On the use of an analytic source model for dose calculations in precision image-guided small animal radiotherapy.

Patrick V Granton1, Frank Verhaegen.   

Abstract

Precision image-guided small animal radiotherapy is rapidly advancing through the use of dedicated micro-irradiation devices. However, precise modeling of these devices in model-based dose-calculation algorithms such as Monte Carlo (MC) simulations continue to present challenges due to a combination of very small beams, low mechanical tolerances on beam collimation, positioning and long calculation times. The specific intent of this investigation is to introduce and demonstrate the viability of a fast analytical source model (AM) for use in either investigating improvements in collimator design or for use in faster dose calculations. MC models using BEAMnrc were developed for circular and square fields sizes from 1 to 25 mm in diameter (or side) that incorporated the intensity distribution of the focal spot modeled after an experimental pinhole image. These MC models were used to generate phase space files (PSFMC) at the exit of the collimators. An AM was developed that included the intensity distribution of the focal spot, a pre-calculated x-ray spectrum, and the collimator-specific entrance and exit apertures. The AM was used to generate photon fluence intensity distributions (ΦAM) and PSFAM containing photons radiating at angles according to the focal spot intensity distribution. MC dose calculations using DOSXYZnrc in a water and mouse phantom differing only by source used (PSFMC versus PSFAM) were found to agree within 7% and 4% for the smallest 1 and 2 mm collimator, respectively, and within 1% for all other field sizes based on depth dose profiles. PSF generation times were approximately 1200 times faster for the smallest beam and 19 times faster for the largest beam. The influence of the focal spot intensity distribution on output and on beam shape was quantified and found to play a significant role in calculated dose distributions. Beam profile differences due to collimator alignment were found in both small and large collimators sensitive to shifts of 1 mm with respect to the central axis.

Entities:  

Mesh:

Year:  2013        PMID: 23615380     DOI: 10.1088/0031-9155/58/10/3377

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  8 in total

1.  Investigating the accuracy of microstereotactic-body-radiotherapy utilizing anatomically accurate 3D printed rodent-morphic dosimeters.

Authors:  Steven T Bache; Titania Juang; Matthew D Belley; Bridget F Koontz; John Adamovics; Terry T Yoshizumi; David G Kirsch; Mark Oldham
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

Review 2.  Tumour and normal tissue radiobiology in mouse models: how close are mice to mini-humans?

Authors:  Bridget F Koontz; Frank Verhaegen; Dirk De Ruysscher
Journal:  Br J Radiol       Date:  2016-09-26       Impact factor: 3.039

3.  A sparse orthogonal collimator for small animal intensity-modulated radiation therapy part I: Planning system development and commissioning.

Authors:  Kaley Woods; Dan Nguyen; Ryan Neph; Dan Ruan; Daniel O'Connor; Ke Sheng
Journal:  Med Phys       Date:  2019-11-04       Impact factor: 4.071

Review 4.  Monte Carlo methods for device simulations in radiation therapy.

Authors:  Hyojun Park; Harald Paganetti; Jan Schuemann; Xun Jia; Chul Hee Min
Journal:  Phys Med Biol       Date:  2021-09-14       Impact factor: 4.174

5.  A kernel-based dose calculation algorithm for kV photon beams with explicit handling of energy and material dependencies.

Authors:  Anna Merle Reinhart; Martin F Fast; Peter Ziegenhein; Simeon Nill; Uwe Oelfke
Journal:  Br J Radiol       Date:  2016-10-27       Impact factor: 3.039

6.  Small field dosimetry for the small animal radiotherapy research platform (SARRP).

Authors:  Mihaela Ghita; Stephen J McMahon; Hannah F Thompson; Conor K McGarry; Raymond King; Sarah O S Osman; Jonathan L Kane; Amanda Tulk; Giuseppe Schettino; Karl T Butterworth; Alan R Hounsell; Kevin M Prise
Journal:  Radiat Oncol       Date:  2017-12-28       Impact factor: 3.481

7.  A scoping review of small animal image-guided radiotherapy research: Advances, impact and future opportunities in translational radiobiology.

Authors:  Kathryn H Brown; Mihaela Ghita; Ludwig J Dubois; Dirk de Ruysscher; Kevin M Prise; Frank Verhaegen; Karl T Butterworth
Journal:  Clin Transl Radiat Oncol       Date:  2022-04-06

8.  The impact of dual energy CT imaging on dose calculations for pre-clinical studies.

Authors:  Ana Vaniqui; Lotte E J R Schyns; Isabel P Almeida; Brent van der Heyden; Stefan J van Hoof; Frank Verhaegen
Journal:  Radiat Oncol       Date:  2017-11-21       Impact factor: 3.481

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.