Literature DB >> 23670506

Experimental study on the efficiency and safety of the manual hyperinflation maneuver as a secretion clearance technique.

Tatiana de Arruda Ortiz1, Germano Forti, Márcia Souza Volpe, Carlos Roberto Ribeiro Carvalho, Marcelo Brito Passos Amato, Mauro Roberto Tucci.   

Abstract

OBJECTIVE: To evaluate, in a lung model simulating a mechanically ventilated patient, the efficiency and safety of the manual hyperinflation (MH) maneuver as a means of removing pulmonary secretions.
METHODS: Eight respiratory therapists (RTs) were asked to use a self-inflating manual resuscitator on a lung model to perform MH as if to remove secretions, under two conditions: as routinely applied during their clinical practice; and after receiving verbal instructions based on expert recommendations. In both conditions, three clinical scenarios were simulated: normal lung function, restrictive lung disease, and obstructive lung disease.
RESULTS: Before instruction, it was common for an RT to compress the resuscitator bag two times, in rapid succession. Proximal pressure (Pprox) was higher before instruction than after. However, alveolar pressure (Palv) never exceeded 42.5 cmH₂O (median, 16.1; interquartile range [IQR], 11.7-24.5), despite Pprox values as high as 96.6 cmH₂O (median, 36.7; IQR, 22.9-49.4). The tidal volume (VT) generated was relatively low (median, 640 mL; IQR, 505-735), and peak inspiratory flow (PIF) often exceeded peak expiratory flow (PEF), the median values being 1.37 L/s (IQR, 0.99-1.90) and 1.01 L/s (IQR, 0.55-1.28), respectively. A PIF/PEF ratio < 0.9 (which theoretically favors mucus migration toward the central airways) was achieved in only 16.7% of the maneuvers.
CONCLUSIONS: Under the conditions tested, MH produced safe Palv levels despite high Pprox. However, the MH maneuver was often performed in a way that did not favor secretion removal (PIF exceeding PEF), even after instruction. The unfavorable PIF/ PEF ratio was attributable to overly rapid inflations and low VT.

Entities:  

Mesh:

Year:  2013        PMID: 23670506      PMCID: PMC4075822          DOI: 10.1590/S1806-37132013000200012

Source DB:  PubMed          Journal:  J Bras Pneumol        ISSN: 1806-3713            Impact factor:   2.624


Introduction

Manual hyperinflation (MH) is a proposed technique that is purported to promote secretion clearance and to re-expand areas of atelectasis, thereby improving lung compliance and oxygenation in patients on mechanical ventilation.( , ) Despite a lack of scientific evidence confirming its benefits on clinical outcomes,( - ) MH is often used in the ICU as a chest physiotherapy technique. The maneuver is widely accepted as effective( , ) and, in Brazil, is largely embraced as a means of removing retained secretions.( - ) The rationale for the use of the technique as an aid for secretion removal is that it simulates a cough.( ) When applied before suctioning, it theoretically moves secretions toward the central airways,( ) thus increasing the efficacy of the suctioning procedure. There has been considerable debate regarding the safety and efficiency of MH.( , ) It has been shown that the efficiency of the MH maneuver is affected by the operator and the type of manual resuscitator used,( , ) as well as by the resistance and compliance of the respiratory system,( , ) largely affecting the pressures and flows generated in the respiratory system. Under certain conditions, the use of this technique can generate high peak airway pressures, increasing the risk of barotrauma.( ) According to expert recommendations,( , , ) in order to promote secretion clearance, MH should consist of a slow, deep inspiration, an inspiratory pause, and a quick release of the resuscitation bag to promote passive exhalation with high expiratory flow rates. However, the way in which the MH maneuver is performed can vary from country to country, impeding the understanding of its effects.( , , ) Since the 1980s, there has been increasing evidence that the peak expiratory flow to peak inspiratory flow (PEF/PIF) ratio is a critical factor for the removal of lung secretions, especially in heavily sedated or paralyzed patients.( - ) In fact, more than their ratio, the difference between the two, in terms of their absolute values, seems to be the major determinant: above a given threshold, whenever the PIF exceeds the PEF, secretions migrate deeper into the lung.( ) Intuitively, respiratory therapists (RTs) have been promoting maneuvers to increase expiratory flows, analogous to those observed during coughing. However, relatively little attention has been given to their inspiratory counterpart, despite expert recommendations to apply a slow, deep inspiration during MH. The inspiratory phase of MH can be performed in many different ways, depending on the personal experience of the operator, and there is little evidence that the MH maneuver generates an adequate flow bias (i.e., with PEF exceeding PIF) or that the ultimate goal of the maneuver (i.e., improved secretion clearance) is achieved. The aim of this study was to evaluate the efficiency and safety of MH (as performed by RTs in a lung model simulating a mechanically ventilated patient) as a means of increasing secretion clearance under two different conditions: MH performed in accordance with the routine clinical practice of RTs; and MH performed in accordance with expert recommendations.( , , ) The efficiency of MH was determined through analysis of the volume and flow patterns generated in relation to the anticipated effects on secretion removal. The safety of MH was evaluated on the basis of the inspiratory pressures achieved (i.e., whether those pressures remained within safe limits during the maneuver). Our ultimate objective was to obtain a qualitative analysis of how MH is performed by experienced professionals and not to address how it is performed in Brazil.

Methods

This was an experimental study conducted in the Laboratory for Medical Research 09, specializing in Pulmonology, at the University of São Paulo School of Medicine, in the city of São Paulo, Brazil. The study was approved by the Research Ethics Committee of the University of São Paulo School of Medicine Hospital das Clínicas. All participants gave written informed consent. The study comprised two phases, evaluating the effects of MH, as performed by RTs, in promoting the clearance of pulmonary secretions. In the first phase (the pre-instruction phase), the RTs were given no explicit verbal instructions on how to apply the maneuver. In the second phase, the RTs were given instructions based on expert recommendations, as detailed below. The two phases are hereafter referred to as the pre- and post-instruction phases. All maneuvers were performed with self-inflating manual resuscitator (SPUR(r); Ambu, Ballerup, Denmark), with a capacity of 1,500 mL. The maneuver was applied by eight experienced RTs (four males and four females), with an average of 2.6 years in practice (range, 2-4 years) in ICUs in the city of São Paulo. Five of the eight had graduated from universities located within the state of São Paulo, and all had completed one of the one-year postgraduate training programs in respiratory therapy offered by university hospitals. The maneuver was applied to a mechanical model of the respiratory system known as a training and test lung (TTL 2600; Michigan Instruments, Grand Rapids, MI), connected to a tracheal tube (8.5 mm). The airway flow was measured proximally (between the manual resuscitator and proximal pressure sensor) by a pneumotach. Two pressure transducers were connected to the model: one to measure proximal pressure (Pprox-at the airway, between the flow sensor and tracheal tube); and another to record alveolar pressure (Palv). Figure 1 illustrates the experimental setup.
Figure 1

Experimental setup.

In both study phases, we altered the resistance and compliance of the lung model in order to simulate three different clinical scenarios: a patient with obstructive lung disease (resistance at 20 cmH2O/L per second and compliance at 0.08 L/cmH2O); a normal patient (resistance at 5 cmH2O/L per second and compliance at 0.05 L/cmH2O); and a patient with restrictive lung disease (resistance at 5 cmH2O/L per second and compliance at 0.025 L/cmH2O). In the pre-instruction phase (as previously mentioned), the RTs were instructed to perform MH to promote pulmonary secretion clearance in accordance with their routine clinical practice. The lung model was set according to those three clinical scenarios explained above and covered with a bed sheet, allowing RTs to sense the movement of the bellows while remaining blinded to the clinical scenario selected. After approximately eight cycles of MH, the settings were changed to simulate the next scenario (in the sequence described above) until all three scenarios had been run. In the post-instruction phase, each RT received brief verbal instructions on how to perform MH according to expert recommendations,( , , ) and all of the steps performed in the pre-instruction phase were repeated. The verbal instructions were given within the space of approximately 2 min, and the RTs were not trained to follow a given pattern of insufflation. The instructions were always given by the same researcher, who instructed each RT as follows: "Now you should perform MH with a slow inflation and a 2-s inspiratory pause, followed by rapid release of the bag". While giving the instructions, the researcher demonstrated the maneuver with the manual resuscitator used in the study. The instructions were given twice. The analog signals from the flow and pressure transducers were amplified, digitized, and recorded at 200 Hz, using a data acquisition and off-line analysis system developed within the software Laboratory Virtual Instrumentation and Engineering Workbench (LabVIEW; National Instruments, Austin, TX, USA). For each experimental condition, three of the (approximately) eight respiratory cycles were randomly selected for analysis. The beginning of the inspiratory phase was defined as zero flow immediately before bagging, and the end of the respiratory cycle was defined as zero flow at the end of exhalation. The Pprox was defined as the peak pressure value during the inspiratory phase, measured by the proximal pressure transducer. The PIF was defined as the peak flow value during the inspiratory phase, measured by the flow transducer, and the PEF was defined as the peak flow during expiratory phase. Tidal volume (VT) was calculated by integration of the inspiratory flow signal. The Palv was defined as the peak pressure value during the inspiratory phase, measured by the alveolar pressure transducer. Inspiratory time (TI) was defined as the duration of the inflation plus the inspiratory pause. Data are shown as median and 25-75% interquartile range (IQR). Repeated measures ANOVA was used to evaluate, for each variable, the following within-subject factors: the three clinical scenarios; and the two phases (before and after instructions). All two-way interactions between those factors were also tested. Inspiratory versus expiratory flow and alveolar versus proximal pressures were also tested as within-subject factors to compare peak flow and peak pressure variables, respectively. Values of p < 0.05 were considered statistically significant.

Results

For each of the six experimental conditions (three clinical scenarios in each study phase), we analyzed three respiratory cycles, corresponding to eighteen respiratory cycles for each of the eight RTs. Therefore, we analyzed a total of 144 respiratory cycles. Among all of the respiratory cycles analyzed, the maximum Palv observed was 42.5 cmH2O (median, 16.1 cmH2O; IQR, 11.7-24.5), despite the much higher Pprox values (maximum, 96.6 cmH2O-median, 36.7; IQR, 22.9-49.4). The VT values were relatively low (maximum, 955 mL-median, 640; IQR, 505-735), and the TI was short (median, 1.29 s; IQR, 0.95-1.72). Overall, PIF was significantly higher than PEF (p < 0.02). A PIF/PEF ratio < 0.9, which theoretically favors mucus migration toward the central airways,( ) was achieved in only 24 (16.7%) of the 144 maneuvers evaluated. This favorable ratio occurred primarily in the post-instruction phase (in 18 of the 24 maneuvers) and was strongly associated with just one of the RTs tested (who was responsible for 12 of the 24 maneuvers). During the pre-instruction phase, six of the eight RTs performed MH using two compressions of the resuscitator bag and produced a PIF higher than that produced in the post-instruction phase (Figure 2).
Figure 2

Differences in proximal pressures achieved by two different respiratory therapists (A and B) before and after explicit verbal instruction (routine clinical practice vs. expert recommendations-dashed lines and solid lines, respectively).

Table 1 shows a comparison between the pre- and post-instruction phases, in terms of the mechanical variables evaluated. After instruction, MH was performed in a slower manner, with a longer TI and a lower PIF, producing a lower Pprox. There were no statistically significant differences between the two phases in terms of the Palv, VT and PEF. Figure 2 illustrates the difference between the two study phases and between two different RTs.
Table 1

Mechanical variables before and after verbal instruction (routine clinical practice vs. expert recommendations).

VariablePre-instructionPost-instructionp
Proximal pressure (cmH2O)44.5 (33.4-63.4)26.8 (18.6-37.5)0.004
Alveolar pressure (cmH2O)16.3 (11.6-25.8)14.8 (11.7-23.8)0.93
Peak inspiratory flow (L/s)1.84 (1.28-2.19)1.14 (0.87-1.44)0.001
Peak expiratory flow (L/s)1.04 (0.57-1.33)0.99 (0.55-1.27)0.28
Tidal volume (mL)628 (497-699)647 (518-746)0.63
Inspiratory time (s)0.95 (0.78-1.19)1.71 (1.44-2.13)<0.001
PIF/PEF ratio1.80 (1.29-2.34)1.15 (0.87-1.80)0.004

Values are expressed as median (25-75% interquartile range)

: peak inspiratory flow

: peak expiratory flow

Values are expressed as median (25-75% interquartile range) : peak inspiratory flow : peak expiratory flow Comparing all three clinical scenarios tested and the two phases of the study, we found that Pprox was markedly higher than was Palv in all instances, with the exception of the restrictive lung disease scenario in the post-instruction phase (Figure 3). In the pre-instruction phase, PIF values were consistently higher than were PEF values. In the post-instruction phase, the PIF/PEF ratio was closer to 1:1. The only situation in which that ratio was consistently unfavorable (PIF still far exceeding PEF) was the obstructive lung disease scenario (Figure 4).
Figure 3

Median values of proximal and alveolar pressures (interquartile ranges as error bars) obtained in the pre-instruction (routine clinical practice) and post-instruction (in accordance with expert recommendations) phases of the study. The major effect of instruction was the generation of lower proximal pressures. The dashed line indicates 30 cmH2O. *p< 0.01 (difference between pre-instruction and post-instruction). #p < 0.05 (difference between proximal and alveolar pressure).

Figure 4

Median values for peak inspiratory and expiratory flows (interquartile ranges as error bars) obtained in the pre-instruction (routine clinical practice) and post-instruction (in accordance with expert recommendations) phases of the study. The major effect of instruction was the generation of lower peak inspiratory flow. The dashed line indicates 1 L/sec. *p< 0.01 (difference between pre-instruction and postinstruction). #p < 0.05 (difference between peak inspiratory and expiratory flows).

Discussion

The major finding of the present study was that, even after receiving explicit instructions, the RTs evaluated here performed MH in a way that probably would not aid secretion removal, with PIF commonly exceeding PEF. Compressing the resuscitator bag multiple times, in rapid succession, resulted in high values of PIF and Pprox. However, this finding was typically associated with a relatively low VT, probably because of short bag-compression times. Therefore, the Palv was often low at the start of exhalation, resulting in a low PEF. Both factors (rapid, multiple compressions and low Palv) appear to be responsible for the unfavorable relationship between PIF and PEF. Finally, although the MH maneuver might not promote secretion clearance, our results suggest that this maneuver, as performed in our study, is unlikely to cause barotrauma, a concern expressed by other authors because of the high Pprox it generates.( ) In the present study, we also observed high Pprox values. However, the Palv values-roughly represented by plateau pressures, which correlate better with barotrauma than do Pprox values-were within the safe range. The MH maneuver was originally described as consisting of a slow, deep inspiration, with an inspiratory pause and rapid release of the resuscitator bag.( ) Since then, many different MH techniques have been described, including one providing a VT that is greater than the baseline VT for the patient in question( ); one providing a VT that is 50% greater than that delivered by the ventilator( ); and one consisting of a slow (3-s) inspiration to achieve a peak airway pressure of 40 cmH2O.( ) In contrast, we found that RTs working in the city of São Paulo have customized the maneuver according to personal practice and bias, frequently applying two compressions of the resuscitator bag, without an inspiratory pause, resulting in PIF being higher than PEF. One possible explanation is that the maneuver applied in that way stimulates coughing and, consequently, improves secretion clearance, or at least gives the RT that impression. As observed in this study, however, applying the maneuver in that way might make it ineffective, with unfavorable relationships between inspiratory and expiratory flows, especially if the patient has a depressed cough reflex or is unable to cough efficiently. According to previous studies,( , ) above a given flow threshold, the direction of mucus transport (in or out of the respiratory system) is governed by the highest peak flow: a PEF 10% higher than the PIF (i.e., a PIF/PEF ratio < 0.9) will favor secretion movement from the distal to the central airways. A recent study evaluating the transport of artificial mucus in a test lung system( ) reported that mucus transport is better explained by the absolute difference between PIF and PEF than by the PIF/PEF ratio. When PEF is higher than PIF, a greater difference between the two translates to better mucus transport. In the present study, the PIF/PEF ratio was below 0.9 in only 24 (16.7%) of the 144 maneuvers evaluated, most of those 24 being performed in the post-instruction phase. Consequently, the median PIF was much higher than was the median PEF, which is a cause for great concern. This disappointing result is in agreement with the findings of some other studies in which self-inflating resuscitators were also used.( , ) The manual resuscitator tested in the present study had a self-inflating bag (the type of manual resuscitator most widely used in ICUs in Brazil). Such resuscitators usually generate a lower VT than that achieved with resuscitators that have flow-inflating bags.( , , ) Even after receiving explicit instructions, the RTs produced VT values that were lower than those reported in other studies.( , , , ) That difference might be related solely to the size of the bag employed: 1.5 L in the present study, compared with 1.6-2.0 liters in the other studies cited. Much more favorable PIF/PEF ratios have been reported when flow-inflating devices were used.( ) The verbal instruction on how to perform the MH improved the performance of the RTs in that the post-instruction maneuvers generated lower PIFs and longer TIs. However, those differences did little to improve the efficiency of the technique, because the PEFs were still quite low. That is likely attributable to the fact that the Palv was also quite low at the start of exhalation, which resulted in a low driving pressure for expiratory flow. Other authors have reported great variability in the practical implementation of the MH maneuver,( , , - , ) the execution of which rarely follows its original description. This makes it practically impossible to compare the effectiveness of MH across clinical studies.( ) If we accept the concept that the relationship between PIF and PEF is responsible for the direction in which secretions move,( - ) we should question the use of an inspiratory pause on physiological grounds. Because of stress relaxation, the effective driving pressure for flow after a pause will be always lower than immediately after end-inspiration.( ) Therefore the use of an inspiratory pause could decrease the PEF and, consequently, impair the efficiency of the MH maneuver. The small number of RTs participating in this study prevents us from generalizing our data for application in clinical practice. However, we believe that this qualitative analysis, performed with representative operators (of both genders, recruited from different hospitals from within the same city), illustrated scenarios that commonly occur in some ICUs. Another limitation of this study was the use of a lung model, which cannot be extrapolated to the complexity of human lungs. Nevertheless, we believe that similar results-high PIF/PEF ratios, low VT, and high Pprox-would also be obtained in human patients on mechanical ventilation, especially in those that are heavily sedated, and this should draw the attention of the RTs to the way in which they perform the MH maneuver. Whether applying high PIF during MH will enhance secretion clearance in patients with preserved cough is a question that merits further investigation. That notwithstanding, the key message here is that performing MH with high PIF/PEF ratios in patients who are unable to cough will not aid secretion removal and might even contribute to mucus retention. Another point that should be mentioned is that the MH maneuver can be performed in combination with expiratory chest compression in order to maximize the increase in the PEF( , ); however we were not able to investigate the use of that combination, because of the experimental system used. In addition, as previously mentioned, we cannot discard the possibility that the use of a different resuscitator device (a self-inflating resuscitator with larger internal volume bags or a flow-inflating resuscitator) might produce better results. Given that, in Brazil, MH is typically performed without Pprox monitoring, the option of performing the maneuver with a mechanical ventilator, so that inspiratory flow, VT and pressure can be easily monitored and adjusted,( , ) should be considered. In studies comparing MH with a manual resuscitator and MH with a mechanical ventilator,( , , ) no differences were found between the two modalities in terms of the amount of secretion removed, although the advantages of better monitoring with the mechanical ventilator were acknowledged. It is important to note that, although the MH maneuver was originally designed to reduce lung collapse and to improve oxygenation or lung compliance, those potential benefits were not tested here. In fact, we believe that MH should be used only as a secretion clearance technique. If the technique is applied to recruit collapsed lungs, its beneficial effects are going to be offset, at least in part, by the fact that the patient must be disconnected from the ventilator, thus exposing the lung to the lower pressure of the ambient air at the end of each compression. Finally, we found no evidence that the MH maneuver, as performed here, increases the risk of barotrauma. The Palv value is a result of the insufflating volume and lung compliance, Pprox values that are higher than the Palv are caused by resistance in the orotracheal tube and airways. Although there have been anecdotal reports of high Pprox during MH in adults, with volumes < 1 L,( ) there is no hard evidence that the maneuver is dangerous. In conclusion, when asked to apply MH in accordance with their routine clinical practice, this small sample of RTs performed the maneuver quite differently from what is recommended by experts, producing a concerning pattern of ventilation in terms of secretion clearance. The repetition of the maneuver after explicit instructions reduced the Pprox but did not help much. Alveolar pressures were usually low (because of the small generated VT) despite high proximal pressures and high inspiratory peak flows. As a result, PEF was also low, far lower than the preceding PIF, a condition theoretically impacting the secretions deeper into the lung. Further studies are necessary in this area, especially focusing on the use of flow-inflating devices delivering higher tidal volumes.
  28 in total

1.  The effect of lung compliance and experience on manual hyperinflation.

Authors:  Bernard Rusterholz; Elizabeth Ellis
Journal:  Aust J Physiother       Date:  1998

2.  Manual hyperinflation: a description of the technique.

Authors:  Bredge McCarren; Chin Moi Chow
Journal:  Aust J Physiother       Date:  1996

3.  The effect of circuit type, volume delivered and "rapid release" on flow rates during manual hyperinflation.

Authors:  Lyndal J Maxwell; Elizabeth R Ellis
Journal:  Aust J Physiother       Date:  2003

4.  Pattern of ventilation during manual hyperinflation performed by physiotherapists.

Authors:  L J Maxwell; E R Ellis
Journal:  Anaesthesia       Date:  2007-01       Impact factor: 6.955

5.  Comparison of the effectiveness of manual and ventilator hyperinflation at different levels of positive end-expiratory pressure in artificially ventilated and intubated intensive care patients.

Authors:  Camila Savian; Jennifer Paratz; Andrew Davies
Journal:  Heart Lung       Date:  2006 Sep-Oct       Impact factor: 2.210

6.  Physical therapy in critically ill adult patients: recommendations from the Brazilian Association of Intensive Care Medicine Department of Physical Therapy.

Authors:  Eduardo Ériko Tenório de França; Francimar Ferrari; Patrícia Fernandes; Renata Cavalcanti; Antonio Duarte; Bruno Prata Martinez; Esperidião Elias Aquim; Marta Cristina Paulete Damasceno
Journal:  Rev Bras Ter Intensiva       Date:  2012-03

7.  Manual hyperinflation combined with expiratory rib cage compression for reduction of length of ICU stay in critically ill patients on mechanical ventilation.

Authors:  Juliana Savini Wey Berti; Elisiane Tonon; Carlos Fernando Ronchi; Heloisa Wey Berti; Laércio Martins de Stefano; Ana Lúcia Gut; Carlos Roberto Padovani; Ana Lucia Anjos Ferreira
Journal:  J Bras Pneumol       Date:  2012 Jul-Aug       Impact factor: 2.624

Review 8.  Secretion management in the mechanically ventilated patient.

Authors:  Richard D Branson
Journal:  Respir Care       Date:  2007-10       Impact factor: 2.258

9.  Performance of manual hyperinflation: a skills lab study among trained intensive care unit nurses.

Authors:  Frederique Paulus; Jan M Binnekade; Pauline Middelhoek; Margreeth B Vroom; Marcus J Schultz
Journal:  Med Sci Monit       Date:  2009-08

10.  Mucus clearance by two-phase gas-liquid flow mechanism: asymmetric periodic flow model.

Authors:  C S Kim; A J Iglesias; M A Sackner
Journal:  J Appl Physiol (1985)       Date:  1987-03
View more
  6 in total

1.  Effects of manual hyperinflation in preterm newborns under mechanical ventilation.

Authors:  Camila Chaves Viana; Carla Marques Nicolau; Regina Celia Turola Passos Juliani; Werther Brunow de Carvalho; Vera Lucia Jornada Krebs
Journal:  Rev Bras Ter Intensiva       Date:  2016-09

2.  Effects of manual hyperinflation, clinical practice versus expert recommendation, on displacement of mucus simulant: A laboratory study.

Authors:  Marcia S Volpe; Juliane M Naves; Gabriel G Ribeiro; Gualberto Ruas; Mauro R Tucci
Journal:  PLoS One       Date:  2018-02-12       Impact factor: 3.240

3.  Acute effects of ventilator hyperinflation with increased inspiratory time on respiratory mechanics: randomized crossover clinical trial.

Authors:  Luciano Matos Chicayban
Journal:  Rev Bras Ter Intensiva       Date:  2019-10-14

Review 4.  Manual hyperinflation in children.

Authors:  Daiane Menezes Lorena; Maria Cecília Moraes Frade; Thalis Henrique da Silva
Journal:  Rev Bras Ter Intensiva       Date:  2022-01-24

5.  Chest compression with a higher level of pressure support ventilation: effects on secretion removal, hemodynamics, and respiratory mechanics in patients on mechanical ventilation.

Authors:  Wagner da Silva Naue; Luiz Alberto Forgiarini Junior; Alexandre Simões Dias; Silvia Regina Rios Vieira
Journal:  J Bras Pneumol       Date:  2014 Jan-Feb       Impact factor: 2.624

6.  Lung hyperinflation by mechanical ventilation versus isolated tracheal aspiration in the bronchial hygiene of patients undergoing mechanical ventilation.

Authors:  Crisiela Brum Assmann; Paulo José Cardoso Vieira; Fernanda Kutchak; Marcelo de Mello Rieder; Soraia Genebra Ibrahim Forgiarini; Luiz Alberto Forgiarini Junior
Journal:  Rev Bras Ter Intensiva       Date:  2016 Jan-Mar
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.