Literature DB >> 23725742

PIP silicone breast implants: rupture rates based on the explantation of 676 implants in a single surgeon series.

Omar Quaba1, Awf Quaba.   

Abstract

INTRODUCTION: To determine the true rupture rates of PIP implants from a large single surgeon cohort and to assess whether rupture rates varied depending on time of implant insertion. In addition, the efficacy of ultra sound scanning (USS) in determining rupture is examined.
DESIGN: Predominantly prospectively based analysis of patient records, investigations and surgical findings. PARTICIPANTS: 338 patients (676 implants) were included in the study and they all had removal of their implants. The senior author operated on all patients at some stage of their treatment. 160 patients were imaged pre-operatively with USS. Patients had implants inserted between 1999 and 2007 for cosmetic breast augmentation.
RESULTS: A total of 144 ruptured implants were removed from 119 patients, giving a rupture rate of 35.2% per patient and 21.3% per implant over a mean implantation period of 7.8 years. A statistical difference (P < 0.001) in rupture rates between implants inserted prior to 2003 and those inserted from 2003 was demonstrated, with higher failure rates in the latter group. There was a significant difference in rupture rates depending on pocket placement of the implants. The sensitivity and specificity of USS at detecting rupture was 90.6% and 98.3% respectively. A proportion of patients (29.4%) demonstrated loco-regional spread of silicone to the axilla on scanning.
CONCLUSIONS: Our paper has confirmed high rates of PIP implant failure in the largest published series to date. The significant difference in rupture rates between implants inserted prior to 2003 and those after this time supports the view that industrial silicone was used in the devices after 2003. Implants are more likely to rupture if inserted in the sub muscular plane compared to the sub glandular plane. USS is highly effective at detecting rupture in PIP implants and loco-regional spread is high compared to other devices. We believe this paper provides hard data enabling more informed decision making for patients, clinicians and providers in what remains an active issue affecting thousands of women.
Copyright © 2013 British Association of Plastic, Reconstructive and Aesthetic Surgeons. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  PIP breast implants; Silicone breast implant rupture

Mesh:

Substances:

Year:  2013        PMID: 23725742     DOI: 10.1016/j.bjps.2013.05.003

Source DB:  PubMed          Journal:  J Plast Reconstr Aesthet Surg        ISSN: 1748-6815            Impact factor:   2.740


  7 in total

1.  PIP breast implants: rupture rate and correlation with breast cancer.

Authors:  M Moschetta; M Telegrafo; I Cornacchia; L Vincenti; V Ranieri; A Cirili; L Rella; A A Stabile Ianora; G Angelelli
Journal:  G Chir       Date:  2014 Nov-Dec

2.  Poly Implant Prothèse (PIP) incidence of rupture: a retrospective MR analysis in 64 patients.

Authors:  Mariella Scotto di Santolo; Bianca Cusati; Alfonso Ragozzino; Nicoletta Dell'Aprovitola; Alessandra Acquaviva; Michele Altiero; Antonello Accurso; Albina Riccardi; Massimo Imbriaco
Journal:  Quant Imaging Med Surg       Date:  2014-12

Review 3.  The clinical implications of poly implant prothèse breast implants: an overview.

Authors:  Umar Wazir; Abdul Kasem; Kefah Mokbel
Journal:  Arch Plast Surg       Date:  2015-01-14

4.  Thoracic outlet syndrome following breast implant rupture.

Authors:  Raakhi Mistry; Yugesh Caplash; Pratyush Giri; Daniel Kearney; Marcus Wagstaff
Journal:  Plast Reconstr Surg Glob Open       Date:  2015-04-07

5.  Soft, Brown Rupture: Clinical Signs and Symptoms Associated with Ruptured PIP Breast Implants.

Authors:  Yvette Godwin; Robert T Duncan; Christine Feig; Michelle Reintals; Sarah Hill
Journal:  Plast Reconstr Surg Glob Open       Date:  2014-12-05

6.  Recurrent 8-Year Ongoing Unilateral Breast Seroma Formation after PIP Implant Removal-A Case Report and Review of the Literature.

Authors:  Sonja Kästner; Felix Julian Paprottka; Phillipp Gonser; Manuel De Villegas López; Kai Oliver Kaye
Journal:  Surg J (N Y)       Date:  2018-03-26

7.  Poly Implant Prothèse silicone breast implants: implant dynamics and capsular contracture.

Authors:  Y Bachour; Z C M Heinze; T S Dormaar; W G van Selms; M J P F Ritt; F B Niessen
Journal:  Eur J Plast Surg       Date:  2018-06-01
  7 in total