Improving detection of mild loss of retinal light increment sensitivity at the posterior pole with the microperimeter MP1.

Wadim Bowl1, Birgit Lorenz, Melanie Jäger, Christoph Friedburg.   

Abstract

PURPOSE: With "standard" stimuli (white, Goldmann size III, 200 ms), the Nidek Microperimeter MP1 underestimates retinal light increment sensitivity (LIS). We thoroughly analyze this problem, suggest alternative settings to improve sensitivity to detect dysfunction, and provide true normal values.
METHODS: LIS was tested at 55 positions in the macular region using a 4-2-1 staircase strategy with 200 ms white or red stimuli on a 1.3 cd m⁻² background. Stimulus size was Goldmann III and I, and additionally II in the healthy subjects. All participants underwent a complete ophthalmologic examination, spectral domain optical coherence tomography (OCT), and fundus autofluorescence (FAF).
RESULTS: In normals, distributions of LIS for white Goldmann sizes II and III within the central 6° to 10° were clipped off at 20 dB--the MP1 cannot attenuate them any further. When the stimulus size was reduced to Goldmann I or the color changed to red, median LIS in the fovea (∼15 dB) was approximately 5 dB higher than at 10° eccentricity. Estimated from these results, central LIS for white Goldmann sizes II and III stimuli were 21 and 27 dB, respectively. In four patients with either focal or diffuse macular pathology, as confirmed by funduscopy, OCT, or FAF, reduced LIS was detected clearly with Goldmann size I stimuli, but not III.
CONCLUSIONS: In all subjects reported here, standard central LIS was above the technical limit of the MP1. To measure true thresholds in healthy subjects, either smaller (Goldmann size I) or dimmer stimuli (red) must be used.

Entities:  

Keywords:  macula; microperimetry; perimetry; retinal light increment sensitivity; spatial summation

Mesh:

Year:  2013        PMID: 23716630     DOI: 10.1167/iovs.12-11241

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  3 in total

1.  Fundus-controlled two-color dark adaptometry with the Microperimeter MP1.

Authors:  Wadim Bowl; Knut Stieger; Birgit Lorenz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-04-24       Impact factor: 3.117

Review 2.  Role of microperimetry in evaluating disease progression in age-related macular degeneration: a scoping review.

Authors:  Gopinath Madheswaran; Pinaz Nasim; Shonraj Ballae Ganeshrao; Rajiv Raman; Ramesh S Ve
Journal:  Int Ophthalmol       Date:  2022-01-07       Impact factor: 2.029

3.  Unexpected Genetic Cause in Two Female Siblings with High Myopia and Reduced Visual Acuity.

Authors:  M N Preising; C Friedburg; W Bowl; B Lorenz
Journal:  Biomed Res Int       Date:  2018-05-23       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.