Interactive, multi-modality image registrations for combined MRI/MRSI-planned HDR prostate brachytherapy.

Galen Reed1, J Adam Cunha, Susan Noworolski, John Kurhanewicz, Daniel Vigneron, I-Chow Hsu, Jean Pouliot.   

Abstract

PURPOSE: This study presents the steps and criteria involved in the series of image registrations used clinically during the planning and dose delivery of focal high dose-rate (HDR) brachytherapy of the prostate.
MATERIAL AND METHODS: Three imaging modalities - Magnetic Resonance Imaging (MRI), Magnetic Resonance Spectroscopic Imaging (MRSI), and Computed Tomography (CT) - were used at different steps during the process. MRSI is used for identification of dominant intraprosatic lesions (DIL). A series of rigid and nonrigid transformations were applied to the data to correct for endorectal-coil-induced deformations and for alignment with the planning CT. Mutual information was calculated as a morphing metric. An inverse planning optimization algorithm was applied to boost dose to the DIL while providing protection to the urethra, penile bulb, rectum, and bladder. Six prostate cancer patients were treated using this protocol.
RESULTS: The morphing algorithm successfully modeled the probe-induced prostatic distortion. Mutual information calculated between the morphed images and images acquired without the endorectal probe showed a significant (p = 0.0071) increase to that calculated between the unmorphed images and images acquired without the endorectal probe. Both mutual information and visual inspection serve as effective diagnostics of image morphing. The entire procedure adds less than thirty minutes to the treatment planning.
CONCLUSION: This work demonstrates the utility of image transformations and registrations to HDR brachytherapy of prostate cancer.

Entities:  

Keywords:  HDR; MRSI; image registration; prostate

Year:  2011        PMID: 23606866      PMCID: PMC3627724          DOI: 10.5114/jcb.2011.21040

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

Studies have shown that greater control of localized prostatic tumors is directly correlated with radiation doses delivered [1-3]. However, excessive dose to the normal tissues causes side effects [4-6]. High dose rate (HDR) brachytherapy can provide a focal dose escalation, and prior studies have shown that combined external beam radiation therapy (EBRT) and HDR brachytherapy yields favorable outcomes, particularly in patients with locally advanced disease [7-9]. We have recently initiated an HDR brachytherapy clinical protocol in which a combination of magnetic resonance imaging (MRI) and magnetic resonance spectroscopic imaging (MRSI) is used to define dominant intraprostatic lesions (DIL). This information is used to perform dose escalation of the DIL without compromising the dose coverage of the prostate nor the protection to the urethra, rectum, and bladder. There are two difficulties, however, inherent in this approach. The first is the merging of information from the staging MRI/MRSI exam with the anatomic computed tomography (CT) exam required for treatment planning; the second is the capacity of the treatment modality/treatment planning system to deliver radiation doses in a precise fashion. The combined MRI/MRSI exam has been shown to be an effective method of detecting and localizing intraprostatic lesions [10-13]. The high sensitivity and specificity yielded by this exam has demonstrated great utility in focal treatment via 125I brachytherapy implants [14] and combined HDR brachytherapy + EBRT [15, 16]. However, since the endorectal probe used for MRI/MRSI acquisition can induce significant prostatic compression and deformation, a non-rigid transformation is required to accurately register the probe-in MRSI data with the probe-out CT images used for treatment planning. To model this deformation, prior studies have implemented finite-element based biomechanical simulations [17, 18], elastic spline deformation [19], thin plate spline deformation [19-23], symmetric forces algorithms [24], Newton-Ralphson algorithms [25], and in-plane linear scaling assuming constant volume [14]. Kim et al. [26] showed that the deformation induced by the inflatable endorectal coil is negligible in the z (superior/inferior) dimension since the induced distortions are less than the axial MRI slice thickness of 3 mm. The same study also showed that the coil induces a non-negligible rotation about the x (right/left) axis. These findings suggest that an x axis rotation followed by an axial in-plane non-rigid deformation sufficiently models the prostatic deformation induced by the endorectal coil. In this study, we developed and implemented a new method combining rigid translations (for centroid alignment), rotations (to model the x axis rotation), and an in-plane non-rigid control-point based morphing method that utilizes a local weighted-mean transformation [27]. Additionally, we present the steps and criteria involved in the series of image registrations used clinically during the planning of the dose delivery process.

Material and methods

For each patient enrolled in the study, a pre-therapy staging MRI/MRSI exam using an endorectal probe (i.e. probe in, or PI) was acquired. The MRI and MRSI images were registered so that the MRSI-defined DIL could be delineated on the anatomic MRI volume. An axial MRI series with the endorectal probe not inserted (i.e. probe out, or MRI-PO) was also acquired. Pairs of points at corresponding anatomic features on the MRI-PI and MRI-PO series were chosen and used to determine the PI-PO transformation. This transformation algorithm was used to correlate spectroscopic data (MRSI-PI) to the MRI-PO data. Mutual information was calculated to measure warping accuracy. On the HDR brachytherapy treatment day, the planning CT and the MRSI-PO were imported into the brachytherapy planning system and registered to each other. The prostate anatomy alone was used to guide the fusion. The target and organs at risk were delineated using the CT, while the DIL was contoured using spectroscopic information from the MRSI-PO. Catheters were then digitized, and the plan was optimized using the Inverse Planning by Simulated Annealing (IPSA) algorithm [16]. The optimization was instructed to provide a clinically effective dose to the entire prostate while concurrently delivering a dose escalation to the DILs. This was done without increasing the dose to surrounding non-cancerous tissues. Once the plan was approved, the dose for the first fraction was delivered.

Subjects

Six patients were scanned and subsequently treated with this protocol. Their age ranged from 44 to 80 years old (mean 61). All patients had biopsy-proven Gleason 3 + 3 disease, and five of the six subjects showed Gleason 3 + 4 pattern in histological analysis. The mean PSA measurement for all subjects was 8.8 ng/mL (ranging from 4.3 ng/mL to 19.5 ng/mL). Written informed consent was obtained for all subjects following a protocol approved by the UCSF Committee of Human Research.

Identification of DIL with MRS and MRI

MR imaging was acquired with a GE 3T SIGNA scanner (GE Medical Systems, Waukesha, WI) using body-coil excitation. The GE 8-channel pelvic phased array and Medrad endorectal coil (Medrad, Pittsburgh, PA) were used for signal reception. Axial T1-weighted images were acquired with TR/TE = 600/12. Fast spin echo T2-weighted images were acquired (TR/TE = 6000/102, 14 cm FOV, 256 × 256 matrix) in the axial, sagittal, and coronal planes. Images and spectroscopic data were first acquired with the endorectal probe inserted. The probe was then removed, and an additional sagittal locator and an axial FSE T2 image data set were then acquired using the pelvic phased array for signal reception. Spectroscopic data was acquired using a specialized prostate 3D MRSI sequence [28, 29] with a 12 × 8 × 8 acquisition grid, 5.4 mm isotropic resolution yielding 0.157 cm3 voxels. Water and lipid suppression was achieved using dual-band spectral spatial pulses [30]. Very-selective spatial saturation (VSS) pulses were used to suppress periprostatic lipids [31]. In each voxel 1024 points were acquired over a 1000 Hz frequency domain. The k-space data was zero-filled in the superior/inferior and anterior/posterior dimensions to a final array size of 12 × 16 × 16. The spectral data was then apodized with a 3 Hz Gaussian filter, Fourier transformed, baseline corrected, frequency aligned, and peaks numerically-integrated. Numerical integration of each of the prostate metabolite peaks (choline, creatine, and citrate) and the suppressed water peak was performed using the known frequency positions of each of these peaks [32]. Spectroscopic voxels were classified using the standardized scoring system proposed by Jung et al. [33] where 1 = definitely normal, 2 = probably normal, 3 = equivocal, 4 = probably abnormal, and 5 = definitely abnormal. After the generation of these scores by a trained reader, grayscale images of these scores were created at the MRSI resolution, so that the suspicious voxels could be delineated by color on the high-resolution T2 volume (Fig. 1). All peripheral zone voxels were tagged; central zone voxels were only tagged when suspicious for malignancy.
Fig. 1

A shows a typical healthy prostate NMR spectrum, with high citrate and polyamine sigal and low choline. B shows the spectral data of a patient enrolled in the study; elevated choline and reduced citrate in the left peripheral zone are indicative of malignancy. C shows the suspicious regions (scores of 3, 4, and 5) delineated on the T2 image. All peripheral zone voxels were tagged; central zone voxels were only tagged when suspicious for malignancy. Only intraprostatic voxels are shown

A shows a typical healthy prostate NMR spectrum, with high citrate and polyamine sigal and low choline. B shows the spectral data of a patient enrolled in the study; elevated choline and reduced citrate in the left peripheral zone are indicative of malignancy. C shows the suspicious regions (scores of 3, 4, and 5) delineated on the T2 image. All peripheral zone voxels were tagged; central zone voxels were only tagged when suspicious for malignancy. Only intraprostatic voxels are shown

Warping the MRSI grid

Rotation

Sagittal images acquired with and without the endorectal probe were analyzed to determine the probe-induced rotation of the prostate about the x axis. The prostatic alignment of each series is determined by the margin of the peripheral zone and central zone as shown in Fig. 2. The image volumes are rotated by the net rotation angle θ − θ using tri-linear interpolation.
Fig. 2

Mid-sagittal probe-in (A) and probe-out (B) images showing the prostatic rotation induced by the endorectal coil. The angles were measured between the y axis and the peripheral zone/central zone margin denoted by the dotted line

Mid-sagittal probe-in (A) and probe-out (B) images showing the prostatic rotation induced by the endorectal coil. The angles were measured between the y axis and the peripheral zone/central zone margin denoted by the dotted line

Translation

Prostatic margins were then hand-drawn on both the axial probe-out volume and the rotated axial probe-in volume. Masks were drawn on the ten to twelve corresponding slices in each volume. The centroids of the masks were calculated, and the rotated axial probe-in volumes were translated so their centroids line up with the centroid of the probe-out volume. Defining RCI and RCO as the centers and CI and CO as the prostate centroids of the probe-in and probe-out volumes respectively, the translation aligns the prostate in both volumes provided they are acquired using the same landmark.

Non-rigid deformation

After the rotation and translations were applied, a non-rigid deformation was used to model the prostatic deformation induced by the endorectal coil. To model this deformation, the probe-in and probe-out T2 imaging series were compared slice-wise. In each slice, control point pairs were assigned to matching anatomical landmarks within the prostate (Fig. 3). The control point locations were assigned using the MATLAB control point selection tool. These point pairs were then used to define a local weighted-mean [27] PI-PO transformation. This non-rigid transformation placed higher weighting on image regions with a higher control point density. Therefore, the operator could apply more control points onto the regions where higher accuracy was required with the registration. Examples are regions of higher prostatic compression (the posterior aspect) and regions of suspected malignancy. The transformation was applied to the MRI-PI volume (to calculate morphing accuracy) and to the MRSI-PI data. Finally, the MRSI-PI data (after rotation, translation, and deformation) were delineated on the probe-out T2 volume (Fig. 4).
Fig. 3

Probe-in (A) and probe-out (B) axial slices showing control point placement. The probe-in volume is rotated by the prostatic angle so the slices are assumed to be coplanar. Despite the lower signal-to-noise ratio on the probe-out image, numerous landmarks are still identifiable on both images

Fig. 4

The left image shows the original acquisition image with the suspicious regions delineated. The center image shows the axial probe-out volume with the rotate, translated, and warped spectral delineations. MRSI scores of 4 and 5 are delineated on the planning CT (right)

Probe-in (A) and probe-out (B) axial slices showing control point placement. The probe-in volume is rotated by the prostatic angle so the slices are assumed to be coplanar. Despite the lower signal-to-noise ratio on the probe-out image, numerous landmarks are still identifiable on both images The left image shows the original acquisition image with the suspicious regions delineated. The center image shows the axial probe-out volume with the rotate, translated, and warped spectral delineations. MRSI scores of 4 and 5 are delineated on the planning CT (right)

Evaluation

Mutual information (MI) was calculated between the MRI-PO images and both the warped and non-warped MRI-PI images after rotations and translations were applied. This value was calculated similar to previous works [34] using the histogram method. MI was calculated only in the smallest-fitting rectangular boundary around the hand-drawn mask for each slice so that the erroneous morphing in the extra-prostatic regions (where no control points are placed) did not excessively sway the metric.

Target delineation for dose planning optimization

MRSI – planning CT registration

On the HDR brachytherapy treatment day, a CT scan with 3 mm slice thickness was acquired immediately after the patient recovered from the implant procedure. The imaging volume included the whole of the prostate and was limited superiorly to include the bladder and inferiorly to visualize all the catheters down to the perineum. The planning CT and the MRSI-PO were imported in the brachytherapy planning system (OncentraBrachy™, Nucletron, Veenendal, the Netherlands), and a rigid body registration was obtained. The prostate anatomy alone was used to guide the fusion, since the prostatic position can vary with respect to extra-prostatic landmarks. Typically, 3 pair-points were defined: two by the urethra in the base and apex areas, and a third one more lateral to the prostate in the median plane. Target and organs at risk were delineated on the CT while the DILs were contoured on the MRSI-PO using spectroscopic information. Catheters were then digitized, the plan was optimized using IPSA, and the dose for the first fraction was delivered.

Definition of volumes of interest

Clinical target volumes (prostate) and organs at risk (urethra, bladder, rectum and bulb) were contoured on each CT slice (Fig. 5A). The DILs – defined by MRSI voxel scores of 4 or 5 – were manually contoured on the CT, but the transparency level was adjusted to make the MRS information visible on the CT (Fig. 4 right). After the catheters were digitized on the CT, all information requested for the optimization of the dose distribution was available (Fig. 5B).
Fig. 5

Digitized catheters and delineated target organs: rectum (brown), urinary bladder (yellow mesh), and urethra (solid yellow). The MRSI-defined DIL is shown in pink on the left, and the right figure shows an isodose contour

Digitized catheters and delineated target organs: rectum (brown), urinary bladder (yellow mesh), and urethra (solid yellow). The MRSI-defined DIL is shown in pink on the left, and the right figure shows an isodose contour

Dose optimization with inverse planning IPSA

The inverse planning optimization algorithm was then used to increase the dose delivered to the dominant intra-prostatic lesions defined by MRI/MRSI while providing the usual dose coverage of the prostate and the protection to the urethra, rectum, bulb and bladder. A class solution – a set of weights provided to the optimization to guide it to satisfy the dose requirements of each region of interest – was previously developed [15] for dose escalation of a DIL defined by combined MRI/MRSI in inverse planned HDR prostate brachytherapy. Using this class solution, a certain level of DIL-boost was feasible for the majority of patients under the RTOG-0321 dosimetric requirements depending on rectal and bladder doses. The class solution in inverse planned HDR prostate brachytherapy for dose escalation of a DIL defined by combined MRI/MRSI was an excellent starting point to explore a customized set of dose constraints to obtain a satisfactory treatment plan for each patient in the ongoing protocol.

Results and Discussion

The complete procedure was clinically integrated and has been used for the treatment of six HDR brachytherapy patients with MRSI information. Out of the 6 patients scanned, the mean prostatic rotation induced by the coil was 20 ± 9, and the mean gland volume was 27 + 14 cm3. Urethra, peripheral zone margins, prostate boundaries, and various hyper/hypointense features on the images served as effective landmarks for the MRI-PI–MRI-PO fusion (Fig. 3). At least 12 point pairs – selected primarily from regions with high deformation, such as the posterior aspect of the peripheral zone, and regions of spectroscopic abnormality – are required as inputs to the MATLAB image transformation function. Visual inspection as well as the computed MI (Fig. 6) served as effective diagnostics of the morphing. MI showed 25% ± 13% increase for the six patients. A paired t-test showed this improvement to be significant (p = 0.0071). On the day of the first fraction, performing the MRSI-CT fusion was followed by the delineation of the DIL adding less than fifteen minutes to the entire planning process. The 3 pair-points registration procedure is considered valid when the sum of squared distances between each pair-point is less than 2 mm. A careful visual inspection of the fusion in the prostate area was also performed.
Fig. 6

Mutual information before and after applying the described transformations to the MRI volumes for each of the six patients in the trial. Each data point represents the mean values of all slices in the prostate

Mutual information before and after applying the described transformations to the MRI volumes for each of the six patients in the trial. Each data point represents the mean values of all slices in the prostate

Conclusions

An MRSI to MRI/CT alignment protocol was developed to exploit the high specificity of combined MRI/MRSI for detecting and localizing prostate cancer within the prostate, and to accurately transfer this information to the planning images. This makes information about the prostate cancer location routinely available, and allowing the use of inverse planning IPSA to boost dominant intraprostatic lesions during HDR brachytherapy, while preserving the prostate coverage and keeping the dose delivered to the organs at risk at the same level compared to an inverse planned dose distribution without DIL boost. Information from one image data set could be accurately transferred to another in the series of MRS → MRI → CT. This workflow was routinely used for the dose planning, including DIL boost. This work illustrates the clinical benefit of image registration tools.
  31 in total

1.  Dualband spectral-spatial RF pulses for prostate MR spectroscopic imaging.

Authors:  A A Schricker; J M Pauly; J Kurhanewicz; M G Swanson; D B Vigneron
Journal:  Magn Reson Med       Date:  2001-12       Impact factor: 4.668

Review 2.  Three-dimensional magnetic resonance spectroscopic imaging of brain and prostate cancer.

Authors:  J Kurhanewicz; D B Vigneron; S J Nelson
Journal:  Neoplasia       Date:  2000 Jan-Apr       Impact factor: 5.715

3.  Inverse planning for HDR prostate brachytherapy used to boost dominant intraprostatic lesions defined by magnetic resonance spectroscopy imaging.

Authors:  Jean Pouliot; Yongbok Kim; Etienne Lessard; I-Chow Hsu; Daniel B Vigneron; John Kurhanewicz
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-07-15       Impact factor: 7.038

4.  Automated finite-element analysis for deformable registration of prostate images.

Authors:  Jessica R Crouch; Stephen M Pizer; Edward L Chaney; Yu-Chi Hu; Gig S Mageras; Marco Zaider
Journal:  IEEE Trans Med Imaging       Date:  2007-10       Impact factor: 10.048

5.  Registration of MR prostate images with biomechanical modeling and nonlinear parameter estimation.

Authors:  Ron Alterovitz; Ken Goldberg; Jean Pouliot; I-Chow Joe Hsu; Yongbok Kim; Susan Moyher Noworolski; John Kurhanewicz
Journal:  Med Phys       Date:  2006-02       Impact factor: 4.071

6.  Interactive deformation registration of endorectal prostate MRI using ITK thin plate splines.

Authors:  M Rex Cheung; Karthik Krishnan
Journal:  Acad Radiol       Date:  2009-03       Impact factor: 3.173

7.  Prostate depiction at endorectal MR spectroscopic imaging: investigation of a standardized evaluation system.

Authors:  Juyoung A Jung; Fergus V Coakley; Daniel B Vigneron; Mark G Swanson; Aliya Qayyum; Vivian Weinberg; Kirk D Jones; Peter R Carroll; John Kurhanewicz
Journal:  Radiology       Date:  2004-12       Impact factor: 11.105

8.  Mapping of the prostate in endorectal coil-based MRI/MRSI and CT: a deformable registration and validation study.

Authors:  J Lian; L Xing; S Hunjan; C Dumoulin; J Levin; A Lo; R Watkins; K Rohling; R Giaquinto; D Kim; D Spielman; B Daniel
Journal:  Med Phys       Date:  2004-11       Impact factor: 4.071

Review 9.  Localized carcinoma of the prostate (stages T1B, T1C, T2, and T3). Review of management with external beam radiation therapy.

Authors:  C A Perez; G E Hanks; S A Leibel; A L Zietman; Z Fuks; W R Lee
Journal:  Cancer       Date:  1993-12-01       Impact factor: 6.860

10.  Expandable and rigid endorectal coils for prostate MRI: impact on prostate distortion and rigid image registration.

Authors:  Yongbok Kim; I-Chow J Hsu; Jean Pouliot; Susan Moyher Noworolski; Daniel B Vigneron; John Kurhanewicz
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

View more
  4 in total

1.  Long-term outcome of magnetic resonance spectroscopic image-directed dose escalation for prostate brachytherapy.

Authors:  Martin T King; Nicola J Nasser; Nitin Mathur; Gil'ad N Cohen; Marisa A Kollmeier; Jasper Yuen; Hebert A Vargas; Xin Pei; Yoshiya Yamada; Kristen L Zakian; Marco Zaider; Michael J Zelefsky
Journal:  Brachytherapy       Date:  2016-04-20       Impact factor: 2.362

2.  Pre-implant magnetic resonance and transrectal ultrasound imaging in high-dose-rate prostate brachytherapy: comparison of prostate volumes, craniocaudal extents, and contours.

Authors:  Simone Grisotto; Annamaria Cerrotta; Brigida Pappalardi; Mauro Carrara; Antonella Messina; Chiara Tenconi; Riccardo Valdagni; Carlo Fallai
Journal:  J Contemp Brachytherapy       Date:  2018-08-31

Review 3.  Brachytherapy in the therapy of prostate cancer - an interesting choice.

Authors:  Janusz Skowronek
Journal:  Contemp Oncol (Pozn)       Date:  2013-11-14

4.  Phase I study of dose escalation to dominant intraprostatic lesions using high-dose-rate brachytherapy.

Authors:  Christopher H Chapman; Steve E Braunstein; Jean Pouliot; Susan M Noworolski; Vivian Weinberg; Adam Cunha; John Kurhanewicz; Alexander R Gottschalk; Mack Iii Roach; I-Chow Hsu
Journal:  J Contemp Brachytherapy       Date:  2018-06-29
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.