| Literature DB >> 23612893 |
John W D Speirs1, Thomas H Burke, Stefanie Y Lee, Brigitte D Ala.
Abstract
BACKGROUND: Complete packing of intracranial aneurysms has demonstrated a significant decrease in aneurysm recurrence rates with increased volumetric filling. The HydroCoil Embolization System (HES) was developed to increase volumetric filling within the aneurysm sac to maintain long term occlusion. To further enhance ease of HES deployment, a new next generation embolic coil, the HydroFill coil, was developed.Entities:
Keywords: Aneurysm; Angiography; Coil; Device; Intervention
Mesh:
Substances:
Year: 2013 PMID: 23612893 PMCID: PMC3812844 DOI: 10.1136/neurintsurg-2012-010643
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1HydroFill Coil—Hydrogel core with a platinum overcoil.
Aneurysm size and calculated volume fill
| Patient No | Location of aneurysm | Size of aneurysm (neck) (mm) | Calculated volume filled | |
|---|---|---|---|---|
| Total (%) | HydroFill (%) | |||
| 1 | R pericallosal | 4.4× 5.0×5.6 (3.5) | 13 | 5 |
| 2 | R PCom | 3.6×3.2 × 5.0 (2.7) | 68 | 44 |
| 3 | L ACom | 8.5×5.7×6.2 (5.2) | 27 | 10 |
| 4 | R MCA | 5.4×7.6×5.6 (4.9) | 26 | 8 |
| 5 | L pericallosal | 9.2×7.4×6.1 (3.5) | 27 | 11 |
| 6 | R supraclinoid | 2.5×3.1×2.6 (2.1) | 135 | 93 |
| R supraclinoid | 3.0×3.3×2.5 (2.3) | 77 | 77 | |
| 7 | L MCA | 3.4×4.2×4.2 (2.5) | 78 | 31 |
| R ICA | 7.4×6.0×5.5 (3.0) | 36 | 21 | |
| R paraclinoid | 4.7×4.4×4.0 (1.8) | 60 | 23 | |
| 8 | R ACom | 2.0×2.6×3.0 (2.5) | 97 | 80 |
| 9 | L A1 | 4.2×3.7×2.3 (1.3) | 133 | 116 |
| 10 | L ACom | 4.8×4.3×4.4 (2.4) | 31 | 14 |
| 11 | R MCA | 6.6×5.4×3.4 (2.2) | 89 | 47 |
A1, proximal segment of the anterior cerebral artery; ACom, anterior communicating artery; ICA, internal carotid artery; L, left; MCA, middle cerebral artery; PCom, posterior communicating artery; R, right.
Figure 2Acute subarachnoid hemorrhage (patient No 9). Left internal carotid angiogram shows a posterior wall A1 segment aneurysm (A, B). Immediate post embolization image (C) demonstrates near complete occlusion with small neck remnant (arrow). Twenty minute delayed control angiogram (D) demonstrates total occlusion of the entire sac. Eighteen month contrast enhanced MR angiography demonstrates continued stable occlusion of the aneurysm sac (E, F).
Figure 3Un-prepped HydroCoil Embolization System 10 and HydroFill prior to insertion.