscopists without any difficulty at
Endoscope Center at the National University Health System, Singapore for screening
esophagealpatients for targeted biopsies. Considering its encouraging diagnostic
performance (Fig. 4(c)), fiber-optic FP/HW Raman spectroscopy
has promise to become one part of routine endoscopic practices. Currently, clinical
trials on a larger series of esophagealpatients based on the simultaneous FP/HW
Raman spectroscopy technique are in progress to further assess its clinical merits
for precancer and early cancer diagnosis in the esophagus.In summary, we demonstrate that the simultaneous FP/HW in vivo Raman spectra
of esophageal tissue can be acquired in real-time during clinical endoscopic
examination. Significant FP/HW Raman spectral differences between ESCC and normal
esophageal tissue are observed. The use of complementary biomolecular information
harvested through the integrated FP/HW Raman spectroscopy significantly enhances the
detection and diagnosis of esophageal neoplasia as compared to either the FP or HW
Raman technique alone. This unique biomolecular endoscopic approach based on the
simultaenous FP/HW Raman spectroscopy could open a new avenue for objective
diagnosis of esophageal neoplasia in vivo at the molecular level.
Methods
Raman Instrumentation
The Raman fiber-optic simultaneous FP and HW spectroscopy platform developed for
in vivo tissue Raman measurements at endoscopy consists of a
785 nm diode laser (maximum output: 300 mW, B&W
TEK Inc.), a high-throughput reflective imaging spectrograph (Acton
LS-785 f/2, Princeton Instruments Inc.) equipped with a customised
830 gr/mm gold-coated grating that has a ~85%
diffraction efficiency in the NIR range of 800–1200 nm.
A thermo electric-cooled (−70 °C),
NIR-optimised deep-depletion charge-coupled device (CCD) camera (PIXIS
400BR-eXcelon, Princeton Instruments Inc.) was used to collect the FP/HW Raman
signals. A customised parabolic-aligned fiber bundle
(64 × 100 μm fibers,
numerical aperture (NA) = 0.22) was coupled into the
entrance slit to compensate for the image aberration in the broad spectral range
(800–3600 cm−1)1736, which significantly improves the signal-to-noise ratio
(SNR) (20-fold improvement) as well as the spectral resolution
(~9 cm−1) of the Raman
system as compared to a conventional straight slit imaging spectrograph. A
depth-selective bevelled fiber-optic Raman endoscopic probe (1.8 mm
in outer diameter) compatible with most medical endoscopes was developed for
in vivo tissue Raman measurements at endoscopy. The customised Raman
endoscopic probe comprises central light delivery fiber
(200 μmin in diameter,
NA = 0.22) surrounded by 18 bevelled collection fibers
(200 μm in diameter, NA = 0.22).
A 1.0 mm sapphire ball lens (NA = 1.78) is
further coupled to the fiber tip of the probe to focus the excitation light onto
the tissue surface37. Monte Carlo simulation shows that
~85% of the total Raman signal collected by the bevelled fiber-optic
Raman endoscopic probe is arising from the top
~300 μm esophageal layer with an estimated
tissue probing volume of ~0.017 mm3;
whereas ~15% of the Raman collection is from the stromal layer
(~300-800) μm of the esophageal tissue with an estimated
tissue probing volume of ~0.003 mm3,
facilitating the in vivo detection of esophageal early cancer and
precancer37.The 785-nm laser excitation power is
~12 mW on the tissue, which is less than the American
National Standards Institute (ANSI) maximum permissible skin exposure limit set
out for a 785-nm laser beam38. Further finite difference thermal
modeling3940 based on the optical properties of esophageal
tissue4142 shows that even without consideration of other
cooling effects (e.g., perfusion and evaporation in tissue), the maximum tissue
temperature rise is only about 0.07 °C after 1-min of
785-nm laser radiation with an incident power density of
~1.5 W/cm2 during tissue Raman
measurements. This temperature rise estimated is far below the level to cause
photothermal damage to tissue and cells43, suggesting that the
laser power used in this study is safe for in vivo tissue Raman
measurements. The system was wavelength-calibrated using a mercury/argon
calibration lamp (HG-1 and AR-1, Ocean Optics Inc., Dunedin, FL) in the FP
range, and 4-acetamidophenol that exhibits strong well-defined peaks in the HW
region at 2931 cm−1 and
3064 cm−1 (ASTM E1840 standard) is
used for the HW wavelength calibration. The spectral response correction for the
wavelength-calibrated system was conducted using a standard lamp (RS-10,
EG&G Gamma Scientific, San Diego, CA)44. The raw FP/HW
Raman spectra measured from in vivo esophageal tissue represent a
combination of weak tissue Raman scattering, intense AF background and the
noise. The raw spectra are preprocessed by a third-order Savitzky-Golay
smoothing filter (a window width of 3 pixels) to remove the spectral noise. In
the FP region
(800–1800 cm−1), a
fifth-order polynomial was found to be optimal for fitting the AF background in
the noise-smoothed spectrum and this polynomial was then subtracted from the
measured FP spectrum to yield the FP tissue Raman spectrum alone. In the HW
range (2800–3600 cm−1), a
first-order polynomial fit was used for removing the AF background. The tissue
FP/HW Raman spectrum was finally normalised to the integrated area of the entire
FP and HW range, enabling a better comparison of the spectral shapes and
relative Raman band intensities between ESCC and normal esophagus. All the
preprocessing is completed within 0.5 sec, and the processed Raman
spectra and diagnostic outcomes can be displayed on the computer screen in
real-time by using the house-developed online Raman software44.
Clinical trial protocol
The ethical protocol of the present study was approved by the Institutional
Review Board (IRB) of the National Healthcare Group (NHG) of Singapore. Prior to
Raman measurements, all patients (21 to 80 years old) signed an informed consent
permitting in vivo Raman spectroscopic measurements during endoscope
examinations. The trial has been registered (registration number:
ISRCTN15587241) on June 23, 2015 into ISRCTN registry (http://www.isrctn.com/), which is a
publicly accessible primary register that participates in the World Health
Organization (WHO) International Clinical Trial Registry Platform. Also, the
trial was performed in accordance with International Conference on Harmonization
(ICH) for Good Clinical Practice (GCP) guidelines, Declaration of Helsinki
(2000). A total of 48 esophagealpatients (31 men and 17 women with mean ages of
51) were enrolled in the gastrointestinal Raman endoscopic examinations for
various esophageal indications (e.g., anemia, bleeding, eating problems, pain
etc.). One notes that the patients presenting with co-morbid diseases, severe
acute/chronic medical conditions or bleeding disorders, in which biopsies may be
contraindicated were excluded from this study. During endoscopic examination of
suspicious lesions, the Raman probe was placed in gentle contact with the
esophageal mucosa surface, and the positioning of the Raman probe against the
tissue sites was verified on the computer monitor by the endoscopists in-charge
at endoscopy. Multiple spectra (∼8-10) for each tissue site were
measured with scanning times of 0.1 to 0.5 sec, which permits a
rapid survey of large tissue areas. Raman spectra acquired in non-contact with
esophageal tissues were automatically discarded online using Raman diagnostic
software developed by our group44. A total of 1172 FP/HW
esophageal Raman spectra (normal (n = 938) and ESCC
(n = 234)) were successfully acquired in vivo
from 101 tissue sites. Immediately after the tissue Raman acquisitions, each
tissue site measured was biopsied and sent for histopathological examination by
three senior gastrointestinal pathologists who were blinded to the Raman scans.
The histopathology assessments serve as the gold standard to determine the
diagnostic performance of the simultaneous FP/HW fiber-optic Raman technique for
identifying ESCC from normal esophageal tissues.
Statistical Analysis
Cohen’s κ statistics were calculated to examine the
agreement among the pathologists29. The unpaired two-sided
Student’s t-test was used to evaluate Raman spectral differences
between ESCC and normal esophagus15. Mean-centering was performed
prior to multivariate statistical analysis to remove common variance from the
dataset of in vivo esophageal tissue Raman spectra. Partial least squares
(PLS)- discriminant analysis (DA) was further applied for tissue diagnosis model
development45. PLS-DA employs the fundamental principle of
PCA but further rotates the components (latent variables (LVs)) by maximizing
the covariance between the spectral variation and group affinity, so that the
LVs explain the diagnostic relevant variations rather than the most prominent
variations in the spectral dataset45. Leave-one-patient out,
cross-validation was used to assess and optimize the PLS-DA model complexity,
while reducing the risk of overfitting45. The overall
discriminatory accuracy of the FP, HW, and integrated FP/HW diagnostic models
were further evaluated through the use of the receiver operating characteristic
(ROC) analysis4546. The PLS-DA model developed was further
applied to an independent validation dataset to evaluate the predictive
diagnostic performance of the integrated FP/HW Raman spectroscopy technique for
ESCC identification. The above multivariate statistical analysis was performed
using in-house written scripts in the Matlab programming environment.
Additional Information
How to cite this article: Wang, J. et al. Simultaneous fingerprint and
high-wavenumber fiber-optic Raman spectroscopy improves in vivo diagnosis of
esophageal squamous cell carcinoma at endoscopy. Sci. Rep.
5, 12957; doi: 10.1038/srep12957 (2015).
Authors: Jason T Motz; Martin Hunter; Luis H Galindo; Joseph A Gardecki; John R Kramer; Ramachandra R Dasari; Michael S Feld Journal: Appl Opt Date: 2004-01-20 Impact factor: 1.980
Authors: Zhiwei Huang; Annette McWilliams; Harvey Lui; David I McLean; Stephen Lam; Haishan Zeng Journal: Int J Cancer Date: 2003-12-20 Impact factor: 7.396
Authors: Diana Lim; Eric S Renteria; Drake S Sime; Young Min Ju; Ji Hyun Kim; Tracy Criswell; Thomas D Shupe; Anthony Atala; Frank C Marini; Metin N Gurcan; Shay Soker; Joshua Hunsberger; James J Yoo Journal: Biodes Manuf Date: 2021-07-19