Literature DB >> 23622035

The RETRIEVE trial: safety and effectiveness of the retrievable crux vena cava filter.

H Bob Smouse1, Robert Mendes, Marc Bosiers, Thuong G Van Ha, Tami Crabtree.   

Abstract

PURPOSE: To evaluate the safety and effectiveness of the Crux vena cava filter in patients at risk for pulmonary embolism (PE).
MATERIALS AND METHODS: The Crux Biomedical Evaluation of the Crux Inferior Vena Cava Filter System trial was an international prospective, multicenter, single-arm clinical trial in 125 patients implanted with the Crux filter between June 2010 and June 2011. Follow-up was 180 days after filter placement and 30 days after filter retrieval. The primary objective was to determine whether the clinical success rate was at least 80%. Clinical success was defined as technical success of deployment and freedom from definite PE, filter migration, and device-related adverse events requiring intervention.
RESULTS: The clinical success rate was 96.0% (120 of 125), with a one-sided lower limit of the 95% confidence interval of 91.8%. The rate of technical success was 98.4% (123 of 125). There were three cases of definite PE (2.4%), two cases of deployment failure, and no cases of device migration, embolization, fracture, or tilting. Investigators observed nine cases of thrombus (all nonocclusive) in or near the filter (six during retrieval evaluation vena cavography, two during computed tomography [CT] scans for PE symptoms, and one during CT for cancer management) and 13 cases of deep vein thrombosis. Device retrieval was attempted at a mean of 84.6 days±57.6 (range, 6-190 d) after implantation and was successful for 98.1% of patients (53 of 54). All deaths (n = 14) were determined to be unrelated to the filter or PE.
CONCLUSIONS: The Crux vena cava filter performed safely, with high rates of clinical, technical, and retrieval success.
Copyright © 2013 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23622035     DOI: 10.1016/j.jvir.2013.01.489

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  4 in total

1.  Inferior Vena Cava Filters: Indications, Outcomes, and Evidence.

Authors:  Jennifer P Montgomery; John A Kaufman
Journal:  Curr Treat Options Cardiovasc Med       Date:  2015-09

Review 2.  Evidence-Based Evaluation of Inferior Vena Cava Filter Complications Based on Filter Type.

Authors:  Steven E Deso; Ibrahim A Idakoji; William T Kuo
Journal:  Semin Intervent Radiol       Date:  2016-06       Impact factor: 1.513

Review 3.  A Critical Review of Available Retrievable Inferior Vena Cava Filters and Future Directions.

Authors:  Jennifer P Montgomery; John A Kaufman
Journal:  Semin Intervent Radiol       Date:  2016-06       Impact factor: 1.513

Review 4.  Gone in 60 days: our first experience with a bioconvertible IVC filter.

Authors:  Gerard Lambe; Johnny O' Mahony; Michael Courtney; Noel Donlon; Claire Donohoe; JMark Ryan
Journal:  Ir J Med Sci       Date:  2021-11-02       Impact factor: 2.089

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.