Intraocular pressure of supine patients using four portable tonometers.

Shunsuke Nakakura1, Etsuko Mori, Minami Yamamoto, Yuuri Tsushima, Hitoshi Tabuchi, Yoshiaki Kiuchi.   

Abstract

PURPOSE: To evaluate the congruity of intraocular pressure (IOP) measurements from supine patients, which were obtained using four portable tonometers.
METHODS: Intraocular pressure measurements were obtained from the right eye of 72 supine patients. We used the iCare (Tiolat Oy, Helsinki, Finland) rebound tonometer, the Diaton (BICOM Inc., Long Beach, NY) transpalpebral tonometer, the Tonopen XL (Reichert inc., Depew, NY), and a Kowa hand-held applanation tonometer (HAT; Kowa Company, Ltd., Nagoya, Japan). Relationships between mean IOPs were evaluated using Pearson correlation coefficients, and the mean differences between tonometers, using one-way analysis of variance followed by Tukey-Kramer post-hoc analysis. Levels of agreement were evaluated using Bland-Altman analysis.
RESULTS: The mean IOPs (mean ± SD) were 18.2 ± 3.5 mm Hg for iCare, 14.8 ± 3.4 mm Hg for Diaton, 16.7 ± 3.7 mm Hg for Tonopen XL, and 16.8 ± 2.8 mm Hg for Kowa HAT. Pearson correlation coefficients between iCare, Tonopen XL, and Kowa HAT ranged from 0.382 to 0.577, whereas those between Diaton and other tonometers ranged from 0.041 to 0.286. Post-hoc analysis indicated significant differences between all pairs except Tonopen XL and Kowa HAT. The mean difference between measurements from iCare and Diaton was 3.39 ± 3.39 mm Hg; iCare and Tonopen XL, 1.47 ± 3.52 mm Hg; iCare and Kowa HAT, 1.49 ± 2.90 mm Hg; Diaton and Tonopen XL, -1.93 ± 4.90 mm Hg; Diaton and Kowa HAT, -1.90 ± 4.15 mm Hg; and Tonopen XL and Kowa HAT, 0.02 ± 3.61 mm Hg. Computation of the width of the 95% limits of agreement resulted in a wide bias range when comparing Diaton with all tonometers. Relatively good agreements were observed between iCare, Tonopen XL, and HAT.
CONCLUSIONS: Intraocular pressure measurements obtained in a supine position by four portable tonometers were not interchangeable. Although iCare and Tonopen XL significantly overestimated IOP values in eyes with a higher IOP when compared with Kowa HAT, the agreements between iCare, Tonopen XL, and Kowa HAT were at clinically acceptable levels.

Entities:  

Mesh:

Year:  2013        PMID: 23708927     DOI: 10.1097/OPX.0b013e3182972df4

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  7 in total

1.  Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects.

Authors:  Yoshitake Kato; Shunsuke Nakakura; Naoko Matsuo; Kayo Yoshitomi; Marina Handa; Hitoshi Tabuchi; Yoshiaki Kiuchi
Journal:  Int Ophthalmol       Date:  2017-04-09       Impact factor: 2.031

2.  Utility of the Tono-Pen in Measuring Intraocular Pressure in Trinidad: A Cross-sectional Study.

Authors:  A Billy; P E David; A K Mahabir; C P Seerattan; J M Street; V D Walcott; R J Yarna; D C Murray; R G Maharaj
Journal:  West Indian Med J       Date:  2015-05-12       Impact factor: 0.171

3.  Intraocular Pressure Measurement in Childhood Glaucoma under Standardized General Anaesthesia: The Prospective EyeBIS Study.

Authors:  Alicja Strzalkowska; Nina Pirlich; Julia V Stingl; Alexander K Schuster; Jasmin Rezapour; Felix M Wagner; Justus Buse; Esther M Hoffmann
Journal:  J Clin Med       Date:  2022-05-18       Impact factor: 4.964

4.  Change in intraocular pressure during point-of-care ultrasound.

Authors:  Cameron Berg; Stephanie J Doniger; Brita Zaia; Sarah R Williams
Journal:  West J Emerg Med       Date:  2015-03-06

5.  A Comparison of Rebound to Indentation Tonometry in Supine Sedated Children with Glaucoma.

Authors:  Dora H AlHarkan; Fatemah T Al-Shamlan; Deepak P Edward; Arif O Khan
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Apr-Jun

6.  Reproducibility of optic disk evaluation in supine subjects with a Heidelberg Retina Tomograph II laser tomographic scanner.

Authors:  Yosuke Harada; Tomoyuki Akita; Joji Takenaka; Yuko Nakamura-Kadohiro; Junko Tanaka; Yoshiaki Kiuchi
Journal:  Clin Ophthalmol       Date:  2016-08-24

Review 7.  Icare® rebound tonometers: review of their characteristics and ease of use.

Authors:  Shunsuke Nakakura
Journal:  Clin Ophthalmol       Date:  2018-07-12
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.