| Literature DB >> 23783737 |
Michael Jørgen Hansen1, Anders Peter S Adamsen, Anders Feilberg.
Abstract
The aim of the present study was to examine the recovery of odorants during the dilution in an olfactometer designed according to the European standard for dynamic olfactometry. Nine odorants in the ppmv-range were examined including hydrogen sulfide, methanethiol, dimethyl sulfide, acetic acid, propanoic acid, butanoic acid, trimethylamine, 3-methylphenol and n-butanol. Each odorant was diluted in six dilution steps in descending order from 4,096 to 128 times dilutions. The final recovery of dimethyl sulfide and n-butanol after a 60-second pulse was only slightly affected by the dilution, whereas the recoveries of the other odorants were significantly affected by the dilution. The final recoveries of carboxylic acids, trimethylamine and 3-methylphenol were affected by the pulse duration and the signals did not reach stable levels within the 60-second pulse, while sulfur compounds and n-butanol reach a stable signal within a few seconds. In conclusion, the dilution of odorants in an olfactometer has a high impact on the recovery of odorants and when olfactometry is used to estimate the odor concentration, the recoveries have to be taken into consideration for correct measurements.Entities:
Mesh:
Substances:
Year: 2013 PMID: 23783737 PMCID: PMC3715246 DOI: 10.3390/s130607860
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.Schematic drawing of the experimental setup.
Concentrations of odorants prior to dilution in the olfactometer and the concentrations measured at each dilution step by PTR-MS. The expected concentrations are shown in brackets.
| 35 | 49 | 63 | 60 | 61 | 75 | 89 | 109 | 57 | |
| DL | 0.2 | 0.02 | 0.1 | 0.2 | 0.1 | 0.04 | 0.1 | 0.2 | 0.2 |
| Undiluted sample, ppmv | 5.0 | 5.2 | 5.4 | 5.0 | 5.7 | 4.3 | 5.1 | 3.2 | 5.0 |
|
| |||||||||
| Dilution factor | ppbv ± 1 SD | ||||||||
|
| |||||||||
| 4,096 | 0.5 ± 0.2 (1.2) | 0.9 ± 0.1 (1.3) | 1.5 ± 0.2 (1.3) | <0.2 | 0.6 ± 0.1 (1.4) | 0.2 ± 0.1 (1.1) | 0.2 ± 0.1 (1.3) | 0.5 ± 0.1 (0.8) | 1.3 ± 0.3 (1.2) |
| 2,048 | 1.1 ± 0.2 (2.4) | 1.7 ± 0.1 (2.5) | 2.6 ± 0.2 (2.7) | <0.2 | 1.3 ± 0.2 (2.8) | 0.6 ± 0.1 (2.1) | 0.6 ± 0.1 (2.5) | 1.2 ± 0.2 (1.6) | 2.2 ± 0.3 (2.4) |
| 1,024 | 2.1 ± 0.4 (4.9) | 3.5 ± 0.3 (5.0) | 5.4 ± 0.3 (5.3) | 0.5 ± 0.2 (4.9) | 3.3 ± 0.3 (5.5) | 1.3 ± 0.1 (4.2) | 1.1 ± 0.1 (5.0) | 2.2 ± 0.2 (3.2) | 4.5 ± 0.5 (4.9) |
| 512 | 4.1 ± 0.6 (9.7) | 7.9 ± 0.6 (10) | 11 ± 0.4 (11) | 1.8 ± 0.2 (9.9) | 7.9 ± 0.4 (11) | 3.5 ± 0.2 (8.5) | 3.4 ± 0.2 (10) | 4.7 ± 0.4 (6.3) | 9.3 ± 0.4 (9.8) |
| 256 | 6.2 ± 0.5 (19) | 16 ± 0.7 (20) | 22 ± 0.5 (21) | 6.3 ± 0.4 (20) | 17 ± 0.5 (22) | 9.7 ± 0.3 (17) | 10 ± 0.3 (20) | 9.0 ± 0.4 (13) | 18 ± 0.5 (20) |
| 128 | 14 ± 1 | 35 ± 0.7 (40) | 43 ± 1 | 22 ± 1 | 37 ± 0.9 (44) | 24 ± 0.7 (34) | 27 ± 0.6 (40) | 19 ± 0.6 (25) | 36 ± 1 |
H2S: hydrogen sulfide; MT: methanethiol; DMS: dimethyl sulfide; TMA: trimethylamine; AA: acetic acid; PA: propanoic acid; BA: butanoic acid; 3MP: 3-methylphenol; NB: n-butanol;
m/z: mass-to-charge ratio;
DL: Detection limit estimated as three times the standard deviation on a blank sample (dry air).
Figure 2.Recovery of odorants in the outlet tubing of an olfactometer as a function of the dilution factor.
Figure 3.Development in the recovery during a 60-second pulse for selected odorants in an olfactometer. The odorants were introduced to the olfactometer after 10 s and the pulse stopped after 70 s.
Figure 4.PTR-MS response time for individual odorants measured with a 1.2 m PEEK sampling line heated to 60 °C. The PTR-MS response time was defined as the time to reach 90% of the final concentration when changing from a highly diluted sample to an undiluted sample (relative time = 10 s).
Figure 5.Correlation between the concentration of dimethyl sulfide measured in the outlet tubing from the olfactometer and in the nose cone where the human nose is placed.