Literature DB >> 2373943

Frequency and reliability of alarms in the monitoring of cardiac postoperative patients.

E M Koski1, A Mäkivirta, T Sukuvaara, A Kari.   

Abstract

Postoperative monitoring of cardiac operated patients requires appropriately functioning monitor alarms as well as intensive nursing activity. The limit alarms can be used for detection of life-threatening situations and monitoring of physiological changes in the patient's state. We studied the significance and the frequency of audible alarms during the postoperative intensive care of ten cardiac patients. Of 1307 occasions when such an alarm was activated during the study period of approximately 26 hours per patient, only 139 (10.6%) were significant. The highest proportion of significant audible limit alarms was found during the immediate postoperative period. Heart rate alarms were more reliable than alarms of the other parameters monitored in the study. Possibilities for improving the physiological monitoring and alarm system are discussed.

Entities:  

Mesh:

Year:  1990        PMID: 2373943     DOI: 10.1007/bf01724206

Source DB:  PubMed          Journal:  Int J Clin Monit Comput        ISSN: 0167-9945


  7 in total

1.  Building ICU artifact detection models with more data in less time.

Authors:  C L Tsien; I S Kohane; N McIntosh
Journal:  Proc AMIA Symp       Date:  2001

2.  Influence of pulse oximeter settings on the frequency of alarms and detection of hypoxemia: Theoretical effects of artifact rejection, alarm delay, averaging, median filtering or a lower setting of the alarm limit.

Authors:  A T Rheineck-Leyssius; C J Kalkman
Journal:  J Clin Monit Comput       Date:  1998-04       Impact factor: 2.502

3.  Clinician-Driven Design of VitalPAD-An Intelligent Monitoring and Communication Device to Improve Patient Safety in the Intensive Care Unit.

Authors:  Luisa Flohr; Shaylene Beaudry; K Taneille Johnson; Nicholas West; Catherine M Burns; J Mark Ansermino; Guy A Dumont; David Wensley; Peter Skippen; Matthias Gorges
Journal:  IEEE J Transl Eng Health Med       Date:  2018-03-05       Impact factor: 3.316

4.  An expert system for monitor alarm integration.

Authors:  C Oberli; J Urzua; C Saez; M Guarini; A Ciprianio; B Garayar; G Lema; R Canessa; C Sacco; M Irarrazaval
Journal:  J Clin Monit Comput       Date:  1999-01       Impact factor: 2.502

5.  Alarm-inducing variability in cardiac postoperative data and the effects of prealarm delay.

Authors:  A Mäkivirta; E M Koski
Journal:  J Clin Monit       Date:  1994-05

Review 6.  Medical audible alarms: a review.

Authors:  Judy Edworthy
Journal:  J Am Med Inform Assoc       Date:  2012-10-25       Impact factor: 4.497

Review 7.  Systematic Review of Physiologic Monitor Alarm Characteristics and Pragmatic Interventions to Reduce Alarm Frequency.

Authors:  Christine Weirich Paine; Veena V Goel; Elizabeth Ely; Christopher D Stave; Shannon Stemler; Miriam Zander; Christopher P Bonafide
Journal:  J Hosp Med       Date:  2015-12-14       Impact factor: 2.960