Breast tumor detection using piezoelectric fingers: first clinical report.

Xin Xu1, Cynthia Gifford-Hollingsworth, Richard Sensenig, Wei-Heng Shih, Wan Y Shih, Ari D Brooks.   

Abstract

BACKGROUND: Mammography is key to detection of breast cancer in high-risk populations. Currently, aside from palpation and risk-assessment questionnaires, there is no prescreening test that can improve the accuracy, safety, and cost effectiveness of screening low-risk populations. The piezoelectric finger (PEF) is a radiation-free, portable, and low-cost breast tumor detector we developed to be used as a prescreening tool. STUDY
DESIGN: Patients presenting with breast abnormalities detected by palpation or imaging were enrolled in this IRB-approved study. The PEF testing was performed with the patient in supine position before undergoing biopsy or surgical excision. The locations of the lesions detected by PEF were compared with those confirmed on imaging or pathology.
RESULTS: A total of 40 patients were enrolled and 46 lesions were confirmed by imaging or pathology. The PEF reported 55 lesions, with 9 false positives and 2 true positives not originally found on imaging or palpation. The overall sensitivity of the PEF test was 87% (40 of 46). In women 40 years old or younger, overall sensitivity was or 100% (19 of 19). In women who had a lesion visible on mammography, PEF had a sensitivity of 83% (24 of 29). Of these, in women aged 40 years or younger, PEF identified all 7 mammographically visible lesions, including 2 malignant lesions. When compared with ultrasound, PEF correctly identified 87% (34 of 39) in this group. Of these, in women aged 40 years or younger, PEF identified 100% (19 of 19) of all ultrasound-visible lesions.
CONCLUSIONS: The PEF identified abnormalities in all 39 patients who presented with breast abnormalities and did not demonstrate any false negatives that would prevent the patients from additional evaluation, which makes it a good prescreening tool. In addition, PEF demonstrated 100% sensitivity in women aged 40 years or younger, a traditionally low-risk population.
Copyright © 2013 American College of Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23623223     DOI: 10.1016/j.jamcollsurg.2013.02.022

Source DB:  PubMed          Journal:  J Am Coll Surg        ISSN: 1072-7515            Impact factor:   6.113


  10 in total

1.  Development of array piezoelectric fingers towards in vivo breast tumor detection.

Authors:  Xin Xu; Youngsoo Chung; Ari D Brooks; Wei-Heng Shih; Wan Y Shih
Journal:  Rev Sci Instrum       Date:  2016-12       Impact factor: 1.523

Review 2.  Breast Cancer Screening: a Paradigm Change Is Needed-the PB Desai Oration.

Authors:  Ismail Jatoi
Journal:  Indian J Surg Oncol       Date:  2019-12-03

3.  A model study of 3-dimensional localization of breast tumors using piezoelectric fingers of different probe sizes.

Authors:  Xin Xu; Wei-Heng Shih; Wan Y Shih
Journal:  Rev Sci Instrum       Date:  2019-01       Impact factor: 1.523

4.  Inverse finite-element modeling for tissue parameter identification using a rolling indentation probe.

Authors:  Hongbin Liu; Kiattisak Sangpradit; Min Li; Prokar Dasgupta; Kaspar Althoefer; Lakmal D Seneviratne
Journal:  Med Biol Eng Comput       Date:  2013-09-15       Impact factor: 2.602

5.  Technical note: Low clinical efficacy, but good acceptability of a point-of-care electronic palpation device for breast cancer screening for a lower middle-income environment.

Authors:  Dustin Valdez; Teofila Cruz; Stephanie Rania; Grazyna Badowski; Kevin Cassel; Thomas Wolfgruber; Scott Grosskreutz; Louis J Dulana; Roy Adonay; Gertraud Maskarinec; John A Shepherd
Journal:  Med Phys       Date:  2022-02-10       Impact factor: 4.071

Review 6.  Future directions for the early detection of recurrent breast cancer.

Authors:  Erika J Schneble; Lindsey J Graham; Matthew P Shupe; Frederick L Flynt; Kevin P Banks; Aaron D Kirkpatrick; Aviram Nissan; Leonard Henry; Alexander Stojadinovic; Nathan M Shumway; Itzhak Avital; George E Peoples; Robert F Setlik
Journal:  J Cancer       Date:  2014-03-16       Impact factor: 4.207

7.  A cost-effective handheld breast scanner for use in low-resource environments: a validation study.

Authors:  Robyn B Broach; Rula Geha; Brian S Englander; Lucy DeLaCruz; Holly Thrash; Ari D Brooks
Journal:  World J Surg Oncol       Date:  2016-10-28       Impact factor: 2.754

8.  Clinical Utility of a Hand-Held Scanner for Breast Cancer Early Detection and Patient Triage.

Authors:  Julie M Clanahan; Sanjana Reddy; Robyn B Broach; Anne F Rositch; Benjamin O Anderson; E Paul Wileyto; Brian S Englander; Ari D Brooks
Journal:  JCO Glob Oncol       Date:  2020-02

Review 9.  Bio-piezoelectricity: fundamentals and applications in tissue engineering and regenerative medicine.

Authors:  Nagwa Ahmed Kamel
Journal:  Biophys Rev       Date:  2022-06-28

Review 10.  Early Diagnosis of Breast Cancer.

Authors:  Lulu Wang
Journal:  Sensors (Basel)       Date:  2017-07-05       Impact factor: 3.576