Cell counting of body fluids: comparison between three automated haematology analysers and the manual microscope method.

P Danise1, M Maconi, A Rovetti, D Avino, A Di Palma, M Gerardo Pirofalo, C Esposito.   

Abstract

INTRODUCTION: Haematological analysis of body fluids (BF) specimens can provide clinicians with valuable diagnostic information because it can indicate one of several serious medical conditions. Although up to now the microscopic counting and the differentiation of WBC in a BF smear have been used as a reference. The introduction of semiautomated and automated methods of analysis has reduced interoperator variability and improved turnaround time and precision. The aim of our study was to evaluate the accuracy and the correlation between the three methods and with the reference method.
METHODS: We examined 110 body fluid samples. Total counting of each sample has been conducted with all systems: Pentra DX120, ADVIA 2120 and XE-2100 and the manual method.
RESULTS: We found statistically significant correlation between the data obtained in the ascitic and pleuric liquid but not in the cerebrospinal fluid.
CONCLUSION: The introduction of automated method for BF analysis is more and more useful in the routine job of a laboratory analysis. It is therefore very important to evaluate the performance of the different automated haematology technologies, because there is a lack of literature in this field. The comparison between the Pentra DX 120, the other technologies and the manual counting showed instrumental overlapping capabilities.
© 2013 John Wiley & Sons Ltd.

Entities:  

Keywords:  Body fluids; cell count; comparison; haematology analysers; white blood cells

Mesh:

Year:  2013        PMID: 23647736     DOI: 10.1111/ijlh.12093

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  4 in total

1.  Automated analysis for differentiating leukocytes in body fluids using the software "biological liquid application" on ADVIA2120/2120i hematology analyzer.

Authors:  Alessandra Falda; Paolo Doretto
Journal:  J Clin Lab Anal       Date:  2018-05-25       Impact factor: 2.352

2.  Evaluation of Sysmex XN-1000 hematology analyzer for cell count and screening of malignant cells of serous cavity effusion.

Authors:  Weiyi Xu; Qian Yu; Lixia Xie; Baode Chen; Ling Zhang
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

3.  Automated and manual microscopic analyses for leukocyte differential counts in exudative pleural effusions: Real-world disagreement and clinical application.

Authors:  Jaehee Lee; Yu Kyung Kim; Ji Eun Park; Yong Hoon Lee; Sun Ha Choi; Hyewon Seo; Seung Soo Yoo; Shin Yup Lee; Seung-Ick Cha; Jae Yong Park; Chang Ho Kim
Journal:  Medicine (Baltimore)       Date:  2022-09-16       Impact factor: 1.817

4.  Performance Evaluation of Body Fluid Cellular Analysis Using the Beckman Coulter UniCel DxH 800, Sysmex XN-350, and UF-5000 Automated Cellular Analyzers.

Authors:  Jooyoung Cho; Joowon Oh; Sang Guk Lee; You Hee Lee; Jaewoo Song; Jeong Ho Kim
Journal:  Ann Lab Med       Date:  2020-03       Impact factor: 3.464

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.