Assessment of changes in left ventricular systolic function with oesophageal Doppler.

X Monnet1, J-M Robert, M Jozwiak, C Richard, J-L Teboul.   

Abstract

BACKGROUND: We tested the ability of mean acceleration (Acc) and peak velocity (V peak) of the aortic velocity signal measured by oesophageal Doppler to reflect left ventricular (LV) systolic performance.
METHODS: We included critically ill patients in whom a fluid challenge (n=25) or the introduction of dobutamine, 5 µg kg(-1) min(-1) (n=25), was planned by the attending physician. Before and after therapeutic interventions, we measured Acc and V peak (CardioQ device) and LV ejection fraction (LVEF) using echocardiography.
RESULTS: For all pairs of measurements, the absolute values of Acc and V peak correlated with LVEF (r=0.36 and 0.57, respectively). The correlation was significantly higher for V peak than for Acc. Volume expansion did not significantly change LVEF and Acc, but significantly increased V peak by 7 (8)%. Dobutamine increased LVEF by 30 (15)%, Acc by 33 (25)%, and V peak by 20 (10)%. Considering the pooled effects of volume expansion and dobutamine, changes in Acc and V peak and those of LVEF were correlated (r=0.53 and 0.67, respectively). When excluding changes <18% (i.e. the least significant change for LVEF), the concordance rate was 96% for Acc and 100% for V peak.
CONCLUSIONS: V peak and, to a lesser extent, Acc measured by oesophageal Doppler behaved as markers of LV systolic performance as they were almost insensitive to fluid administration and changed to a much larger extent with dobutamine. These indices could be used to estimate LV systolic performance and to assess the effects of inotropic therapy.

Entities:  

Keywords:  cardiac output; cardiac output, shock; measurement, equipment; monitors, dobutamine, measurement techniques

Mesh:

Substances:

Year:  2013        PMID: 23798745     DOI: 10.1093/bja/aet212

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  4 in total

Review 1.  The use of the oesophageal Doppler in perioperative medicine: new opportunities in research and clinical practice.

Authors:  Bernardo Bollen Pinto; Glen Atlas; Bart F Geerts; Karim Bendjelid
Journal:  J Clin Monit Comput       Date:  2016-08-27       Impact factor: 2.502

Review 2.  Evolving concepts of hemodynamic monitoring for critically ill patients.

Authors:  Olfa Hamzaoui; Xavier Monnet; Jean-Louis Teboul
Journal:  Indian J Crit Care Med       Date:  2015-04

3.  End-tidal carbon dioxide variation after a 100- and a 500-ml fluid challenge to assess fluid responsiveness.

Authors:  Matthias Jacquet-Lagrèze; Florent Baudin; Jean Stéphane David; Jean-Luc Fellahi; Patrick B Hu; Marc Lilot; Vincent Piriou
Journal:  Ann Intensive Care       Date:  2016-04-22       Impact factor: 6.925

4.  How to detect a positive response to a fluid bolus when cardiac output is not measured?

Authors:  Zakaria Ait-Hamou; Jean-Louis Teboul; Nadia Anguel; Xavier Monnet
Journal:  Ann Intensive Care       Date:  2019-12-16       Impact factor: 6.925

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.