Literature DB >> 2380749

Oxygen analyzers: failure rates and life spans of galvanic cells.

R M Meyer1.   

Abstract

Competing technologies exist for measuring oxygen concentrations in breathing circuits. Over a 4-year period, two types of oxygen analyzers were studied prospectively in routine clinical use to determine the incidence and nature of malfunctions. Newer AC-powered galvanic analyzers (North American Dräger O2med) were compared with older, battery-powered polarographic analyzers (Ohmeda 201) by recording all failures and necessary repairs. The AC-powered galvanic analyzer had a significantly lower incidence of failures (0.12 +/- 0.04 failures per machine-month) than the battery-powered polarographic analyzer (4.0 +/- 0.3 failures per machine-month). Disposable capsules containing the active galvanic cells lasted 12 +/- 7 months. Although the galvanic analyzers tended to remain out of service longer, awaiting the arrival of costly parts, the polarographic analyzers were more expensive to keep operating when calculations included the cost of time spent on repairs. Stocking galvanic capsules would have decreased the amount of time the galvanic analyzers were out of service, while increasing costs. In conclusion, galvanic oxygen analyzers appear capable of delivering more reliable service at a lower overall cost. By keeping the galvanic capsules exposed to room air during periods of storage, it should be possible to prolong their life span, further decreasing the cost of using them. In addition, recognizing the aberrations in their performance that warn of the exhaustion of the galvanic cells should permit timely recording and minimize downtime.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2380749     DOI: 10.1007/bf02832146

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  10 in total

1.  Standards for patient monitoring during anesthesia at Harvard Medical School.

Authors:  J H Eichhorn; J B Cooper; D J Cullen; W R Maier; J H Philip; R G Seeman
Journal:  JAMA       Date:  1986 Aug 22-29       Impact factor: 56.272

2.  A random survey of anesthesia machines and ancillary monitors in 45 hospitals.

Authors:  V Kumar; M S Hintze; A M Jacob
Journal:  Anesth Analg       Date:  1988-07       Impact factor: 5.108

3.  Test of a fuel cell oxygen analyzer.

Authors:  T A Torda; G C Grant
Journal:  Br J Anaesth       Date:  1972-10       Impact factor: 9.166

4.  Oxygen analyzer dangerous--senses nitrous oxide as battery fails.

Authors:  S Piernan; M F Roizen; J W Severinghaus
Journal:  Anesthesiology       Date:  1979-02       Impact factor: 7.892

5.  Electrodes for the measurement of oxygen and carbon dioxide tensions.

Authors:  A C Smith; C E Hahn
Journal:  Br J Anaesth       Date:  1969-09       Impact factor: 9.166

6.  Evaluation of oxygen monitors for use during anesthesia.

Authors:  D R Westenskow; W S Jordan; R Jordan; S T Gillmor
Journal:  Anesth Analg       Date:  1981-01       Impact factor: 5.108

7.  Monitoring of respiratory function during anesthesia.

Authors:  I P Sergejev
Journal:  Int Anesthesiol Clin       Date:  1981

8.  Oxygen analyzers in anaesthesia: performance in a simulated clinical environment.

Authors:  J P Bengtson; H Sonander; O Stenqvist
Journal:  Acta Anaesthesiol Scand       Date:  1986-11       Impact factor: 2.105

9.  An evaluation of fourteen oxygen analysers for use in patient breathing circuits.

Authors:  A H Ilsley; W B Runciman
Journal:  Anaesth Intensive Care       Date:  1986-11       Impact factor: 1.669

10.  Oxygen analysers. An evaluation of five fuel cell models.

Authors:  P G Roe; C K Tyler; R Tennant; P K Barnes
Journal:  Anaesthesia       Date:  1987-02       Impact factor: 6.955

  10 in total
  2 in total

Review 1.  Open-circuit respirometry: a historical review of portable gas analysis systems.

Authors:  Duncan J Macfarlane
Journal:  Eur J Appl Physiol       Date:  2017-10-17       Impact factor: 3.078

Review 2.  Analysis of oxygen, anaesthesia agent and flows in anaesthesia machine.

Authors:  Rakesh Garg; Ramesh Chand Gupta
Journal:  Indian J Anaesth       Date:  2013-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.