Porcine carotid artery replacement with biodegradable electrospun poly-e-caprolactone vascular prosthesis.

Wojciech Mrówczyński1, Damiano Mugnai2, Sarra de Valence3, Jean-Christophe Tille4, Ebrahim Khabiri2, Mustafa Cikirikcioglu2, Michael Möller3, Beat H Walpoth5.   

Abstract

OBJECTIVE: There is a continuous search for shelf-ready small-caliber vascular prostheses with satisfactory early and late results. Biodegradable scaffolds, repopulated by recipient's cells regenerating a neovessel, can be a suitable option for adult and pediatric, urgent and elective cardiovascular procedures.
METHODS: This was a short-term experimental assessment of a new biodegradable vascular prosthesis for arterial replacement in the pig. Eleven pigs underwent bilateral carotid artery replacement with biodegradable electrospun poly-ε-caprolactone (PCL) nanofiber prostheses (internal diameter, 4 mm; length, 5 cm); or expanded polytetrafluoroethylene (ePTFE) prostheses as control. Perioperative anticoagulation was achieved with intravenous heparin (double baseline activated clotting time). Postoperatively, until conclusion of the study at 1 month, animals received aspirin and clopidogrel daily. Transit time flow was measured intraoperatively and at sacrifice. Doppler ultrasound (1 and 4 weeks) and a selective carotid angiography (4 weeks) were performed to assess patency. All explanted grafts were analyzed by histology, morphometry, and scanning electron microscopy in order to study graft-host interaction.
RESULTS: Surgical handling and hemostasis of the new prostheses were excellent. Patency rate was 78% (7/9) for PCL grafts, compared with 67% (4/6) for ePTFE grafts. Transit time flow and Doppler ultrasound showed no significant changes in flow and velocity or diameter over time in both groups. Both prostheses showed no detectable in vivo compliance as compared with native carotid artery. Percent neoendothelialization was 86% for PCL and 58% for ePTFE grafts (P = .008). Neointima formation was equal in both grafts. More adventitial infiltration of macrophages, myofibroblasts, and capillaries was seen in PCL grafts with a milder foreign-body reaction when compared with ePTFE implants. Both grafts showed similar endoluminal thrombus formation.
CONCLUSIONS: Biodegradable, electrospun PCL grafts showed good surgical and mechanical properties, no aneurysm formation, and similar short-term patency compared with ePTFE grafts. Rapid endothelialization and cell ingrowth confirms favorable PCL graft-recipient biological interaction. Despite good early results, long-term follow-up is required before clinical application.
Copyright © 2014 Society for Vascular Surgery. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23707057     DOI: 10.1016/j.jvs.2013.03.004

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  21 in total

1.  Differential outcomes of venous and arterial tissue engineered vascular grafts highlight the importance of coupling long-term implantation studies with computational modeling.

Authors:  Cameron A Best; Jason M Szafron; Kevin A Rocco; Jacob Zbinden; Ethan W Dean; Mark W Maxfield; Hirotsugu Kurobe; Shuhei Tara; Paul S Bagi; Brooks V Udelsman; Ramak Khosravi; Tai Yi; Toshiharu Shinoka; Jay D Humphrey; Christopher K Breuer
Journal:  Acta Biomater       Date:  2019-06-12       Impact factor: 8.947

2.  [Establishment of a rabbit model of small diameter vascular graft replacement].

Authors:  Jia-Qing Zhang; Kun-Tang Chen; Fu-Wei Zhang; Shao-Bin Li; Yuan-Zhou Wu; Jing Feng; Wu-Jun Wang; Yu-Sheng Yan
Journal:  Nan Fang Yi Ke Da Xue Xue Bao       Date:  2017-05-20

3.  Electrospun biodegradable elastic polyurethane scaffolds with dipyridamole release for small diameter vascular grafts.

Authors:  Primana Punnakitikashem; Danh Truong; Jyothi U Menon; Kytai T Nguyen; Yi Hong
Journal:  Acta Biomater       Date:  2014-08-08       Impact factor: 8.947

4.  Preclinical study of patient-specific cell-free nanofiber tissue-engineered vascular grafts using 3-dimensional printing in a sheep model.

Authors:  Takuma Fukunishi; Cameron A Best; Tadahisa Sugiura; Justin Opfermann; Chin Siang Ong; Toshiharu Shinoka; Christopher K Breuer; Axel Krieger; Jed Johnson; Narutoshi Hibino
Journal:  J Thorac Cardiovasc Surg       Date:  2016-11-14       Impact factor: 5.209

5.  Well-organized neointima of large-pore poly(L-lactic acid) vascular graft coated with poly(L-lactic-co-ε-caprolactone) prevents calcific deposition compared to small-pore electrospun poly(L-lactic acid) graft in a mouse aortic implantation model.

Authors:  Shuhei Tara; Hirotsugu Kurobe; Kevin A Rocco; Mark W Maxfield; Cameron A Best; Tai Yi; Yuji Naito; Christopher K Breuer; Toshiharu Shinoka
Journal:  Atherosclerosis       Date:  2014-10-17       Impact factor: 5.162

6.  Fast-degrading bioresorbable arterial vascular graft with high cellular infiltration inhibits calcification of the graft.

Authors:  Tadahisa Sugiura; Shuhei Tara; Hidetaka Nakayama; Tai Yi; Yong-Ung Lee; Toshihiro Shoji; Christopher K Breuer; Toshiharu Shinoka
Journal:  J Vasc Surg       Date:  2016-09-26       Impact factor: 4.268

Review 7.  Elastomers in vascular tissue engineering.

Authors:  Matti A Hiob; Gareth W Crouch; Anthony S Weiss
Journal:  Curr Opin Biotechnol       Date:  2016-05-02       Impact factor: 9.740

Review 8.  Quickening: Translational design of resorbable synthetic vascular grafts.

Authors:  Chelsea E T Stowell; Yadong Wang
Journal:  Biomaterials       Date:  2018-05-05       Impact factor: 12.479

Review 9.  Biofabrication of tissue engineering vascular systems.

Authors:  Qiao Zhang; Èlia Bosch-Rué; Román A Pérez; George A Truskey
Journal:  APL Bioeng       Date:  2021-05-07

10.  Fabrication of Highly Oriented Cylindrical Polyacrylonitrile, Poly(lactide-co-glycolide), Polycaprolactone and Poly(vinyl acetate) Nanofibers for Vascular Graft Applications.

Authors:  Sairish Malik; Subramanian Sundarrajan; Tanveer Hussain; Ahsan Nazir; Seeram Ramakrishna
Journal:  Polymers (Basel)       Date:  2021-06-24       Impact factor: 4.329

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.