Literature DB >> 23690134

Reproducibility of current perception threshold with the Neurometer(®) vs the Stimpod NMS450 peripheral nerve stimulator in healthy volunteers: an observational study.

Ban C H Tsui1, Timothy J Shakespeare, Danika H Leung, Jeremy H Tsui, Gareth N Corry.   

Abstract

PURPOSE: Current methods of assessing nerve blocks, such as loss of perception to cold sensation, are subjective at best. Transcutaneous nerve stimulation is an alternative method that has previously been used to measure the current perception threshold (CPT) in individuals with neuropathic conditions, and various devices to measure CPT are commercially available. Nevertheless, the device must provide reproducible results to be used as an objective tool for assessing nerve blocks.
METHODS: We recruited ten healthy volunteers to examine CPT reproducibility using the Neurometer(®) and the Stimpod NMS450 peripheral nerve stimulator. Each subject's CPT was determined for the median (second digit) and ulnar (fifth digit) nerve sensory distributions on both hands - with the Neurometer at 5 Hz, 250 Hz, and 2000 Hz and with the Stimpod at pulse widths of 0.1 msec, 0.3 msec, 0.5 msec, and 1.0 msec, both at 5 Hz and 2 Hz. Intraclass correlation coefficients (ICC) were also calculated to assess reproducibility; acceptable ICCs were defined as ≥ 0.4.
RESULTS: The ICC values for the Stimpod ranged from 0.425-0.79, depending on pulse width, digit, and stimulation; ICCs for the Neurometer were 0.615 and 0.735 at 250 and 2,000 Hz, respectively. These values were considered acceptable; however, the Neurometer performed less efficiently at 5 Hz (ICCs for the second and fifth digits were 0.292 and 0.318, respectively).
CONCLUSION: Overall, the Stimpod device displayed good to excellent reproducibility in measuring CPT in healthy volunteers. The Neurometer displayed poor reproducibility at low frequency (5 Hz). These results suggest that peripheral nerve stimulators may be potential devices for measuring CPT to assess nerve blocks.

Entities:  

Mesh:

Year:  2013        PMID: 23690134     DOI: 10.1007/s12630-013-9965-z

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  5 in total

1.  The technique comparison of brachial plexus blocks by ultrasound guided with blocks by nerve stimulator guided.

Authors:  Guo-Ying Liu; Zhen-Qiang Chen; Hai-Yan Jia; Zhi-Gang Dai; Xue-Jun Zhang
Journal:  Int J Clin Exp Med       Date:  2015-09-15

Review 2.  Distal Sensorimotor Neuropathy: Improvements in Diagnosis.

Authors:  Prashanth R J Vas; Sanjeev Sharma; Gerry Rayman
Journal:  Rev Diabet Stud       Date:  2015-08-10

3.  Thermal quantitative sensory testing in healthy Dutch children and adolescents standardized test paradigm and Dutch reference values.

Authors:  Gerbrich E van den Bosch; Monique van Dijk; Dick Tibboel; Abraham J Valkenburg
Journal:  BMC Pediatr       Date:  2017-03-16       Impact factor: 2.125

4.  Both ipsilateral and contralateral localized vibratory stimulations modulated pain-related sensory thresholds on the foot in mice and humans.

Authors:  Atsushi Doi; Juntaro Sakasaki; Chikato Tokunaga; Fumiya Sugita; Syota Kasae; Keisuke Nishimura; Yushi Sato; Takako Kuratsu; Sariya Hashiguchi; Min-Chul Shin; Megumu Yoshimura
Journal:  J Pain Res       Date:  2018-08-28       Impact factor: 3.133

5.  Supplementation with Korean Red Ginseng Improves Current Perception Threshold in Korean Type 2 Diabetes Patients: A Randomized, Double-Blind, Placebo-Controlled Trial.

Authors:  Kahui Park; YuSik Kim; Junghye Kim; Shinae Kang; Jong Suk Park; Chul Woo Ahn; Ji Sun Nam
Journal:  J Diabetes Res       Date:  2020-01-16       Impact factor: 4.011

  5 in total