| Literature DB >> 23781417 |
Vandana Dhillon1, Alok Kumar Chaudhary.
Abstract
A simple HPTLC method having high accuracy, precision and reproducibility was developed for the routine estimation of moxifloxacin hydrochloride in the tablets available in market and was validated for various parameters according to ICH guidelines. moxifloxacin hydrochloride was estimated at 292 nm by densitometry using Silica gel 60 F254 as stationary phase and a premix of methylene chloride: methanol: strong ammonia solution and acetonitrile (10:10:5:10) as mobile phase. Method was found linear in a range of 9-54 nanograms with a correlation coefficient >0.99. The regression equation was: AUC = 65.57 × (Amount in nanograms) + 163 (r(2) = 0.9908).Entities:
Keywords: HPTLC; moxifloxacin hydrochloride and Validation
Year: 2010 PMID: 23781417 PMCID: PMC3658017 DOI: 10.4103/2229-4708.72232
Source DB: PubMed Journal: Pharm Methods ISSN: 2229-4708
Figure 1HPTLC chromatogram of the constituted tablet placebo.
Figure 2Chromatogram of the moxifloxacin hydrochloride Tablet
Calibration data for linearity
Regression analysis
Method precision of the analytical method
Recovery study