| Literature DB >> 23624022 |
Kylie A Beattie1, Chris Luscombe, Geoff Williams, Jordi Munoz-Muriedas, David J Gavaghan, Yi Cui, Gary R Mirams.
Abstract
INTRODUCTION: Drugs that prolong the QT interval on the electrocardiogram present a major safety concern for pharmaceutical companies and regulatory agencies. Despite a range of assays performed to assess compound effects on the QT interval, QT prolongation remains a major cause of attrition during compound development. In silico assays could alleviate such problems. In this study we evaluated an in silico method of predicting the results of a rabbit left-ventricular wedge assay.Entities:
Keywords: AP (D); Action Potential (Duration); Action potential; Cardiac safety; Compound screening; Concentration for 50% Inhibition; ECG; ECVAM; European Centre for the Validation of Alternative Methods; FLIPR; FLuorescence Imaging Plate Reader; GSK; GlaxoSmithKline; I(Kr); I(Ks); IC(50); ICH; International Conference for Harmonization; Ion channels; Mathematical model; Methods; QSAR; QT interval; Quantitative Structure Activity Relationship; Rabbit ventricular wedge; TdP; Torsades de Pointes; electrocardiogram; hERG; human-Ether-a-go-go Related Gene; minus log(10) of IC(50); pIC(50); rapid delayed rectifier potassium current; slow delayed rectifier potassium current
Mesh:
Substances:
Year: 2013 PMID: 23624022 PMCID: PMC4142193 DOI: 10.1016/j.vascn.2013.04.004
Source DB: PubMed Journal: J Pharmacol Toxicol Methods ISSN: 1056-8719 Impact factor: 1.950
Fig. 1a) An action potential generated from a single cell simulation and b) a pseudo-ECG from a one-dimensional tissue simulation at a range of concentrations for compound 2659 (see Supplementary data). Simulations were performed using PatchXpress data, parameterising the drug block model with the IC50 value and Hill coefficient in the single cell simulation and with just the IC50 value (and assuming the Hill coefficient is 1) in the one-dimensional simulation. The intervals used for calculation of the APD90 value and QT interval from the simulated control result are indicated in a) and b). Arrows indicate the effect observed in the APD90 value/QT interval with increasing compound concentration. In c) the percent change in APD90/QT interval determined from the simulation results and experimental results (both individual preparation results and the average of these) in the rabbit ventricular wedge assay is plotted for comparison.
Classification matrix presenting categorical classification of simulation and experimental results when considering the assay’s ability to predict QT prolongation. The results were obtained when using PatchXpress assay data, assuming compounds can interact with multiple channels and parameterising the drug block model with both the IC50 value and the Hill coefficient for a set of 77 compounds. The accuracy, sensitivity, specificity, and positive and negative predictive values are calculated as indicated.
| Simulation prolonger | Simulation non-prolonger | ||
|---|---|---|---|
| Experimental prolonger | 18 | 7 | Sensitivity = 18/(18 + 7) |
| Experimental non-prolonger | 10 | 42 | Sensitivity = 42/(42 + 10) |
| Positive predictive value = 18/(18 + 10) = 64.3% | Negative predictive value = 42/(42 + 7) = 85.7% | Accuracy = (18 + 42)/(18 + 7 + 10 + 42) = 77.9% |
Prolongation metric values calculated (with 95% confidence intervals indicated) for evaluation of the assay’s predictivity of QT prolongation in the rabbit ventricular wedge experiments when considering the different uses of the available ion channel data described. Entries are ordered alphabetically by the data type used in the simulations.
| Case number | Ion channel data use combination | Accuracy (%) | Kappa | Sensitivity | Specificity | Positive | Negative |
|---|---|---|---|---|---|---|---|
| 1 | IonWorks/FLIPR data, multiple channel block, fitting for IC50 value (121 compounds) | 77.7 | 0.50 | 73.0 | 79.8 | 61.4 | 87.0 |
| 2 | IonWorks/FLIPR data, multiple channel block, fitting for IC50 value & Hill coefficient (121 compounds) | 75.2 | 0.45 | 70.3 | 77.4 | 57.8 | 85.5 |
| 3 | IonWorks/FLIPR data, multiple channel block including KCNQ1 interactions, fitting for IC50 value (121 compounds) | 77.7 | 0.51 | 75.7 | 78.6 | 60.9 | 88.0 |
| 4 | PatchXpress data, hERG block only, fitting for IC50 value (77 compounds) | 72.7 (61.9–81.4) | 0.48 (0.29–0.67) | 96.0 (80.5–99.3) | 61.5 (48.0–73.5) | 54.5 (40.1–68.3) | 97.0 (84.7–99.5) |
| 5 | PatchXpress data, multiple channel block, fitting for IC50 value (77 compounds) | 74.0 | 0.41 | 60.0 | 80.8 | 60.0 | 80.8 |
| 6 | PatchXpress data, multiple channel block, fitting for IC50 value & Hill coefficient (77 compounds) | 77.9 | 0.51 | 72.0 | 80.8 | 64.3 | 85.7 |
| 7 | PatchXpress data, multiple channel block, fitting for IC50 value, one-dimensional simulation (77 compounds) | 71.4 | 0.41 | 76.0 | 69.2 | 54.3 | 85.7 |
| 8 | QSAR data, multiple channel block (372 compounds) | 55.4 | 0.17 | 91.4 | 27.6 | 49.3 | 80.6 |
Shortening metric values calculated (with 95% confidence intervals indicated) for evaluation of the assay’s predictivity of QT shortening in the rabbit ventricular wedge experiments when considering the different uses of the available ion channel data described. Entries are ordered alphabetically by the data type used in the simulations. The missing entry is due to the inability to calculate the metric due to there being no compounds in the relevant categories for its computation.
| Case number | Ion channel data use combination | Accuracy | Kappa | Sensitivity | Specificity | Positive | Negative |
|---|---|---|---|---|---|---|---|
| 1 | IonWorks/FLIPR data, multiple channel block, fitting for IC50 value (121 compounds) | 69.4 | 0.13 | 29.0 | 83.3 | 37.5 | 77.3 |
| 2 | IonWorks/FLIPR data, multiple channel block, fitting for IC50 value & Hill coefficient (121 compounds) | 71.1 | 0.23 | 41.9 | 81.1 | 43.3 | 80.2 |
| 3 | IonWorks/FLIPR data, multiple channel block including KCNQ1 interactions, fitting for IC50 value (121 compounds) | 69.4 | 0.13 | 29.0 | 83.3 | 37.5 | 77.3 |
| 4 | PatchXpress data, hERG block only, fitting for IC50 value (77 compounds) | 76.6 | 0.00 | 0.00 | 100 | – | 76.6 |
| 5 | PatchXpress data, multiple channel block, fitting for IC50 value (77 compounds) | 76.6 | 0.26 | 33.3 | 89.8 | 50.0 | 81.5 |
| 6 | PatchXpress data, multiple channel block, fitting for IC50 value & Hill coefficient (77 compounds) | 75.3 | 0.24 | 33.3 | 88.1 | 46.2 | 81.3 |
| 7 | PatchXpress data, multiple channel block, fitting for IC50 value, one-dimensional simulation (77 compounds) | 72.7 | 0.22 | 38.9 | 83.1 | 41.2 | 81.7 |
| 8 | QSAR data, multiple channel block (372 compounds) | 79.8 | 0.10 | 14.1 | 93.5 | 31.0 | 84.0 |
Fig. 2Plot of concentrations (EC10 values) at which a 10% change in APD90 value/QT interval is expected, as interpolated from simulation and experimental results, for each compound. The results from case 6, using PatchXpress data, are presented. Compounds exhibiting more than 10% change in their experimental or simulation results after drug administration, as compared to the control measurement, at at least one test concentration were included. This enabled interpolation of the EC10 values. Points plotted with an asterisk (*) are compounds for which more than a 10% change in the QT interval length (as compared to the control measurement) is exhibited in both simulation and experimental results. These points are plotted in one of four quadrants according to the classification of the simulation and experimental results: both experimental and simulation results show prolongation (quadrant 1), both show shortening (quadrant 3), experimental results show shortening and simulation results show prolongation (quadrant 2), simulation results show shortening and experimental results show prolongation (quadrant 4). Points plotted with a circle (○) are compounds for which only one of the experimental and simulation results show more than a 10% change in the QT interval as compared to the control. Interpolation is used to determine the EC10 value from this result. For the remaining experimental or simulation result, the percent change in QT interval is between −10% and 10% (showing “no effect”) at all concentrations tested. This result is assumed to have an EC10 value corresponding to the maximum concentration tested in the rabbit wedge assay amongst all the compounds which show “no effect” (with the percent change in APD90 or QT interval between −10% and 10%), which has a value of 500 μM. Half log unit lines which correspond to the error commonly associated with the ion channel assays are included as an indication of the range of accepted error.