Literature DB >> 23389175

Simultaneous multiple-depths en-face optical coherence tomography using multiple signal excitation of acousto-optic deflectors.

Mantas Zurauskas1, John Rogers, Adrian Gh Podoleanu.   

Abstract

We present a novel low-coherence interferometer configuration, equipped with acousto-optic deflectors that can be used to simultaneously acquire up to eight time domain optical coherence tomography en-face images. The capabilities of the configuration are evaluated in terms of depth resolution, signal to noise ratio and crosstalk. Then the configuration is employed to demonstrate simultaneous en-face optical coherence tomography imaging at five different depths in a specimen of armadillidium vulgare.

Entities:  

Mesh:

Year:  2013        PMID: 23389175     DOI: 10.1364/OE.21.001925

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  2 in total

1.  Dermascope guided multiple reference optical coherence tomography.

Authors:  Roshan Dsouza; Hrebesh Subhash; Kai Neuhaus; Josh Hogan; Carol Wilson; Martin Leahy
Journal:  Biomed Opt Express       Date:  2014-08-01       Impact factor: 3.732

2.  Frequency multiplexed long range swept source optical coherence tomography.

Authors:  Mantas Zurauskas; Adrian Bradu; Adrian Gh Podoleanu
Journal:  Biomed Opt Express       Date:  2013-05-02       Impact factor: 3.732

  2 in total