Literature DB >> 23683525

Revision of ceramic head fracture after third generation ceramic-on-ceramic total hip arthroplasty.

Kyung-Hoi Koo1, Yong-Chan Ha, Shin-Yoon Kim, Kang-Sup Yoon, Byung-Woo Min, Sang-Rim Kim.   

Abstract

We performed 24 revisions of fractures of third generation ceramic heads. The stem was not changed in 20 revisions; a new ceramic-on-ceramic bearing was used in four and a metal-on-polyethylene bearing in 16. The stem was changed in four revisions; a new ceramic-on-ceramic bearing was used in three and a metal-on-polyethylene bearing in one. During the follow-up of 57.5 months, complications occurred in five hips among the 20 stem retained revisions: a fracture of the new ceramic head in two, metallosis with pseudocyst in two, and femoral osteolysis with stem loosening in one. However, there were no complications in the four revisions where the stem was changed. Revision surgery after ceramic head fracture shows high rates of early complications. We recommend stem revision in cases of THA failure due to fracture of a modern ceramic head.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  arthroplasty; ceramic; fracture; hip; revision

Mesh:

Substances:

Year:  2013        PMID: 23683525     DOI: 10.1016/j.arth.2013.04.007

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Experience with sandwich liner and its high rate of failure.

Authors:  John Junghun Shin; Ji-Soo Shin; Jae-Hyun Kim; Sung Kwan Hwang
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-07-16

2.  Navigated Acetabular Cup Fixation for Acetabular Deformity or Revision Total Hip Arthroplasty.

Authors:  Ho Hyun Yun; Jung-Ro Yoon; Jung Jin Yu; Hyo-Sung Seo
Journal:  Hip Pelvis       Date:  2014-09-29

3.  Fourth-Generation Ceramic Head Fracture in Dual Mobility Total Hip Arthroplasty: A Case Report.

Authors:  Michael J Patetta; Erdan Kayupov; Joseph A Karam; Samuel J Chmell
Journal:  J Orthop Case Rep       Date:  2022-01

4.  Do the Reasons for Ceramic-on-ceramic Revisions Differ From Other Bearings in Total Hip Arthroplasty?

Authors:  Henri Migaud; Sophie Putman; Grégory Kern; Ronald Isida; Julien Girard; Nassima Ramdane; Christian P Delaunay; Moussa Hamadouche
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

5.  Comparison of the risk of revision in cementless total hip arthroplasty with ceramic-on-ceramic and metal-on-polyethylene bearings.

Authors:  Claus Varnum; Alma B Pedersen; Per Kjærsgaard-Andersen; Søren Overgaard
Journal:  Acta Orthop       Date:  2015-01-30       Impact factor: 3.717

6.  Long Term Survivorship of a Severely Notched Femoral Stem after Replacing the Fractured Ceramic head with a Cobalt-Chromium Head.

Authors:  Andreas Panagopoulos; Irini Tatani; Panagiotis Megas
Journal:  Open Orthop J       Date:  2016-12-30

7.  Adverse local tissue reaction after 2 revision hip replacements for ceramic liner fracture: A case report.

Authors:  Iulian Antoniac; Mihai Negrusoiu; Mihai Mardare; Claudiu Socoliuc; Ancuţa Zazgyva; Marius Niculescu
Journal:  Medicine (Baltimore)       Date:  2017-05       Impact factor: 1.889

Review 8.  Ceramic Toughening Strategies for Biomedical Applications.

Authors:  Rushui Bai; Qiannan Sun; Ying He; Liying Peng; Yunfan Zhang; Lingyun Zhang; Wenhsuan Lu; Jingjing Deng; Zimeng Zhuang; Tingting Yu; Yan Wei
Journal:  Front Bioeng Biotechnol       Date:  2022-03-07
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.