Competing technologies exist for measuring oxygen concentrations in breathing circuits. Over a 4-year period, two types of oxygen analyzers were studied prospectively in routine clinical use to determine the incidence and nature of malfunctions. Newer AC-powered galvanic analyzers (North American Dräger O2med) were compared with older, battery-powered polarographic analyzers (Ohmeda 201) by recording all failures and necessary repairs. The AC-powered galvanic analyzer had a significantly lower incidence of failures (0.12 +/- 0.04 failures per machine-month) than the battery-powered polarographic analyzer (4.0 +/- 0.3 failures per machine-month). Disposable capsules containing the active galvanic cells lasted 12 +/- 7 months. Although the galvanic analyzers tended to remain out of service longer, awaiting the arrival of costly parts, the polarographic analyzers were more expensive to keep operating when calculations included the cost of time spent on repairs. Stocking galvanic capsules would have decreased the amount of time the galvanic analyzers were out of service, while increasing costs. In conclusion, galvanic oxygen analyzers appear capable of delivering more reliable service at a lower overall cost. By keeping the galvanic capsules exposed to room air during periods of storage, it should be possible to prolong their life span, further decreasing the cost of using them. In addition, recognizing the aberrations in their performance that warn of the exhaustion of the galvanic cells should permit timely recording and minimize downtime.Entities: