Literature DB >> 23726233

Utility of a microwave surgical instrument in sealing lymphatic vessels.

Katsushi Takebayashi1, Hisanori Shiomi, Shigeyuki Naka, Hiroyuki Murayama, Koichiro Murakami, Hiroya Akabori, Tsuyoshi Yamaguchi, Tomoharu Shimizu, Satoshi Murata, Hiroshi Yamamoto, Yoshimasa Kurumi, Tohru Tani.   

Abstract

BACKGROUND: This study assessed the ability of a novel microwave coagulation surgical instrument (MWCX) to seal lymphatic vessels when compared with LigaSure (Valleylab, Boulder, CO), the Harmonic Scalpel (HS; Ethicon Endo-Surgery, Cincinnati, OH), and electric cautery.
METHODS: The burst pressure of pig inguinal lymphatic vessels was assessed after the sealing of vessels with each surgical instrument. The rate of lymphorrhea from pig mesenteric lymphatic vessels was also investigated using indocyanine green and visualized with the Photodynamic Eye system (Hamamatsu Hotoniks, Hamamatsu, Japan).
RESULTS: Burst pressures were higher with MWCX (average, 300 mm Hg), LigaSure (average, 290 mm Hg), and HS (average, 253 mm Hg) when compared with electric cautery (average, 152.3 mm Hg; vs MWCX: P = .002, vs LigaSure: P = .002, vs HS: P = .004). The rate of lymphorrhea was significantly lower with LigaSure (13.3%), HS (18.8%), and MWCX (13.3%) when compared with electric cautery (77.3%; vs LigaSure: P < .001, vs HS: P < .001, vs MWCX: P < .001).
CONCLUSIONS: MWCX was equivalent to LigaSure and HS in terms of the ability to seal lymphatic vessels.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  MWCX; Microwave; Vessel sealing device

Mesh:

Year:  2013        PMID: 23726233     DOI: 10.1016/j.amjsurg.2012.07.048

Source DB:  PubMed          Journal:  Am J Surg        ISSN: 0002-9610            Impact factor:   2.565


  4 in total

1.  Video-Assisted Thyroidectomy Using a Surgical Energy Device: Initial Experience in a Japanese Single-Center Cohort.

Authors:  Jun-Ichi Ohkubo; Tetsuro Wakasugi; Shoko Takeuchi; Shoichi Hasegawa; Azusa Takahashi; Hideaki Suzuki
Journal:  Biomed Hub       Date:  2021-12-10

2.  The invention of microwave surgical scissors for seamless coagulation and cutting.

Authors:  Tohru Tani; Shigeyuki Naka; Soichiro Tani; Hisanori Shiomi; Koichiro Murakami; Atsushi Yamada; Dang Tran Khiem
Journal:  Surg Today       Date:  2018-05-10       Impact factor: 2.549

3.  Bipolar Sealing Device Use in Pancreas Graft Preparation: A Novel Tieless Backtable Surgery Technique.

Authors:  Paulo N Martins; Andrew R Gillooly; Babak Movahedi; Adel Bozorgzadeh
Journal:  Transplant Direct       Date:  2018-09-25

4.  Modern lymphatic dissection techniques for preventing post renal transplant lymphocele.

Authors:  Ciprian Valerian Lucan; Ionut Jurchis; Mihai Suciu; Sonia Emilia Selicean; Salvatore Buttice
Journal:  Clujul Med       Date:  2017-10-20
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.