| Literature DB >> 23730378 |
Abstract
BACKGROUND: Mobile bearing TKA prostheses were designed to minimize polyethylene wear by increasing implant conformity and reducing stresses between the articulating prosthesis components. It is the purpose of this study to assess the mid-term functionality and clinical outcome associated with a highly congruent mobile platform design, the e.motion(®) UC total knee prosthesis.Entities:
Keywords: Mobile bearing; TKA; radiolucent lines; rotating platform; total knee arthroplasty; ultra-congruent.
Year: 2013 PMID: 23730378 PMCID: PMC3664449 DOI: 10.2174/1874325001307010152
Source DB: PubMed Journal: Open Orthop J ISSN: 1874-3250
Characteristics of the 28 Patients who Received an e.motion® UC Knee Prosthesis and Duration of the Follow-Ups
| Mean | ± | SD | Range | |||
|---|---|---|---|---|---|---|
| (min | ; | max) | ||||
| Age (years) | 72.4 | ± | 7.5 | (60.0 | ; | 84.0) |
| Height (cm) | 161.8 | ± | 8.0 | (150.0 | ; | 177.0) |
| Weight (kg) | 76.8 | ± | 14.5 | (55.0 | ; | 120.0) |
| BMI (kg/m2) | 29.3 | ± | 5.2 | (23.0 | ; | 42.0) |
| Age (years) | 77.8 | ± | 7.5 | (65.5 | ; | 89.7) |
| 5-year visit (yrs) | 5.6 | ± | 0.1 | (5.2 | ; | 6.0) |
| 2-year visit (yrs) | 2.4 | ± | 0.3 | (1.9 | ; | 2.7) |
| 1-year visit (yrs) | 1.1 | ± | 0.2 | (0.8 | ; | 1.4) |
| 3-month visit (mths) | 2.9 | ± | 0.7 | (0.07 | ; | 4.0) |
SD: standard deviation.
BMI: body mass index, kg/m2 (= weight/height2).
Comparison in the KOOS Subscales Between Patients who Received the e.motion® UC Knee Prosthesis and Patients who Received the LCS Prosthesis (Lygre et al., 2010)
| e.motion® UC Prosthesis (Follow-Up: 5.6 Years) | LCS Prosthesis | Difference Between Prostheses | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Resurfaced (N=12) | Non-Resurfaced (N=16) | Weighted Mean (N=28) | Resurfaced (N=184) | Non-Resurfaced (N=180) | Weighted Mean (N= 364) | (e.motion® – LCS), Weighted Means | |||||||||
| Mean | ± | SE | Mean | ± | SE | Mean | ± | SE | Mean | ± | SE | ||||
| ADL | 77,5 | ± | 18,7 | 78,1 | ± | 15,8 | 77,8 | 70,8 | ± | 1,9 | 71,6 | ± | 2,1 | 71,2 | 6,6 |
| Pain | 83,1 | ± | 22,9 | 84,9 | ± | 13,9 | 84,1 | 76,1 | ± | 1,8 | 75,8 | ± | 2 | 76 | 8,1 |
| Symptoms | 82,1 | ± | 19,9 | 86,6 | ± | 11,9 | 84,7 | 78 | ± | 1,5 | 79,9 | ± | 1,5 | 78,9 | 5,8 |
| Sports | 37,1 | ± | 21,2 | 37,8 | ± | 38,6 | 37,5 | 35,9 | ± | 2,3 | 38,3 | ± | 2,6 | 37,1 | 0,4 |
| QoL | 70,8 | ± | 23 | 73,8 | ± | 19,1 | 72,5 | 59,6 | ± | 2,3 | 62,3 | ± | 2,3 | 60,9 | 11,6 |
SE: standard error; UC: ultra-congruent; LCS: low-contact stress
The Knee injury and Osteoarthritis Outcome Score (KOOS) is a 42-item self‑administered questionnaire that covers 5 dimensions: pain (“Pain”, 9 items), other disease‑specific symptoms (“Symptoms”, 7 items), activities of daily living (“ADL”, 17 items), sport and recreation function (“Sports”, 5 items), and knee-related quality of life (“QoL”, 4 items).
Data kindly provided by Dr Lygre (Publication: Lygre SH, Espehaug B, Havelin LI, Vollset SE, Furnes O. Does patella resurfacing really matter? Pain and function in 972 patients after primary total knee arthroplasty. An observational study from the Norwegian Arthroplasty Register. Acta Orthop. 2010; 81 (1): 99-107.)
Weighted mean = ((mean resurfaced *184)+(mean non-resurfaced*180))/364.
e.motion®-Prosthesis Followed Up for 5.6 Years. Summary Data for KSS-Total Score
| Interval | N | Mean | STD | Minimum | Maximum |
|---|---|---|---|---|---|
| FU0: preOP | 28 | 70.18 | 23.30 | 19.00 | 111.00 |
| FU1: 6Months | 25 | 141.60 | 35.50 | 52.00 | 199.00 |
| FU2: 1 Year | 13 | 150.00 | 35.35 | 79.00 | 198.00 |
| FU3: 2-3 Years | 25 | 166.48 | 38.56 | 45.00 | 199.00 |
| FU4: 5-6 Years | 18 | 160.33 | 31.17 | 99.00 | 198.00 |