Literature DB >> 23681810

Continuous-flow liquid microjunction surface sampling probe connected on-line with high-performance liquid chromatography/mass spectrometry for spatially resolved analysis of small molecules and proteins.

Gary J Van Berkel1, Vilmos Kertesz.   

Abstract

RATIONALE: A continuous-flow liquid microjunction surface sampling probe extracts soluble material from surfaces for direct ionization and detection by mass spectrometry. Demonstrated here is the on-line coupling of such a probe with high-performance liquid chromatography/mass spectrometry (HPLC/MS) enabling extraction, separation and detection of small molecules and proteins from surfaces in a spatially resolved (~0.5 mm diameter spots) manner.
METHODS: A continuous-flow liquid microjunction surface sampling probe was connected to a six-port, two-position valve for extract collection and injection to an HPLC column. A QTRAP® 5500 hybrid triple quadrupole linear ion trap equipped with a Turbo V™ ion source operated in positive electrospray ionization (ESI) mode was used for all experiments. The system operation was tested with the extraction, separation and detection of propranolol and associated metabolites from drug dosed tissues, caffeine from a coffee bean, cocaine from paper currency, and proteins from dried sheep blood spots on paper.
RESULTS: Confirmed in the tissue were the parent drug and two different hydroxypropranolol glucuronides. The mass spectrometric response for these compounds from different locations in the liver showed an increase with increasing extraction time (5, 20 and 40 s). For on-line separation and detection/identification of extracted proteins from dried sheep blood spots, two major protein peaks dominated the chromatogram and could be correlated with the expected masses for the hemoglobin α and β chains.
CONCLUSIONS: Spatially resolved sampling, separation, and detection of small molecules and proteins from surfaces can be accomplished using a continuous-flow liquid microjunction surface sampling probe coupled on-line with HPLC/MS detection. Published in 2013. This article is a U.S. Government work and is in the public domain in the USA.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23681810     DOI: 10.1002/rcm.6580

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  24 in total

1.  Laser Ablation with Vacuum Capture for MALDI Mass Spectrometry of Tissue.

Authors:  Fabrizio Donnarumma; Fan Cao; Kermit K Murray
Journal:  J Am Soc Mass Spectrom       Date:  2015-09-15       Impact factor: 3.109

2.  High Resolution Tissue Imaging Using the Single-probe Mass Spectrometry under Ambient Conditions.

Authors:  Wei Rao; Ning Pan; Zhibo Yang
Journal:  J Am Soc Mass Spectrom       Date:  2015-03-25       Impact factor: 3.109

Review 3.  Droplet probe: coupling chromatography to the in situ evaluation of the chemistry of nature.

Authors:  Nicholas H Oberlies; Sonja L Knowles; Chiraz Soumia M Amrine; Diana Kao; Vilmos Kertesz; Huzefa A Raja
Journal:  Nat Prod Rep       Date:  2019-05-21       Impact factor: 13.423

Review 4.  Spatially resolved absolute quantitation in thin tissue by mass spectrometry.

Authors:  Vilmos Kertesz; John F Cahill
Journal:  Anal Bioanal Chem       Date:  2021-04       Impact factor: 4.142

5.  Protein identification in imaging mass spectrometry through spatially targeted liquid micro-extractions.

Authors:  Daniel J Ryan; David Nei; Boone M Prentice; Kristie L Rose; Richard M Caprioli; Jeffrey M Spraggins
Journal:  Rapid Commun Mass Spectrom       Date:  2018-03-15       Impact factor: 2.419

6.  MicroLESA: Integrating Autofluorescence Microscopy, In Situ Micro-Digestions, and Liquid Extraction Surface Analysis for High Spatial Resolution Targeted Proteomic Studies.

Authors:  Daniel J Ryan; Nathan Heath Patterson; Nicole E Putnam; Aimee D Wilde; Andy Weiss; William J Perry; James E Cassat; Eric P Skaar; Richard M Caprioli; Jeffrey M Spraggins
Journal:  Anal Chem       Date:  2019-05-31       Impact factor: 6.986

7.  Clinical Application of Ambient Ionization Mass Spectrometry.

Authors:  Li-Hua Li; Hua-Yi Hsieh; Cheng-Chih Hsu
Journal:  Mass Spectrom (Tokyo)       Date:  2017-02-24

Review 8.  Protein identification strategies in MALDI imaging mass spectrometry: a brief review.

Authors:  Daniel J Ryan; Jeffrey M Spraggins; Richard M Caprioli
Journal:  Curr Opin Chem Biol       Date:  2018-11-23       Impact factor: 8.822

9.  Use of an Open Port Sampling Interface Coupled to Electrospray Ionization for the On-Line Analysis of Organic Aerosol Particles.

Authors:  Kenneth D Swanson; Anne L Worth; Gary L Glish
Journal:  J Am Soc Mass Spectrom       Date:  2017-09-11       Impact factor: 3.109

Review 10.  Emerging trends in paper spray mass spectrometry: Microsampling, storage, direct analysis, and applications.

Authors:  Benjamin S Frey; Deidre E Damon; Abraham K Badu-Tawiah
Journal:  Mass Spectrom Rev       Date:  2019-09-06       Impact factor: 10.946

View more