We present a novel swept source optical coherence tomography configuration, equipped with acousto-optic deflectors that can be used to simultaneously acquire multiple B-scans originating from different depths. The sensitivity range of the configuration is evaluated while acquiring five simultaneous B-scans. Then the configuration is employed to demonstrate long range B-scan imaging by combining two simultaneous B-scans from a mouse head sample.
We present a novel swept source optical coherence tomography configuration, equipped with acousto-optic deflectors that can be used to simultaneously acquire multiple B-scans originating from different depths. The sensitivity range of the configuration is evaluated while acquiring five simultaneous B-scans. Then the configuration is employed to demonstrate long range B-scan imaging by combining two simultaneous B-scans from a mouse head sample.
Optical coherence tomography (OCT) [1] is a non invasive
imaging technique that has the potential to provide fast and high resolution depth-resolved
information of the scattering properties of biological samples. To obtain an axial scan, time
domain (TD)-OCT systems rely on the slow reference arm scanning [2]. Later developed spectral domain (SD)-OCT systems enabled fast acquisition [3] of axial scans within a range of several millimeters. One of
the major limitation of the SD-OCT technology remains its depth dependent sensitivity drop, an
effect commonly called as the sensitivity “roll-off”. In swept source (SS)-OCT
systems, the sensitivity roll-off is due to the limited coherence length of the swept source
used, while in spectrometer based (SB)-OCT systems, the limited resolution of the spectrometer
causes the roll-off [2].Recent progress in the development of tunable lasers for SS-OCT lead to swept sources with
coherence length longer than 1 cm [4,5]. However, when increasing the tuning speed, their coherence length
diminishes [6]. There is an interest in imaging objects
with axial extension larger than 1 cm such as the human eye [4], or perform topography of objects with variation in height larger than 1 cm.
Therefore, there is a need for optical sources of longer coherence length, or equivalently, of
methods that can compensate the sensitivity roll-off with depth.We recognize that in typical OCT biomedical applications, such as intravascular and
gastrointestinal imaging, the light cannot penetrate biological tissue deeper than few
millimeters [7] and the loss of backscattered light is a
limiting factor, not the roll-off. However, in several biomedical applications such as balloon
based endoscopic OCT [8] or gastrointestinal imaging
[9], the distance to the sample can vary by few tenths of
millimetres and the high imaging range is desired to obtain consistent quality scans.Various approaches to increase the imaging range have already been reported. Several
techniques have been developed to remove the mirror terms [10-12] and effectively double the axial
range by giving access to full axial range. A different approach, consisted in using a pair of
recirculation loops in the object and reference arms [13]
with a differential optical path difference comparable with the axial range given by the inverse
of the line width of the swept source and using radio frequency shifts imprinted at each
recirculation through the secondary loops, by frequency shifters.In this paper we present a technique which allows substantial extension in the imaging range
by: (i) elimination of mirror terms [14] and (ii)
simultaneous acquisition of multiple axially shifted A-scans, by employing a frequency domain
multiplexing method [15]. The multiple A-scans are
encoded at different radiofrequencies (RF) employing the same approach as in frequency
multiplexed time domain OCT [15]. The use of a
multiplexer, comprising of two lenses and a pair of a Bragg cells [16,17] in conjunction with a multiple
delay element, allows producing several conventional A-scans at different depths simultaneously
in multiple channels OCT. Each channel provides a full axial range A-scan, as it is free from
mirror terms. It is also possible to adjust the differential depth difference in the multiple
delay element, δ, that allows superimposing of several adjacent A-scans. In this way,
decreased sensitivity in an A-scan delivered by a channel is compensated by the increase in
sensitivity in the adjacent channel.
2. Method
2.1 Set-up
The schematic diagram of a long axial range SS-OCT system is depicted in Fig. 1.
Fig. 1
Schematic layout of the long range SS-OCT system. SS: swept source; MO1,2,3,4: microscope
objectives; AOD11, 12, 21, 22: acousto – optic deflectors; POLC: polarization
controllers; GS: galvanometer scanner equipped with a mirror to scan along X direction; Tx:
scanner controller; L1, L2: lenses; M: mirror; DFS1,2: digital frequency synthesizer; DCR:
dispersion compensator rod; AL: achromatic lens; OL: Objective lens; BS: beam-splitter; PC:
personal computer equipped with a NI-5124 digitizer to decode each axial scan and produce
the image on the PC display.
Schematic layout of the long range SS-OCT system. SS: swept source; MO1,2,3,4: microscope
objectives; AOD11, 12, 21, 22: acousto – optic deflectors; POLC: polarization
controllers; GS: galvanometer scanner equipped with a mirror to scan along X direction; Tx:
scanner controller; L1, L2: lenses; M: mirror; DFS1,2: digital frequency synthesizer; DCR:
dispersion compensator rod; AL: achromatic lens; OL: Objective lens; BS: beam-splitter; PC:
personal computer equipped with a NI-5124 digitizer to decode each axial scan and produce
the image on the PC display.The swept source (Superlum, Cork, Ireland, model Broad-sweeper 840) is a tunable
semiconductor light source, operating at 100000 nm/s sweeping speed, with a 828.0 –
873.0 nm spectral range. This corresponds to scanning δλ = 45 nm in 0.45 ms at a
rate γ = 2.2 kHz, with an equivalent repetition rate of 2 kHz. The light is amplified by
an Optical Power Booster (Superlum, Cork, Ireland, model 850-50) to achieve more than 23 mW in
single mode fiber. The light from the booster is then injected into a 50/50 single mode
directional coupler that divides light into an object and a reference arm. The polarization of
light is controlled using Newport F-POL-IL polarization controllers, POLC. The light output of
the first coupler is collimated with two microscope objectives MO1 (New Focus 5724-B-H) and MO2
(10x OFR LLO-4-18-NIR).The light in the reference arm then traverses two acousto-optic deflectors, AOD21 and AOD22,
from Gooch and Housego. The deflectors are operated with a Gooch and Housego digital frequency
synthesizer MSD040-150-0.8ADS2-A5H-8X1. The frequency synthesizer is controlled via a computer
and can be tuned within 40 - 150 MHz range in 1 kHz increments. It can generate a maximum of P
= 8 channels operating simultaneously with up to 200 mW of RF power delivered per channel. The
light passing through the AOD21 can be diffracted into a number of beams, from 1 to 8.Lens L1 (f = 500 mm) is assembled from two closely spaced plano-convex BK7 singlet lenses,
both with focal length 1 m, and placed at 2f distance from the deflectors. The multiple optical
delays are inserted immediately after the L1 so that each diffracted beam passes through a
different glass block as depicted in Fig. 2.
Fig. 2
Frequency division multiplexer. where each diffracted beam traverses a different delay
step. The steps are made from glass plates of optical thickness δ.
Frequency division multiplexer. where each diffracted beam traverses a different delay
step. The steps are made from glass plates of optical thickness δ.The lens L1 converges the fan of diffracted beams on the AOD22, driven by the same set of P
RF signals, of identical frequencies as applied to AOD21. Light levels on each diffracted beam
can be digitally controlled in each of the P channels. A lens L2 (f = 500 mm) is placed
immediately after the AOD 22 to collimate the multiplexed beam. After passing the AOD22 and L2,
all beams in the first diffracted order are collimated and collinear, therefore they can be
easily injected into one of the inputs of a second 50/50 single mode optical fiber coupler,
using the MO4.The object arm is equipped with the same set of Bragg cells and lenses to compensate for
dispersion, however a single RF signal of 60 MHz is applied on both AOD11 and AOD12.In the object arm, light is directed by a mirror M to a thick 50/50 beam-splitter plate, BS,
which reflects the object beam to a galvanometer X-scanner equipped with a mirror, GS. The
X-scanner determines the line in the final image and is driven by a generator TX
with a sawtooth signal, generated by a Thurlby Thandar Instruments TG210
frequency generator. The TTL output from the frequency generator is connected to the PC via a
NI-5124 digitizer. The beam is focused on the object using a pair of achromatic doublet lenses,
AL of focal length 75 mm, and an objective lens OL of focal length 40 mm. The lenses are
arranged in a 4f configuration, which reimages the pupil plane on the galvo-scanner mirror with
magnification equal to 1. The reflected light from the object propagates back along the same
path up to the BS, wherefrom is launched via MO3 into the other input of the second 50/50
single mode fiber coupler, where it interferes with the light from the reference arm. M3 and M4
use aspheric lenses New Focus 5723-B-H. An asymmetric choice of microscope
objective and lenses was made to (i) minimize the diameter of the beams travelling through the
acousto-optic deflectors and (ii) increase the clearance between the beams in the reference arm
at the point where the multiple delay element is inserted.To compensate for the dispersion due to lenses AL and OL and the beam splitter BS, a
dispersion compensation rod DCR is inserted in the beam after the lens L2 in the reference arm.
After the AODs in both arms, only the first-order diffracted beams are selected by using
pinholes placed conveniently (not shown).The interference signal is photo-detected with a New Focus 1807 balanced
photo-detector of 80 MHz electrical bandwidth. The RF spectrum of the photodetected signal is
digitized and analysed using a National Instruments PXI-5124 card, which is
installed in a PC.
2.2 Choice of excitation frequencies
All AODs are used to simultaneously create multiple signal carriers by upshifting the
frequency of each deflected beam. Sufficient separation of diffracted beams by AOD21,
comparable to the size of the beam diameters of a few mm requires a difference of frequency in
the signals applied to the AOD larger than several MHz, chosen here as 9.376 MHz. This
simplifies the construction of a delay element by relaxing the step edge quality requirements
and making the delay elements compatible with a beam diameter of up to 2 mm. The AODs in the
object arm are driven at 60 MHz and the AODs in the reference arm are driven simultaneously at
fp = 65.247 MHz + (p - 1)*9.376 MHz, with p = 1 to P. The drivers can drive up to P
= 8 signals. In the experiments which follow, the maximum value for P was 4. The resulting
carriers in the photo-detected signal therefore pulsate at 2fp −120 MHz, i.e.
at frequencies:where C0 = 130.494-120 =
10.494 MHz and ΔF = 18.752 MHz as depicted in Fig.
3 [15].
Fig. 3
Data processing algorithm.
Data processing algorithm.
2.3 Data processing
The data acquired by the photo-detector, is digitized by the NI PCI-5124
card, whose acquisition is triggered by the TTL sweeping signal provided by the SS and by TTL
signal associated to the signal produced by the frequency generator used to drive the
galvo-scanner. Thus a B-scan image is created.First, the RF spectra representing single sweeps are extracted from the photo-detected signal
using the TTL signal provided by the SS. The SS is driven in a saw-tooth fashion, therefore the
signal due to the return of the frequency to the initial value (50 microseconds) is eliminated.
Each packet extracted in this way is processed to produce an A-scan. The packets then are
resampled according to known procedures in the practice of spectral domain OCT, to linearize
the data, i.e. to place the data in equal frequency slots. Then a fast Fourier transform (FFT)
is performed on each sweep to produce a 1D array, containing the A-scan interference signal
amplitude data for each of the carriers Cp, as illustrated in Fig. 3. Using the data provided by the frequency generator TTL signal,
multiple A-scans are assembled into a continuous A-scan. Then the A-scans are assembled into a
2D array to produce a B-scan OCT image.The frequency Fp of the photo-detected signal in each channel p, due to sweeping
the optical frequency at a rate γ [Hz], for a given value of the OPD in the
interferometer is:where Δk is the tuning bandwidth in
wave numbers, Cp the carrier frequency imprinted by frequency shifting on each
channel p and S is the conversion efficiency of the swept source interferometer, which converts
OPD into RF frequency. Using the numerical values above, S = 137 kHz/mm. Using Eq. (1) in Eq.
(2),Equation (3) shows that different channels p determine different frequency for the
photodetected signal depending on p.
2.4 Choice of differential delay δ
As shown by Eq. (3), the same OPD determines
different frequency components in the photo-detected signal. In a different configuration
applying a similar concept [13], the dependence on p was
eliminated by making the last bracket in Eq. (3)
zero. Here however, this cannot be achieved, given the need to separate the diffracted beams
out of the AOD sufficiently laterally to allow introduction of different delay elements, as
explained above, which justified the choice of a step increase from one carrier to the next of
9.376 MHz. With S = 137 kHz/mm and ΔF = 18.752 MHz, these would determine a value
δi too large for the differential step in OPD between channels
of:The value of δi is too
large in comparison with the axial range of each channel p. Measurement of the axial range of
each channel p, as shown below leads to a range of a few mm. The differential delay between
channels, δ, should be comparable with the maximum axial range achievable per channel in
order to obtain, by combination of channels, an overall constant sensitivity over an extended
axial range. For instance, if a constant overall sensitivity is targeted, then sensitivity
curves of neighboring channels should be shifted axially to each other until superposed over
their FWHM points. This means a differential delay δC = FWHM, where FWHM is
the OPD value of the sensitivity curve versus depth where sensitivity drops to half. The
measurements below will show that FWHM is approximately 2 mm. An alternative strategy would be
to accept a non-constant sensitivity with depth, but over an extended range. We adapted this
strategy here and decided to shift the curves of sensitivity laterally by twice more than the
FWHM of each individual channel. Three slabs of glass giving a delay δ = 4.2 mm where
found which satisfy this requirement. With this value in (4), the last bracket cannot be made
zero (as in [11]), as δ <<
δi. This means that each channel will occupy a restricted bandwidth in a
spectral window around each carrier Cp, to a range determined by multiplying S with
the axial range per channel, Δ. Using a value Δ = 2 - 4 mm, this gives a
bandwidth of plus minus B ≈0.5 - 1 MHz (an exact calculation of bandwidth is presented in the
next paragraph), much less than the frequency difference between adjacent carriers, ΔF =
18.752 MHz, as shown schematically in Fig. 4(a). Figure 4(b) shows a simplified sketch of
superposition of sensitivity profiles versus OPD, where δ ≈2 FWHM. This means
that in order to obtain the final long range B-scan, each spectral window needs to be cropped
from the photo-detected signal and shifted down in frequency by:Using a differential delay of 4.2 mm,
this givesIn this way, the frequency generated by
each channel after being shifted becomes independent on p, giving in all cases the same
value:The final long range B-scan of the curves
in Fig. 4(b) is obtained by adding the intensity values
of each overlapping spectral window. The frequency values in the resulting long scan can be
recalculated in OPD values using an equation derived from Eq. (5):
Fig. 4
a: Sensitivity curves in each channel versus optical path difference, each
centered around a different carrier frequency, Cp; b: schematic illustration of
combining the sensitivity of several channels versus OPD to obtain a more constant variation
of sensitivity.
a: Sensitivity curves in each channel versus optical path difference, each
centered around a different carrier frequency, Cp; b: schematic illustration of
combining the sensitivity of several channels versus OPD to obtain a more constant variation
of sensitivity.
3. Results
3.1 Spectrum
Figure 5 shows the sensitivity curves for 4 delayed channels and the overall cumulated curve
obtained by their summation. A multiple delay element consisting in 3 steps of δ = 4.2
mm and a mirror as object are used in Fig. 1. The set
carrier frequencies of the four channels for OPD = 0 is f1 = 10.494 MHz,
f2 = 19.87 MHz, f3 = 29.246 MHz and f4 = 38.622 MHz. The
multiple curves are raised by varying the reference arm length by moving the fiber launcher
holding the MO4.
Fig. 5
Sensitivity curve for P = 4 channels delayed relatively by δ = 4.2 mm.
Sensitivity curve for P = 4 channels delayed relatively by δ = 4.2 mm.The data points for each sensitivity curve were measured using a HP 8590A spectrum analyzer.
The data collected were then fitted with a 4th order polynomial spline. A curve representing
the long range A-Scan sensitivity was calculated as a sum of fitted curves for the single
A-scan sensitivities.Figure 5 shows that the sensitivity of each individual
channel drops by more than 6 dB for an OPD = 2 mm away from the maximum, giving less than 4 mm
usable range for a single channel. However, combining the data from all 4 carriers extends the
axial range of near constant sensitivity to more than 16 mm imaging range.The regime of operation utilized here is of P = 4 channels with independent demodulation. The
axial range is multiplied by an approximate factor of P = 4. If a swept source with 4 times
longer coherence length was used, the bandwidth of the photodetector unit would have needed to
accommodate a bandwidth of 4B, where B can be evaluated by considering the maximum number of
axial pixels within the axial range δ = 4.2 mm. Using a bandwidth of 45 nm and a central
wavelength of 0.85 μm, the number of pixels is M = 594. Using a scanning time of 0.5 ms,
this gives a value B = 1.19 MHz. However, because here independent demodulation is implemented,
the frequency of each carrier shifts from Cp - B to Cp + B, i.e. a 2B bandwidth is needed on
each channel. In total, the bandwidth is not P times larger, but 2P larger, ie 2PB = 8B, as a
disadvantage of this configuration when compared to the case of a swept source with a 4 times
longer coherence length employed in a conventional single path configuration.However, if the denser packing of carriers is employed, according to Fig. 4(b) and satisfying the Eq.
(4) as in [13], then the bandwidth needed is
only PB.
3.2 Imaging
To illustrate the capability of the system to produce a long range B-scan, in
vitro imaging of a formalin fixed C57BL/J mouse tissue sample was performed. A mouse
head (Fig. 6(e)) was positioned in front of OL and rotated in a way that the image would span through
the major part of the sensitivity range. For this demonstration, the AODs were operated using a
set of two frequencies C1 and C2. Only two channels were used for imaging because the 8 mm
sensitivity range (where variation of sensitivity due to roll-off is less than 6 dB) already
exceeds the focal depth of the lens OL.
Fig. 6
B-scan images of a tilted mouse head obtained with a long-range constant sensitivity
SS-OCT system. a: Conventional B-scan carried by C1; b:
Conventional B-scan carried by C2; c): A long range B-scan produced by the long
range imaging software; d: experimentally measured sensitivity curve in air
representing the system sensitivity variation as a function of reference OPD. The curve is
positioned so that OPD in the graph corresponds to the OPD in a, b
and c; e a photo of the sample. The distance z in the object is
OPD/2. Due to a long exposure time the scanning beam is visible as a white line across the
eye and surrounding tissues.
B-scan images of a tilted mouse head obtained with a long-range constant sensitivity
SS-OCT system. a: Conventional B-scan carried by C1; b:
Conventional B-scan carried by C2; c): A long range B-scan produced by the long
range imaging software; d: experimentally measured sensitivity curve in air
representing the system sensitivity variation as a function of reference OPD. The curve is
positioned so that OPD in the graph corresponds to the OPD in a, b
and c; e a photo of the sample. The distance z in the object is
OPD/2. Due to a long exposure time the scanning beam is visible as a white line across the
eye and surrounding tissues.Figure 6 shows two regular B-scans in Fig. 6(a) and in Fig.
6(b) that are obtained simultaneously. Each is delivered by a separate channel, CH1 and
CH2 respectively, that are tuned on the carriers C1 and C2 respectively.
The B-scans recorded from both channels are stitched by the imaging software to produce a long
range B-scan, as presented in Fig. 6(c).Figures 6(a) and 6(b) illustrate an additional challenge raised by this configuration. The images
present several white lines due to Bragg cell intermodulation products [18]. The lines are always present at the same spectral locations when the
same set of RF frequencies is applied to the Bragg cells. They are present even without
interference. Therefore, these lines are removed by acquiring reference images with no object
present, which subsequently are used for background subtraction from the current images before
stitching them to form the final image in Fig. 6(c).Possibly, the intensity of intermodulation peaks that are present in the RF spectrum may be
significantly reduced by careful selection of acousto-optic materials and excitation RFs.The sensitivity versus OPD curve presented in Fig.
6(d) is aligned to the images so that the graph represents the system sensitivity when
acquiring the data from corresponding lines in the B-scan shown in Fig. 6(c). The position of the scanning object beam, while acquiring the
long range B-scan, is visible in Fig. 6(e) as a line
across the eye and surrounding tissues.
4. Discussion
The method implemented here is different from that reported in [13], where the differential delay between OCT channels was adjusted to match
the value obtained from the difference of frequency between the channels divided by the
conversion constant. S, between the OPD and the RF frequency generated when tuning the SS. Such
a matching was not possible here as explained above. The method presented can be applied to up
to P = 8 carriers, number limited here by the AOD drivers. However, in practice there should be
a correlation between the compounded axial range achievable and the confocal axial range
determined by the interface optics. This is why we compounded images produced by two carriers
only, as the confocal depth range limited by the interface optics was below 5 mm. Further
research into long axial range OCT images should go hand in hand with methods to alleviate the
limitation of sensitivity due to the confocal gate. Such methods may involve for example digital
focusing [19] or use of non-conventional objective lenses
[20-22].Frequency multiplexing could be combined with recently developed emerging swept source
technologies such as micro-electro-mechanical systems (MEMs) tunable vertical cavity surface
emitting lasers (VCSEL) [5] to offer an even further
increased axial range. Such a long axial range may find applications in sensing, profiling or
tracking of objects.The method presented is especially suitable for those swept sources where by increasing the
tuning speed, their coherence length diminishes. In combining such sources with multiple path
configurations as described here, to increase the axial range of the OCT system beyond the limit
set by the source coherence length, the bandwidth of the photodetection unit should also be
taken into account. In this paper the principle of extending the axial range was demonstrated
using a configuration with P = 4 delays in 4 independent channels and using a low rate tunable
swept source, where the bandwidth due to tuning was small. If the line rate is increased to the
frequency of fast commercial available swept sources of 200 kHz (100 times larger than the
tuning rate used here), then a bandwidth B = 119 MHz would be needed. If non superposition of
frequencies generated in the four channels is pursued, as illustrated in Fig. 4(a) and employed here, this would require values for the frequency of
the first 4 carriers of at least f1 = B, f2 = f1 + 2B,
f3 = f2 + 2B and f4 = f3 + 2B, and so on, i.e. of
at least 119, 337, 575 and 813 MHz respectively. The total photodetection bandwidth would be 2PB
= 952 MHz. However, if the swept source is faster, then the adjustment used in [13] becomes achievable, in which case a more compact frequency
space is achievable, as implemented in [13]. In this
case, the bandwidth required is exactly that for 4 times more axial pixels, i.e. 476 MHz
only.Swept sources with tuning frequencies of over 1 MHz [23] have also been reported, which would require acousto-optic modulators and
photodetecting units operating at frequencies over several GHz.In our experiment, the setup presented was optimized to offer a long sensitivity range on the
expense of some allowed variation of sensitivity within the range. The ripples in the overall
sensitivity curve are less than 2 dB. Reduced rippling can be obtained by choosing a
differential delay element δ so that the imaging channels would come even closer
together, in terms of OPD, providing a relatively flat top sensitivity curve when combined, as
suggested by the profile in Fig. 4(b).
5. Conclusions
A system capable to perform long range SS-OCT imaging by combining several mirror terms free
B-scans acquired simultaneously at different depths is presented. The system uses AOM frequency
shifters implementing a multiple delay element with P delays in the reference arm, each
traversed by a deflected beam at a different angle and with a different frequency shift as
imprinted by the AOMs. The part of the reference beam traversing a particular delay in the
multiple delay element is coded in the RF frequency of the photo-detected signal. In this way,
A-scans and B-scans corresponding to different axial ranges can be identified.The system is demonstrated for P = 4 delays which covers more than 17 mm OPD axial range
measured at 6 dB, with a variation of less than 2 dB inside this axial range. The capability of
producing long-range B-scan images was proven by imaging a part of a mouse head extended over 1
cm. The method presented acts on the interferometer configuration and not on the source and
therefore can be combined with any method acting on the swept source.
Authors: Wolfgang Wieser; Thomas Klein; Desmond C Adler; Francois Trépanier; Christoph M Eigenwillig; Sebastian Karpf; Joseph M Schmitt; Robert Huber Journal: Biomed Opt Express Date: 2012-09-24 Impact factor: 3.732