Optical coherence tomography after carotid stenting: rate of stent malapposition, plaque prolapse and fibrous cap rupture according to stent design.

G de Donato1, F Setacci, P Sirignano, G Galzerano, A Cappelli, C Setacci.   

Abstract

OBJECTIVES: This study aims to evaluate the rate of stent malapposition, plaque prolapse and fibrous cap rupture detected by optical coherence tomography (OCT) imaging according to carotid stent design.
DESIGN: It was a prospective single-centre study.
MATERIALS AND METHODS: Forty consecutive patients undergoing protected carotid artery stenting (CAS) and high-definition OCT image acquisition were enrolled in the study. OCT frames were analysed off-line, in a dedicated core laboratory by two independent physicians. Cross-sectional OCT images within the stented segment of the internal carotid artery were evaluated at 1-mm intervals for the presence of strut malapposition, plaque prolapse and fibrous cap rupture according to stent design.
RESULTS: Closed-cell design stents (CC) were used in 17 patients (42.5%), open-cell design stents (OC) in 13 (32.5%) and hybrid design stents (Hyb) in 10 (25%). No procedural or post-procedural neurological complications occurred (stroke/death 0% at 30 days). On OCT analysis the frequencies of malapposed struts were higher with CC compared to OC and Hyb (34.5% vs 15% and 16.3%, respectively; p < 0.01). Plaque prolapse was more frequent with OC vs CC (68.6% vs 23.3%; p < 0.01) and vs Hyb stents (30.8%; p < 0.01). Significant differences were also noted in the rates of fibrous cap rupture between CC and OC (24.2% vs 43.8%; p < 0.01), and between CC and Hyb (24.2% vs 39.6%; p < 0.01), but not between OC and Hyb stents (p = 0.4).
CONCLUSION: Intravascular OCT after CAS revealed that micro-defects after stent deployment are frequent and are related to the design of implanted stents. Stent malapposition is more frequent with CC stents, while plaque prolapse is more common with OC stents. It remains, however, unknown whether these figures now detected with OCT are of any clinical and prognostic significance.
Copyright © 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23582886     DOI: 10.1016/j.ejvs.2013.03.005

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  13 in total

Review 1.  Carotid Artery Stenting-Historical Context, Trends, and Innovations.

Authors:  Miloslav Spacek; Josef Veselka
Journal:  Int J Angiol       Date:  2015-08-19

2.  Detection of in-stent protrusion (ISP) by intravascular ultrasound during carotid stenting: Usefulness of stent-in-stent placement for ISP.

Authors:  T Okazaki; S Sakamoto; K Shinagawa; N Ichinose; D Ishii; T Matsushige; Y Kiura; K Kurisu
Journal:  Eur Radiol       Date:  2018-07-19       Impact factor: 5.315

Review 3.  Optical Coherence Tomography in Cerebrovascular Disease: Open up New Horizons.

Authors:  Ran Xu; Qing Zhao; Tao Wang; Yutong Yang; Jichang Luo; Xiao Zhang; Yao Feng; Yan Ma; Adam A Dmytriw; Ge Yang; Shengpan Chen; Bin Yang; Liqun Jiao
Journal:  Transl Stroke Res       Date:  2022-04-21       Impact factor: 6.829

Review 4.  One swallow does not a summer make but many swallows do: accumulating clinical evidence for nearly-eliminated peri-procedural and 30-day complications with mesh-covered stents transforms the carotid revascularisation field.

Authors:  Piotr Musiałek; L Nelson Hopkins; Adnan H Siddiqui
Journal:  Postepy Kardiol Interwencyjnej       Date:  2017-07-19       Impact factor: 1.426

5.  Comparison of closed-cell and hybrid-cell stent designs in carotid artery stenting: clinical and procedural outcomes.

Authors:  Ersan Tatli; Alptug Tokatli; Mehmet Bulent Vatan; Mustafa Tarik Agac; Huseyin Gunduz; Ramazan Akdemir; Harun Kilic
Journal:  Postepy Kardiol Interwencyjnej       Date:  2017-05-25       Impact factor: 1.426

6.  Perspective review on applications of optics in cerebral endovascular neurosurgery.

Authors:  Christopher R Pasarikovski; Jillian Cardinell; Victor X D Yang
Journal:  J Biomed Opt       Date:  2019-03       Impact factor: 3.170

7.  Optical Coherence Tomography Evaluation of Carotid Artery Stenosis and Stenting in Patients With Previous Cervical Radiotherapy.

Authors:  Xiaohui Xu; Feihong Huang; Xuan Shi; Rui Liu; Yunfei Han; Min Li; Fang Wang; Qingwen Yang; Wusheng Zhu; Ruidong Ye; Xinfeng Liu
Journal:  Front Neurosci       Date:  2022-04-27       Impact factor: 4.677

8.  A Parametric Study of Flushing Conditions for Improvement of Angioscopy Visibility.

Authors:  Kohei Mitsuzuka; Yujie Li; Toshio Nakayama; Hitomi Anzai; Daisuke Goanno; Simon Tupin; Mingzi Zhang; Haoran Wang; Kazunori Horie; Makoto Ohta
Journal:  J Funct Biomater       Date:  2022-06-01

9.  Optimal platelet function test for in-stent tissue protrusion following carotid artery stenting.

Authors:  Masanori Tsujimoto; Yukiko Enomoto; Masafumi Miyai; Yusuke Egashira; Toru Iwama
Journal:  J Int Med Res       Date:  2018-03-19       Impact factor: 1.671

10.  Evaluation of spatial distribution and characterization of wall shear stress in carotid sinus based on two-dimensional color Doppler imaging.

Authors:  Bo Zhang; Yuqin Ma; Fang Ding
Journal:  Biomed Eng Online       Date:  2018-10-19       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.